Product Overview

1

163-DIG Scale Overview

Product Overview

The Gammex Digital Compression Scale, Model 163-DIG, measures the compression force applied 

by mammography systems.
Accurate compression during mammographic image acquisition is essential to minimizing dose to 

the patient while achieving acceptable image quality. Compression minimizes patient motion, 

reduces scattered radiation, increases image contrast, and diminishes radiation exposure to the 

breast by spreading out and reducing the overall thickness of the breast tissue.
Routine testing of compression force with the Gammex Digital Compression Scale allows clinical 

users to comply with internationally recognized standards and manufacturer recommendations. For 

example, countries that follow ACR & MQSA guidelines can use the 163-DIG to fulfill the MQSA 

requirements for compression performance. For more information, please see the ACR 

Mammography Quality Control Manual, Technologist’s Section on Compression.

Getting Started

Capacity

Assemble the Platform

Add the scale platform to the scale body by aligning the pins with the holes on the body and evenly 

pushing in all the pins. 

Battery & Install

Batteries are not included with the 163-DIG product. Refer to 

Scale Specifications

 on page 6 for the 

type of batteries to use. A display of 

Lo

 or the battery symbol 

(

on the 163-DIG readout 

indicates low batteries and replacements are required.
To install batteries in the 163-DIG:

1

On the bottom of the 163-DIG, press and slide the battery cover.

2

Note the polarity (+) and (-) and insert the batteries in the correct orientation.

3

Ensure all contacts make a strong connection with batteries. If necessary, use a small screw 

driver to bend battery terminals for a better connection.

4

Slide the battery cover into place until clicks shut.

Conditions for Use, Transportation, & Storage

When the 163-DIG is not in regular use or transported, removing one or more batteries is 

recommended to preserve battery life. If transporting, removing all batteries is recommended to 

prevent damage to the battery housing. 
Transportation and storage within a soft case (sold separately) is recommended. Contact your 

Gammex sales associate (

gammexsales@sunnuclear.com

) for a transportation case to fit the 163-

DIG. 
Never store and transport in extreme heat or cold conditions, refer to 

Scale Specifications

 on page 6 

for normal operating conditions. Also avoid high humidity, vibration, dusty, or magnetic 

environments in all cases of use, transport, and storage. Failure to follow guidelines voids the 

product warranty. Refer to 

Scale Specifications

 on page 6 for Operating Temperature.

WARNING:  Overloading the 163-DIG might permanently damage the scale – do not 

overload! The scale capacity is 0-55 lbs. (0-25 kg). 

Summary of Contents for 163-DIG

Page 1: ...i Title LJLWDO RPSUHVVLRQ 6FDOH 0RGHO...

Page 2: ...selling or otherwise distributing any part of this manual without the prior writ ten consent of Gammex Inc is prohibited Gammex Inc reserves the right to make periodic modifications to this manual wit...

Page 3: ...uence 4 Zero Tare 4 Hold Pause 4 Overload Warning 4 Energy Saver 4 Detachable Display Panel 5 Scale Specifications 6 Using the 163 DIG 7 General Operation 7 Standard Setup 0 degree 7 Prone Stereotacti...

Page 4: ...type of batteries to use A display of Lo or the battery symbol on the 163 DIG readout indicates low batteries and replacements are required To install batteries in the 163 DIG 1 On the bottom of the 1...

Page 5: ...t potentially lead to voiding product warranty Independent Failure Mode Warning Do not use results from the 163 DIG to make changes to your system do use the results to perform additional system check...

Page 6: ...n over time You can freeze the display of the maximum force reading by pressing the Pause Hold button Press the Zero button to reset Components Figure 2 1 Model 163 DIG Diagram Table 2 1 Parts and Acc...

Page 7: ...ved the Scale will read a negative value Press Zero Tare to re zero the 163 DIG Hold Pause Hold Pause allows users to freeze the display on the Scale Press it again to cancel or use Zero Tare Overload...

Page 8: ...1 On the bottom push the release tab with arrow indicting direction 2 Wriggle and slide the Display Panel upwards in the direction of the Platform while holding the release tab 3 The separated Displa...

Page 9: ...g 0 1 kg 0 01 kg 1 25 kg Display Units g lb oz kg lb oz Scale Dimensions 8 9 x 8 2 x 2 9 in 225 x 208 x 73 mm Platform Dimensions 7 87 x 6 22 in 200 x 158 mm Foam Compression Block Dimensions 7 25 x 4...

Page 10: ...mpression paddle Optionally place the Foam Compression Block on top of the 163 DIG Platform which is also useful for testing curved paddles 3 Power On the 163 DIG Scale and wait for Initialization to...

Page 11: ...lat and stable surface 2 Rotate the 163 DIG onto its narrow side on a flat and stable surface and zero the scale 3 Bring the 163 DIG to the Mammography system and very carefully apply pressure to clam...

Page 12: ...mponents are subject to wear Calibration Process Required equipment 10 kg certified weight or combination of certified weights Process 1 While the 163 DIG is off press and hold the Unit button while h...

Page 13: ...ter use with a damp cloth if necessary Contacting Gammex Support Technical support is available from Gammex Inc Contact the Support department By telephone 1 321 259 6862 option 3 then 3 By email Glob...

Page 14: ...21 CFR 820 Gammex Quality Management System is certified to 13485 by SGS Operator Responsibility The instructions in this manual are intended for trained personnel The operator is solely responsible f...

Page 15: ...any safety or health related issues or concerns regarding the use of Gammex products contact Gammex directly Modifications to Equipment WARNING No modifications may be made to this unit without the e...

Page 16: ...RUSRUDWH HDGTXDUWHUV LVFRYHU ULYH 0LGGOHWRQ 86 VXQQXFOHDU FRP...

Reviews: