Chapter 6: Adding a new Load/Power Supply to Framework
6-10
Key Name
Type
Required
Description
CalCellValue
Float
YES
This is the calibration cell value, in ohms.
CalCell
String
YES
This is the description of the calibration cell which is
used to tell the user what connection to make.
VchRange
Float
YES
This is the voltage, in volts, used to set the Vch range on
the IFC1010 instrument. For the calibration, the range is
usually set to the x100 gain or the 30 mV range.
IchRange
Float
YES
This is the voltage, in volts, used to set the Ich range on
the IFC1010 instrument. For the calibration, the range is
usually set to the x100 gain or the 30 mV range.