Caring for your 405 GSX LE
Phantom comes ready to scan. Do not remove surface material.
Store your 405 GSX LE with cover closed securely.
Always attach the scanning surface cover and store the phantom out of direct sunlight
when it is not in use.
Store your 405 GSX LE at 35º–105ºF (2º–40ºC).
Freezing temperatures will damage the phantom and high temperatures will accelerate
desiccation.
Weigh your 405 GSX LE to monitor desiccation.
Weigh the phantom when you first receive it and then every 6 months. Record the
values on the data sheet.
Do not drop or damage the phantom
.
Return the phantom for inspection and/or repair if it has been dropped or damaged.
Physical damage to the case will cause premature desiccation.
Gammex recommends annual servicing of your 405 GSX LE to ensure proper
operation. Our qualified service technicians will check for desiccation and provide any
needed rejuvenation, scanning/certification to original specifications, and repairs.