Introduction
The 405 GSX LE phantom provides users with a wide range of criteria to evaluate
the performance of their ultrasound machine. The advanced tissue mimicking gel
used in the phantom has been developed to provide a smoother background texture
than conventional gels. The gel reduces backscatter and is fully compatible with the
latest tissue harmonics equipment and technology. The phantom utilizes an advance
composite film scanning surface that improves transmission properties, thus allowing
more of the ultrasound beam to be received and transmitted.
The 405 GSX LE phantom is designed to be used for quality control purposes when
monitoring these important image characteristics:
1.
axial resolution
2.
lateral resolution
3.
dead zone target imaging
4.
vertical and horizontal distance accuracy
5.
image uniformity
The triangular grey scale targets in the phantom can be used to test the resolution of
high performance ultrasound scanners.
Unique to the 405 GSX LE are two horizontal cross fibers located in the middle of the
phantom. These fibers have been included to assist the user in aligning the transducer
and can be used as a reference “marker” to ensure that QA tests are performed
consistently every time.
Limitations of Use
The 405 GSX LE is designed to be used to aid in the quality control testing and monitor-
ing of high performance ultrasound systems only. It is not to be used to make diagnostic
decisions.