5
Scanning your 404GS LE
•
Always place the phantom on a stable, level surface for scanning.
•
The phantom comes ready to scan. Do not peel off the surface material.
•
Use water or a generous amount of coupling gel to ensure good transmission.
Do not use mineral oil, baby oil or lanolin-based gels as a coupling medium.
Poor transmission is a result of insufficient coupling.
•
Do not press the transducer into the scanning surface.
This damages the
scanning surface and will shorten the life of the phantom. For curved transduc-
ers, use water or a thick gel layer.
•
Clean the scanning surface immediately after use. Use a soft cloth or paper
towel and soap and water, if needed.
Caution:
Do not press the transducer
into the scanning surface.
Summary of Contents for 404GS LE
Page 1: ...404GS LE Precision Small Parts Grey Scale Phantom Gammex User s Guide...
Page 13: ...13...
Page 24: ...24 Notes...