TROUBLESHOOTING GUIDE

 

Symptom / 

Issue 

Possible Cause / Resolution 

Pump does not 

ACTIVATE 

 

  If the user experiences issues with pump activation, press/hold the    

MODE key to enter into settings screen. Click ENTER to access 2/2 

and modify 

Flow Rate Level

 to 

5

. If pump is still inactive, pump 

needs to be replaced 

 contac

Support@CO2Meter.com

 

Device does not 

power ON 

Make sure the power supply is connected properly and that 

there is power in the outlet the adapter is connected has the 

appropriate voltage. If using batteries, make sure they are not 

depleted. 

Slow 
response 

Check the air flow channels to make sure they 

are not obstructed. 

Reading Error 

Ensure that the Inlet/Outlet Luer Cap Fittings are 

removed 

to 

enable an accurate means of sampling.  

Red SD Card Icon 

Appears 

This indicates that the SD Card is either damaged, 

missing, or full of data. Replace SD Card.  

LCD Display Turns 

Off During 

Measurements 

If the user experiences LCD display turning off during 

calibration,  measurements,  or data logging 

  please 

reference the 

settings

 section to adjust backlight / auto 

power off from default settings.   

LCD Screen 

Troubleshooting 

During Use 

If the user experiences issues with the screen orientation, 

this setting can be changed by referencing the 

settings

 

section and adjusting the 

screen auto rotate

 to 

NO

.    

Rechargeable Batteries Index

 

‘Battery OK’: 

Measurements 

are possible 

‘Battery Low’: 

The battery needs to be 

recharged, measurements are 

still possible 

Battery 

Exhausted’: 

Measurements are 

not possible 

Battery installation: 

This device is supplied with Li-ion 18650 3.7V rechargeable batteries (3x), which are pre-installed upon device shipment. 

Battery charging: 

During battery charging, the temperature of the device will rise by 5°C~10°C. At this time, the measurements of 

temperature and humidity will be affected by temperature rise. This could cause an effect on the accuracy of 

temperature when charging. Please use a fan to blow toward the Temperature & Humidity Sensor (B ) 

directly, in order to gain a compensated balance of temperature and humidity between that of the 

temperature sensor and surrounding area. (5V/1A USB adapter charger). 

Summary of Contents for CM-1000-00-00-00-25

Page 1: ...Operation Manual Popular Models 100 CO2 CM 1000 00 00 25 00 10 CO2 CM 1000 00 00 00 25 25 O2 CM 1000 00 00 52 00 25 O2 100 CO2 CM 1000 00 52 25 00 Up to 64 Customizable Sensor Configurations Copyrigh...

Page 2: ...GUIDE 7 Powering Operation 7 DATA LOGGING 8 SETTINGS 9 DISPLAY SETTINGS 9 Date Time 1 1 9 Custom Settings 9 USING THE FILTERS 10 CALIBRATION 11 WHY IS CALIBRATION REQUIRED 11 FACTORY CALIBRATION 11 O...

Page 3: ...our local community CO2Meter Inc appreciates your business and looks forward to working with you and your team in the future Please take some time to read through this manual to become familiar with...

Page 4: ...Trap can be purchased separately This filter will allow free flow of sampled air while keeping moisture out of the meter Note Please contact our technical support staff for more details about these ac...

Page 5: ...sting of fire suppression systems The devices built in micro pump ensures reliable operation and consistent readings based on user selected time intervals PRINCIPAL OF OPERATION The sensors inside thi...

Page 6: ...eperiod LCD Display The Full Color Liquid Crystal Display LCD shows the following features CO2 concentration level in parts per million ppm or O2 CO CH4 concentration levels in percentage format RH AM...

Page 7: ...Follow these instructions to ensure proper set up 1 After powered on when pump is activated the device will measure the data in real time 2 Before getting started you need to enter the settings screen...

Page 8: ...bing and filters as required to the top ports see image on page10 3 To begin logging press and hold the PUMP button The inner square of the SD Card Icon in the top right hand corner of the screen will...

Page 9: ...en auto rotate Yes No YES Backlight level 1 32 The larger the number the brighter the screen 11 Background color There are a total of 4 background colors Black Backlight time 10 999 sec Backlight disp...

Page 10: ...y vacuum effect push it through the sensing chamber and then exhaust the air outside the meter through the outlet port Below are examples of the options for exhausting the sampled air closed vs open l...

Page 11: ...hable difference between 1 point and 2 point calibration methods One point sensor calibration involves applying one reference gas whereas 2 point calibration is done by provided two reference gases to...

Page 12: ...the UP E and Down H arrows at the same time to enter the specific Sensor Calibration page 7 Press the Enter button J to highlight Target Sensor inRed 8 Press the Down arrow button H to scroll through...

Page 13: ...ard Weight 480g 1 05 lbs without batteries Dimensions 225 8x99 6x55 mm 8 9 x 3 9 x 2 2 in CM 1000 Pump Pneumatic Data Max Flow 500 ml min Max intermittent duty 380 to 400 mbar Max continuous pressure...

Page 14: ...libration measurements or data logging please reference the settings section to adjust backlight auto power off from default settings LCD Screen Troubleshooting During Use If the user experiences issu...

Page 15: ...he device rises The temperature and humidity sensors are affected Currently measurements are only for reference After batteries are fully charged and the device cools down measurements are reliable 3...

Reviews: