Pin targets are resolved axially if an imaginary horizontal line can be drawn
between the targets without touching either target. The targets on the left are not
resolved. The targets on the right are resolved.
Examine the other axial resolution target groups and compare the resolution at
various depths. Axial resolution may change with depth.
8.
Scan the nearest cystic target group. Each target should be round with a clean
black appearance and well defined edges. Bright specular echoes at the top and
bottom of the targets are normal.
Measure the dimensions of the 6 mm cystic target to check the image geometry.
Use the calipers to measure from top to bottom and side to side. Repeat with the
other cystic target groups as part of the cyst imaging test.
9.
Decrease image depth to the minimum and examine the dead zone target group.
The dead zone targets can be used to measure lateral resolution in the extreme
near field of the transducer. Note the highest dead zone pin you can see should
be the point of reference - Not the scanning surface.
Unresolved
Resolved