14
Doppler Quality Control Instruments
Flow instruments and string test objects are two types of Quality Control test tools for Doppler
ultrasound scanners. Flow instruments, such as the Gammex 1425A, produce physiological flow
rates and can be used to perform the full range of quality and acceptance tests. “Flow phantoms
provide a more realistic signal environment in which a spectrum of velocities is produced across
the flow profile of a simulated vessel. Flow instruments can be calibrated in terms of absolute flow
rates (and, therefore, maximum velocities).”
4
Flow phantoms mimic blood flow in the body; the
continuous stream simulates venous flow.
5
The angled vessel is used to measure sensitivity for different frequency transducers and accuracy
of the Doppler angle. The variable flow speeds test the accuracy of the Doppler shift. Good flow
instruments are self-contained and provide an accurate speed of sound. They are also practical for
routine quality control and comparison of ultrasound systems as well as teaching the principles of
Doppler ultrasound.
1
String phantoms are not physiological.
7
String phantoms display a specular-like reflection from the
string surface which could produce uncertainties in specifying the exact Doppler angle.
10
The knot
signal on the spectral display shows as a regular spike, this can interfere with visualization of the
Doppler spectrum.
7
Also, the water tank can produce unwanted reverberations and reflections.
Doppler Guided Tour
Using the Preset Flow Rates
The microprocessor-based flow controller produces flow rates from 1.7 to 12.5 ml/s. To select con-
tinuous flow, select a preset. The Optimizer displays the flow rate in ml/s and a calculated velocity
in cm/s. There are five user-programmable presets for continuous flow rates.
These presets can be adjusted in 0.1 ml/s incremental steps. To adjust the flow rate, use the
“INCREASE” and “DECREASE” buttons. To save a new flow rate as a preset flow rate for future
use, press the SAVE PRESET button and select from the preset buttons numbered 1-5.
The Pulse mode presets are hard set and cannot be adjusted.
Summary of Contents for 1425A
Page 26: ...26 Notes...
Page 27: ...27 Notes...