GC-IMS
– User Manual
56/214
Averaging display
Displays the current average value. The
averaging value determines how many
raw spectra are averaged to generate
one single spectrum as result in the
stored measurement file. Modifying the
averaging parameter affects the number
of recorded spectra per time period. A
typical average value is 6. The maximum
is set to 99. A value of 0 (Off) disables
averaging.
Drift-Voltage Button
Drift Voltage polarity can be selected and
switched (positive/negative).
Flow display
In Input Field Setpoint the flowrate of Drift
gas (EPC1) and Carrier gas (EPC2) can
be set.
The current flowrates for Drift gas (EPC1)
and Carrier gas (EPC2) are displayed.
The maximum value for EPC1 is 500
ml/min. The maximum value for EPC2 is
150 ml/min. The working value for EPC1
is 150 ml/min. The working value of EPC2
depends of the application measurement
and is influenced by the build-in GC-
Column (length, inner diameter).
Temperature display
In Input Field Setpoint the temperature
values of the heating modules
IMS (T1)
,
Column (T2)
,
Six-Port Valve with
Sample Loop (T3)
,
internal
Transferline (T5)
and
optional external
Transferline (T4)
can be set. The
current Temperatures (T1
–T5) are
displayed. Temperature module T6 is
currently not available. The maximum
setable value is 80 °C.
Drift Voltage mode
Displays the current selected Drift
Voltage state (positive / negative).
Device Parameter
Displays the current temperature, flow
and pressure values.
Summary of Contents for GC-IMS
Page 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Page 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Page 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Page 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Page 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...