WRO 300 H
30
Regular maintenance
HCEN12481-02/2012
4
Insert the wand connector into the chemical intake port of the
WRO 300 H (see figure below), press firmly and turn the
connector downwards to lock. Check that the connector is
securely in place.
5
CHEM SELECT
is shown in the display and
starts to
flash.
6
Insert the open end of the disinfection wand into a container
with sufficient amount of disinfectant (refer to
“Chemical
disinfectants recommended by Gambro” on page 28
). Check
that the open end of the intake wand is below the solution
level and that the clamp is open.
7
From
CHEM SELECT
, press
to enter the list of available
Cleaning and Chemical disinfection programs.
8
Select preferred Chemical disinfection program by using the
arrow buttons.
Insert the connector and press
Turn to lock
Disinfection Wand Connector
Container
DIALOX
MINNCARE