MPCe/LPCe Users Manual
24
Diagnostic Menu
The diagnostic menu provides screens for calibrating and determining historical events recorded in the event log.
Memory information is also provided for service/support diagnostic information.
Calibrate Supplies
The controller can be calibrated to remove any electrical noise in the system. This provides a more accurate
current measurement given all site specific parameters that may be different than those observed during
production. When initiated, the controller goes through the calibration steps on screen for reference. Click OK
when calibration process is completed.
NOTE:
Prior to initiating this procedure, ensure that ion pump can be turned off and cables disconnected at
the back panel of the controller.
Calibrate Touchscreen
To accurately represent button presses with corresponding information on the screen, the touchscreen may
require calibration. When started, the controller goes through the calibration steps on the screen for reference. It
is important that when pressing the target X calibration buttons that the user press the X firmly and hold for at
least ½ second. An additional display test can be added to verify contrast settings show distinction between all
gray scales.