QUICK START GUIDE 

1. Install filters and tubing included if necessary.

2. Turn the meter ON by switching the red power switch up on the left of the device.

3. Turn logging ON by pressing the PUMP button until the SD Card Icon displays a 

green 

inner square.

When the sampling process is completed, you can hold the PUMP button until the SD Card Icon

turns solid 

white.

4.

When the SD Card is White, the pump will stop after 60 seconds of sampling. (user is able to

set the sampling period to any time length).

POWERING THE METER

 

1. Power on: Switch the Red Power Button ( A ) Up, located on the left side of the device, to the UP position.

2. To turn off, Switch the Power Button ( A ) Down.
3. This device uses Li‐ion rechargeable batteries, which can provide a 30+ hours charge.

*Please note - when the device is charged with a USB cable it cannot be turned off*

OPERATION

 

The red POWER switch on the left side of the device turns the unit ON/OFF. The PUMP button on the right side of the 

device’s faceplate, enable/disable the pump/or logging functionality. 

Make sure you read through these instructions thoroughly before using the device. This guide will help you become 

more familiar with the device’s operation. Please read the 

Important Safeguards 

on page 

before continuing. 

IMPORTANT: Follow these instructions to ensure proper set up:

 

1.

After powered on, when pump is activated, the device will measure the data in real‐time.

2.

Before getting started you need to enter the settings screen, by holding the 

Mode Button (I), 

to alter the settings

of the device

.

3.

Utilize the 

UP 

and 

DOWN 

arrow buttons to highlight the appropriate settings you wish to update. Once selected,

the appropriate setting will be highlighted in 

purple 

and can be altered by using the 

UP 

and 

DOWN 

buttons.

Once the appropriate settings have been selected, press and hold the 

MODE 

button to return to the main display

screen.

*Please note, you may wish to navigate to "auto power off" and "back‐light time" in the settings screen to

change the standard 30 second time / 60 minutes period to any custom length interval for non‐
interference*

 

On the main display you will notice 3 zone squares: 

Zone 1 

(top display sensor area) 

Zone 2 

(bottom left ‐ RH) and 

Zone 3 

(bottom right ‐ DP/AL/AMB)

1. To select a zone, use the 

Left (F) / Right (G) arrow buttons 

to select the parameter/ sensor to

be adjusted, the zone will be highlighted in 

red

.

2.

Next, you can press the 

Right arrow button (G) 

to proceed to alter the settings use the 

Up (E) /Down

(H)

arrow buttons to edit the parameter values or alarm levels for that zone.

DISPLAY ALL SENSOR READINGS

 

Press/hold the 

ENTER 

button ( J ) to display all sensor readings. Press the 

MODE 

button ( I ) to return to

the main screen. (See reference chart below, for factory default settings.

Summary of Contents for CM-1000-00-00-00-25

Page 1: ...Operation Manual Popular Models 100 CO2 CM 1000 00 00 25 00 10 CO2 CM 1000 00 00 00 25 25 O2 CM 1000 00 00 52 00 25 O2 100 CO2 CM 1000 00 52 25 00 Up to 64 Customizable Sensor Configurations Copyrigh...

Page 2: ...GUIDE 7 Powering Operation 7 DATA LOGGING 8 SETTINGS 9 DISPLAY SETTINGS 9 Date Time 1 1 9 Custom Settings 9 USING THE FILTERS 10 CALIBRATION 11 WHY IS CALIBRATION REQUIRED 11 FACTORY CALIBRATION 11 O...

Page 3: ...our local community CO2Meter Inc appreciates your business and looks forward to working with you and your team in the future Please take some time to read through this manual to become familiar with...

Page 4: ...Trap can be purchased separately This filter will allow free flow of sampled air while keeping moisture out of the meter Note Please contact our technical support staff for more details about these ac...

Page 5: ...sting of fire suppression systems The devices built in micro pump ensures reliable operation and consistent readings based on user selected time intervals PRINCIPAL OF OPERATION The sensors inside thi...

Page 6: ...eperiod LCD Display The Full Color Liquid Crystal Display LCD shows the following features CO2 concentration level in parts per million ppm or O2 CO CH4 concentration levels in percentage format RH AM...

Page 7: ...Follow these instructions to ensure proper set up 1 After powered on when pump is activated the device will measure the data in real time 2 Before getting started you need to enter the settings screen...

Page 8: ...bing and filters as required to the top ports see image on page10 3 To begin logging press and hold the PUMP button The inner square of the SD Card Icon in the top right hand corner of the screen will...

Page 9: ...en auto rotate Yes No YES Backlight level 1 32 The larger the number the brighter the screen 11 Background color There are a total of 4 background colors Black Backlight time 10 999 sec Backlight disp...

Page 10: ...y vacuum effect push it through the sensing chamber and then exhaust the air outside the meter through the outlet port Below are examples of the options for exhausting the sampled air closed vs open l...

Page 11: ...hable difference between 1 point and 2 point calibration methods One point sensor calibration involves applying one reference gas whereas 2 point calibration is done by provided two reference gases to...

Page 12: ...the UP E and Down H arrows at the same time to enter the specific Sensor Calibration page 7 Press the Enter button J to highlight Target Sensor inRed 8 Press the Down arrow button H to scroll through...

Page 13: ...ard Weight 480g 1 05 lbs without batteries Dimensions 225 8x99 6x55 mm 8 9 x 3 9 x 2 2 in CM 1000 Pump Pneumatic Data Max Flow 500 ml min Max intermittent duty 380 to 400 mbar Max continuous pressure...

Page 14: ...libration measurements or data logging please reference the settings section to adjust backlight auto power off from default settings LCD Screen Troubleshooting During Use If the user experiences issu...

Page 15: ...he device rises The temperature and humidity sensors are affected Currently measurements are only for reference After batteries are fully charged and the device cools down measurements are reliable 3...