GC-IMS
– User Manual
204/214
Action
When using the CGFU-unit, replace the CGFU-
filters.
Possible Cause
System contamination
Action
Start cleaning mode.
Symtom
Find signals from the prior measurement run in the
chromatogram
Possible Cause
The measurement runtime is too short.
Action
Increase the measurement run time
Symtom
The actual size of the measurement file is to large
Possible Cause
Average setting too low.
Action
Increase the Average setting or shorten the
runtime (recommended averaging: 6).
Symtom
The measurement signals cannot mapped well.
Possible Cause
Average setting is too high.
Action
Decrease the Average setting (recommended
averaging: 6).
Symtom
EPC2 Maximum flow of 150 ml/min is not reached.
Possible Cause
The reachable maximum flow depends on the dimensions of
the installed column.
Action
Get the specified maximum flow from
Analytical approval. If that still does not
help contact the G.A.S. service hotline.
Possible Cause
When using a CGFU unit the total flow (EPC1
and EPC2) is limited to 400ml/min.
Action
No Action
Symtom
Six-Port-
valve doesn’t switch
Possible Cause
Lost connection