5. Setting and meaning
When the scale is off, press and hold“UNIT”, then press the “ON/OFF”. The scale will now start
in the setting mode. Pressing “UNIT” can change the parameters; pressing “TARE” can change
the values of parameter.
C1----2 Sensitivity setting 0 1 2 3 4, the larger number is, the less sensible but the better stable.
C2----1 Filtering strength setting 0 1 2 3, the larger number is, the lower reaction speed but the
better stable.
C3----1 Auto power off time setting when using batteries. 0: no shutdown; 1: 15minutes; 2:
30minutes; 3: 45minutes; 4: 60minutes.
6. Display information
F----1 Date over display scope
F----2 A/D-switch is defect
F----3 Key is defect
F----5, F----L Sensor signal is defect
F----H Object weight over capacity
C----F calibration error, instability or not at zero point during calibrating
C----H calibration error, items on the pan or at higher zero point during calibrating
E------ Storage calibration data error, re-calibration is necessary
o Calibration zero instruction, when the empty of platform, back to zero shows, if not, it need to
be re-calibrated.