Section 3. Using the 163-DIG
Prone/Stereotactic Setup (90-degree)
The 163-DIG is suitable for measuring compression force for a wide range of patient setup
orientations. For prone or stereotactic systems that compress at 90-degrees, the 163-DIG can be
used on its side. Optionally, the Display Panel can be removed. Refer to the
on page 5 for more information.
To collect measurements in 90-degree orientation:
1
Power up and zero the 163-DIG flat and stable surface.
2
Rotate the 163-DIG onto its narrow side on a flat and stable surface and zero the scale.
3
Bring the 163-DIG to the Mammography system and very carefully apply pressure to clamp the
scale between the compression paddle and detector.
WARNING: Only let go after the scale has pressure (~10 lbs). Damage from
dropping the Scale is not covered by warranty.
4
Perform measurements per local Quality Standard (ACR, EUREF, etc.)
5
Place hands under 163-DIG at all times when compression force is released.
Figure 3-2. Example setup of 163-DIG on Prone/Stereotactic System. The scale is held in place by
user until the machine applies the appropriate pressure.