Chapter 6: Adding a new Load/Power Supply to Framework
6-19
[AppliedDCTest]
This section is used during calibration to confirm the application of DC values to the cell. Nothing is calibrated
during this step, rather it is just a check to confirm expected DC levels. Make sure the voltage produced is within
the capability of both the power device and the LPI. Ohm’s law, V=IR, dictates the voltage that will be produced for
a given IApplied. The example below depicts the information for a BP4610 power supply from NF Corporation.
Each of the keys for this section will be discussed below.
[AppliedDCTest]
CalCellValue = 0.2
CalCell = 200 mOhm Cal Cell
VchRange = 0.2
IchRange = 3.0
IAppliedDCI = 8.0
TApplied = 5.0
TRest = 20
Enabled = 1