Multi Gas Clip Simple Plus User’s Manual
UM MGC-S-PLUS v1.00a
Portable gas detectors you can count on
1 of 18
WARNING STATEMENTS/AVERTISSEEMENT
Do not substitute components as this may interfere with the intrinsic safety of the device .
Ne remplacez pas les composants car cela pourrait interférer avec la sécurité intrinsèque de l’appareil .
DO NOT substitute any other battery type than specified and supplied by Gas Clip Technologies .
Only use Gas Clip Technologies parts in the detector . Non-approved parts will void the warranty and are
considered unsafe .
Before each use, check that all sensor and alarm ports are clear of any obstructions, i .e . debris or blockage .
The detector contains a lithium battery that must be disposed of by a qualified recycler . Check local regulations for
proper disposal .
DO NOT use IR communications when an explosive atmosphere may be present .
If you suspect any malfunction or have any technical problems, contact GCT at 1-877-525-0808 .
The battery may present a fire or chemical burn hazard if mistreated . Do not disassemble, heat above 100°C
(212°F) or incinerate . Contact Gas Clip Technologies for replacement instructions . Use of another battery may
present a risk of fire or explosion .
Keep new and used batteries away from children .
Do not expose the detector to sensor poisons such as, but not limited to, alcohol and citrus based products .
Exposure to poisons may impair the accuracy and/or response time of the detector .
If suspected sensor poisoning has occurred, recheck the detector (both calibrate and bump test) .
Caution: Before each day’s usage, sensitivity must be tested on a known concentration of methane equivalent
to 25-50% of full-scale concentration . Accuracy must be within 0-20% of actual . Accuracy may be corrected by
calibrating the detector .
Any rapid up-scale reading, followed by a declining or erratic reading, may indicate a gas concentration beyond
upper scale limit which may be hazardous .
Toute rapide haut de gamme lecture suivie d'une lecture erratique baisse ou peut indiquer une concentration de
gaz supérieure à la limite superieure de l'échelle qui peut être dangereux .
Strong Electromagnetic Interference (EMI) may cause incorrect operations .
Only the combustible gas detection portion of this instrument has been assessed for performance by CSA
International .
READ FIRST BEFORE OPERATION
Gas Clip Technologies (GCT) Multi Gas Clip Simple Plus (MGC-S-PLUS) detectors are personal safety devices designed to
detect the presence of specific toxic gases: carbon monoxide (CO), hydrogen sulfide (H
2
S), oxygen (O
2
) and combustible
gases/Lower Explosive Limit (LEL) . Before operation, ensure that you have been properly trained on the use of the
equipment and the appropriate actions to take in the event of an alarm condition .