5. Setting and meaning 

 

When the scale is off, press and hold“UNIT”, then press the “ON/OFF”. The scale will now start 
in the setting mode. Pressing “UNIT” can change the parameters; pressing “TARE” can change 
the values of parameter.   

C1----2 Sensitivity setting 0 1 2 3 4, the larger number is, the less sensible but the better stable.   

C2----1 Filtering strength setting 0 1 2 3, the larger number is, the lower reaction speed but the 

better stable.   

C3----1  Auto  power  off  time  setting  when  using  batteries.  0:  no  shutdown;  1:  15minutes;  2: 

30minutes; 3: 45minutes; 4: 60minutes.   

 

 

6. Display information 

 

F----1 Date over display scope   

F----2 A/D-switch is defect   

F----3 Key is defect   

F----5, F----L Sensor signal is defect   

F----H Object weight over capacity   

C----F calibration error, instability or not at zero point during calibrating   

C----H calibration error, items on the pan or at higher zero point during calibrating   

E------ Storage calibration data error, re-calibration is necessary   

o Calibration zero instruction, when the empty of platform, back to zero shows, if not, it need to 

be re-calibrated. 

 

 

7. Packing list 

Operating manual: 1 copy 
Power line: 1 piece 
Scale pan: 1 set 
Warranty card, certification: 1 set 
Power line: 1 piece 
Screw: 2 sets 
Holder: 1 set 
 
 
 

 