SGC and SGC Plus User’s Manual

UM SGC v3.0

1 of 15

WARNING STATEMENTS

 If the detector is past the “Activate Before Date” on the package, please do not activate.
 Do not attempt part replacement or substitution as this could impair the intrinsic safety rating and will void the 

warranty of the product.

 Before daily use check the following: 

1) Sensor and alarm ports are clear of any obstructions i.e. debris or blockage
2) Perform the Self-test to ensure proper operation of visual, audible, and vibrating alarms
3) Confirm receipt of a check mark in the upper left hand corner signaling a successful Self-test
4) Inspect the detector for any physical signs of damage

 DO NOT use IR communications when an explosive atmosphere may be present.
 The battery may present a fire or chemical burn hazard if mistreated. Do not disassemble, heat above 100°C (212°F) 

or incinerate. Contact Gas Clip Technologies for replacement instructions. Use of another battery may present a 
risk of fire or explosion. 

 Keep new and used batteries away from children.
 DO NOT expose the detector to sensor poisons such as, but not limited to: alcohol, citrus-based cleaners, silicones, 

lead compounds (e.g. tetraethyl lead), sulfur compounds, phosphorus, halogenated hydrocarbons and aerosols. 
Exposure to poisons may impair the accuracy and/or response time of the detector. 

 If suspected sensor poisoning has occurred, recheck the detector (both calibrate and bump test).

 Calibrate the SGC O₂ detector at least every 30 days. Make sure to calibrate in a clean air environment. See SGC O₂ 

detector section for instructions on calibration.

 The CO and H

2

S versions of both the Single Gas Clip and SGC Plus do not require calibration for the life of the 

product.

 The detector should be bump tested before each day’s use with a known concentration of gas to confirm its ability 

to respond to gas. The recommended target gas concentrations for the different detector versions are as follows--
H

2

S: 25 ppm, CO: 200 ppm, O₂: 18%. Calibrate the detector if the readings are not within the specified limits.

Bump tests can be performed either manually or through the SGC Dock. If a manual test is to be performed, make 
sure to test in a clean air environment.

 Any rapid up-scale reading followed by a declining or erratic reading may indicate a gas concentration beyond 

upper scale limit which may be hazardous.

 Strong Electromagnetic Interference (EMI) may cause incorrect operations.

  If a detector fails the self-test or bump test, please discontinue use.

 The detector contains a lithium battery that must be disposed of by a qualified recycler. Check local regulations for 

proper disposal.

 Do not substitute internal components as this may interfere with the intrinsic safety of the device.
 Do not attempt to replace the battery or sensor. This product is designed to be disposable. Changing these 

components will void the warranty.

 If you suspect any malfunction or have any technical problems please contact GCT at 877-525-0808.

 

READ FIRST BEFORE OPERATION

Gas Clip Technologies (GCT) Single Gas Clip and SGC Plus detectors are personal safety devices designed to detect the 
presence of specific toxic gases such as Carbon Monoxide (CO), Hydrogen Sulfide (H

2

S) or Oxygen (O₂) deficiency. Before 

operation, please ensure you have been properly trained on the use of the equipment and appropriate actions to take in 
the event of an alarm condition.

Summary of Contents for Multi Gas Clip

Page 1: ...Single Gas Clip Plus Single Gas Clip User s Manual G A S C L I P T E C H C O M...

Page 2: ...n 5 Turning the Detector On 5 Turning the Detector Off 5 Performing a Self test 5 Alarms 7 Detector Maintenance 8 Bump Test Interval bUP 8 O2 Detector Calibration 8 Cleaning 9 Hibernation SGC Plus Onl...

Page 3: ...ion The CO and H2S versions of both the Single Gas Clip and SGC Plus do not require calibration for the life of the product The detector should be bump tested before each day s use with a known concen...

Page 4: ...ur l talonnage Les versions CO et H2S du Single Gas Clip et du SGC Plus ne n cessitent pas d talonnage pendant toute la dur e de vie du produit Le d tecteur doit tre test avant chaque utilisation quot...

Page 5: ...of 15 DETECTOR COMPONENTS ENTRY DESCRIPTION 1 Audible Alarm Port 2 Display 3 Alarm Bar LEDs 4 Maintenance LEDs 5 Power Button 6 Sensor 7 Infrared IR Communication Window 8 Alligator Clip with Safety...

Page 6: ...play Details The detector utilizes a special high viewing angle LCD that is designed to enhance the screen visibility In the absence of gas it displays life remaining In those cases where gas is prese...

Page 7: ...and the sensor is tested Below we outlined the most common screen configurations however if the detector has been programmed via the GCT IR Link or has been exposed to gas additional screens may appea...

Page 8: ...ctor will be reset Note while the value will be cleared on the display the value will be held in the detector s log See the event log section for further details This value can be cleared on the next...

Page 9: ...life Please discontinue use Alarm Types LOW ALARM Audible alarm One 1 slow beep every second Visual alarm One 1 slow flash every second Vibrating alarm One 1 slow vibration every second HIGH ALARM an...

Page 10: ...button press to skip the bump test If the SGC is bumped while showing GAS then it will record as a bump test in the event log instead of as an exposure If no gas is applied it will return to the norma...

Page 11: ...ve the event log by pressing the save event log before hibernating 4 Click on the Hibernate Button acknowledge the event log message 5 Keep in front of the GCT IR Link until the Hibernate OK message i...

Page 12: ...of logs you want to download a Event Logs Only b Events Partial Data Log last week of data c Events New Data Logs Only All data logs from last time you pulled the logs to current d Full Logs approxim...

Page 13: ...CALCAP Replacement calibration cap for the Single Gas Clip or the Single Gas Clip Plus Single Gas Clip Dock P N SGC DOCK portable chargeable all in one docking station in a durable portable case for a...

Page 14: ...ts Newer events replace older events Warranty Full 2 years SGC or 3 years SGC Plus Gas Range Resolution Accuracy T90 Gases H2S 0 100 ppm 0 1 2 ppm 30 s CO 0 300 ppm 1 5 ppm 30 s O 0 30 vol 0 1 0 7 vol...

Page 15: ...with a damp cloth Intrinsically Safe S curit lntrins que Class I Division 1 Groups A B C and D T4 Ex ia IIC T4 Ga Class I Zone 0 AEx ia IIC T4 Ga 40 C H2S 40 C CO 35 C O Ta 50 C CAN CSA C22 2 No 60079...

Page 16: ...arranty length of coverage for products sold without an Activate Before Date shall be two 2 years beginning upon the original end user s documented date of purchase or one 1 year from the product s or...

Page 17: ...aforementioned information Please note that the defective product must be returned to GCT address which is 305 W FM 1382 Suite 540 Cedar Hill TX 75104 with provided prepaid shipping label GCT will ret...

Reviews: