Page 23 of 28
Gamma Vacuum DIGITEL MPCq Users Manual
Document 900034, Rev B
9.1 Hex
Command
9.2 Description
9.3 Data
Field
9.4 Response
9.5 Data/Response
Description
DE
TSP GET TSP STATUS
Reads TSP status.
R, X
YY
Where:
R is TSP (1-2)
X is option. Currently option 0 is only
supported.
YY is a two digit status code
‘’00” - for TSP Initializing
“01” - for TSP Off
“02” - for TSP Armed (Running)
“03” - for TSP Armed (Waiting, pressure
window, supply is off)
“04” - for TSP Armed (Waiting, pressure
window, pump pressure outside the defined
pressure window)
“05” - for TSP Armed (Waiting, Interlock not
satisfied)
“06” - for TSP Armed (Waiting, Other TSP is
firing)
“07” - for TSP Firing