Multi Gas Clip Simple Plus User’s Manual
UM MGC-S-PLUS v1.00a
Portable gas detectors you can count on
10 of 18
•
When a bump interval has been programmed into the detector, but a bump test is not yet due
, the current
bump test status will show: “L. BUMP___” (the date of the last bump test) followed by “BUMP XXXd.” (the number
of days until the next scheduled bump test is due). Quickly press the Power/Menu Button while the display shows
“BUMP XXXd.” to force a manual bump test. Audible, visual & vibrating prompting to apply gas will start. The
display will alternate between “BUMP DUE” and “APPLY GAS”. Clip the MGC-S Calibration Cap (provided with the
detector) into place over the sensor ports. Apply gas to the detector at a flow rate of 0.5 LPM and do not disturb
while test is being performed. Once all of the sensors have been tested, the detector will go into alarm. Remove
the MGC-S Calibration Cap and the detector will return to normal operation after a short period of time. The bump
due date will be automatically reset based on the new bump date. The bump test will be recorded in the Bump Log.
Press
Power/Menu Button
•
When a bump test interval has not been programmed into the detector
, the current bump test status will
show: “L. BUMP___” (the date of the last bump test). Quickly press the Power/Menu Button while the display
shows “L. BUMP___” to force a manual bump test. Audible, visual & vibrating prompting to apply gas will start. The
display will alternate between “BUMP XXXd.” and “APPLY GAS”. Clip the MGC-S Calibration Cap (provided with the
detector) into place over the sensor ports. Apply gas to the detector at a flow rate of 0.5 LPM and do not disturb
while test is being performed. Once all of the sensors have been tested, the detector will go into alarm. Remove
the MGC-S Calibration Cap and the detector will return to normal operation after a short period of time. The bump
test will be recorded in the Bump Log.