WRO 300 H
HCEN12481-02/2012
Operation
23
2
Operation
2.1 Start
Press
until light goes on.
If Remote start/stop is provided, WRO 300 H will start auto-
matically when requested from the dialysis machine.
2.2 Stop
Press
until light goes out and the WRO 300 H goes to
Standby mode.
WARNING
This device does not remove chlorine and chloramines. Carbon filtra-
tion to remove these substances is therefore required if the total chlo-
rine concentration might exceed 0.1 mg/l (ppm). Severe patient injury
may otherwise occur. Ensure by testing, or by other means, that the
total chlorine concentration is below 0.1 mg/l (ppm) prior to initiating
dialysis treatment.
NOTE
will light up and the WRO 300 H will go into Active Cooldown if
the internal water temperature exceeds 45 °C (
will still be lit).
The procedure will continue until the temperature is below 45 °C.
NOTE
If power failure occurs during operation mode the procedure conti-
nues in the same phase as when the power failure occurred when
the WRO 300 H starts up again.
RUN
X
ȝ
S
/
cm
STANDBY