WRO 300 H
HCEN12104 Rev 09.2010
Technical data
43
Power supply
Mains Voltage
100 to 240 V AC (line voltage
range 85 to 264 V AC)
Frequency
50 or 60 Hz (line frequency range
45 to 65 Hz)
Power Consumption Max 1500 W (100 V AC)
1380 W (115 V AC)
1850 W (220-240 V AC)
Appliance inlet
According to IEC 60320; C14
Cable
Cable with protective earth
Conductor length max. 3.5 m
minimum 1 mm² (220-240 V AC)
AWG 16 at 100 and 115 V AC
Mains plug
•
Continental Europe - 230 V
AC, mains plug CEE-7 / VII
•
United Kingdom - 230 V AC,
mains plug BS 1363/A
•
Australia - 240 V AC, mains
plug AS/NZS 3112
•
North America type - 100 and
115 V AC, Hospital grade,
earthed plug, type IEC 60 083;
A5-15
•
China - 220 V AC mains plug
PRC/3
•
Brazil - 220 V AC, mains plug
BR/3
•
Denmark - 230 V AC, mains
plug DK3/HGAS
Fuses
10 A slow blow (220-240 V AC)
15 A slow blow (115 V AC)
15 A slow blow (100 V AC)
Earth Leakage Current
Max. 250 µA (220-240 V AC)
Max. 145 µA (115 V AC)
Max. 140 µA (100 V AC)
Values