background image

GasLab Pro 

Multi Gas

 Sampling 

Data 

Logger

Operating Instructions 

CM-1000

Product Overview

Thank you for selecting

 the

 CM-1000 

multi gas sampling 

data 

logger.

 The CM-1000 is designed to 

simultaneously measure multiple gas concentrations 

through sampling methods

. With long-term data storage, 

the user can assess t

he previous records

. When 

the 

CM-1000 measures the gas concentration reaching the 

alarm setting or higher, the data display and alarm functions are activated. 

The 

CM-1000 is equipped with 

an RS485 interface for connecting to a computer for remote 

monitoring. The

 Multi Gas 

Sampling 

Data 

Logger

 can be used in 

a wide-range of industries and applications. 

Features:

 

V

 

O

ptions for CO

2

 measurement range: 

1%, 

5%, 

10%, 20% 

30%,

 and

 100%

 

V

 

Measure CO

2

, CO, O

2

, CH

4

, RH, AMB, DP, ALTI 

V

 Large LCD for easy reading of gas concentration and temperature and humidity. 

V

 

Audible alarm 

V

 

Data logging with SD card

 

V

 Through the RS485 interface connection, multiple CM-1000 units can be connected to the back-end 

computer system as a security monitoring application.

V

 Built

-

in LCD 

back light

 for easy reading in the dark.

V

 This device is supplied with Li- ion 18650 3.7V rechargeable batteries, can be used for a long time.

 

A. Power Key

B. Temperature & Humidity Sensor

C. LCD

D. Rubber Boot

E. Up Key

F. Left Key

G. Right Key

H. Down Key

I. Mode Key

J. Enter Key

K. Pump Key

L. USB Socket

M. SD Card Slot

N. RS485 Jack

O. Analog / Voltage Output Port

P. Luer Cap

Q. Screw Position

R. Battery Cover

Key description: 

MODE (

I

Enter the setting menu. 

Up (

E

) / Down (

H

) / Left (

F

) / Right (

G

Change parameter values or select options. 

Enter (

J

Execute a command 

PUMP (

K

Start the pump or data logger.

 

LCD display symbol description:

 

Symbol 

Meaning 

Description

CO

2

  concentration,  parts per 

million (ppm)

 

The current CO

2

 concentration 

CO concentration, parts per 

million (ppm) 

The current CO concentration

 

O

2

 concentration, % (percent) 

The current O

2

 concentration

 

CH

concentration,  parts per 

million (ppm)

 

The current CH

4

 concentration

 

Relative humidity 

Relative humidity 

Ambient temperature

 

Ambient temperature

 

Dew point 

The current dew point 

Atmospheric pressure 

Compensate the pressure changes with 

appropriate altitude of location when 

measure. 

Time weighted average (8 hours)  CO

2

, 8-hr time weighted average 

Short

Term Exposure Limit 

(15 min. weighted average) 

CO

2

, 15 min. weighted average 

Maximum value 

Maximum value of gas measurement 

Minimum value 

Minimum value of gas measurement

 

High/low alarm 

Customize the range of color Bar 

Battery capacity 

Battery capacity without segment display. 

Real-time display 

12/24-hour time display 

Warning

 

Need to calibrate the sensor or exceed the 

pump life time. 

Pump 

The pump icon rotates to indicate that the 

pump has been started.

 

Micro SD card 

Micro SD card working status (When the 

icon flashes, it means that the Micro SD 

card is broken or full; when the icon remains 

stationary, it indicates normal operation.)

 

Operation Instructions 

1. Power on:

Switch the Power Key (

A

) on.

2. Measurement:

After power on, the device starts to measure and update the data every second.

Press Enter key (

J

) to select zone. The selection order is: Zone 1

Zone 2

Zone 3. The selected zone will

display a red frame. Press Left (

F

) / Right (

G

) key to select the parameter or sensor to be adjusted. The

selected parameter or sensor displays the red background color, press Right key (

G

) to proceed the setting

and then press Up (

E

) / Down (

H

) key to edit the parameter value.

Zone 1 sensor display: CO

2

, O

2

, CO, CH

4

Zone 2 sensor display: O

2

, CO, CH

4

, RH

Zone 3 sensor display: CO, CH

4

, AMB, DP, ALTI

Zone 1 parameter display and setting 

Item 

Select 1 

Select 2 

Select 3 

Select 4 

Nothing 

MAX 

MIN 

STEL 

TWA 

AL L / AL H 

Press Up (

E

) / Down  (

H

) key to 

edit the parameter value. 

Note: AL L and AL H are expressed as a percentage. Example: 0.04 = 400ppm 

Zone 3 parameter unit selection: 

Sensor 

Unit 1 

Unit 2 

Unit 3 

Unit 4 

DP 

DP=°

DP=°

 

AMB 

°

°

 

ALTI 

mf 

mmHg 

hpa 

3. Display all sensor readings

Press and hold Enter key (

J

) to display all sensor readings. Press Mode key (

I

) to return to the main

screen.

4. Setting

Press and hold Mode key (

I

) to enter the setup menu, then press Mode key (

I

) to continue to the next

page. Press Up (

E

) / Down (

H

) key to select options or change parameter values. And then press Enter key

(

J

) to confirm it. Press and hold Mode key (

I

) to return to the main screen.

The first page: 

Date 

01/01/2017 ~ 2099 

Time 

12/24-hour time 

Continue alarm 

Yes (The buzzer sounds continuously.) / No (The buzzer sounds once.) 

Screen auto rotate 

Yes / No 

Backlight level 

1~32 (The larger the number, the brighter the screen) 

Background color 

□□□

 

(There are a total of 4 background colors.)

Backlight time 

10~999 sec (Backlight display time) 

Auto power off 

1~999 min 

Graph line 2 

CO

2

/ O

2

/CO/ CH

(The second trend line shows the gas measurements.) 

About 

CO2Meter – www.co2meter.com 

The second page: 

Analog for Zone 1 

0-10V/4~20mA (Zone 1 voltage or current output)

Analog for 4-20mA 

CO

2

/O

2

/CO/CH

(Another analog output option) 

Data log time 

Data logging interval time 

Flow rate level 

1~5 (The larger the number, the greater the pump output) 

Exhaust time 

10~240 sec 

Ventilate time 

0~120 sec 

Gas compensation 

Yes / No (CO

and O

2

Sensor times info 

Sensor recalibrated time and pump life time (The time is displayed in the 

countdown mode. If you need to calibrate the sensor or exceed the pump 

life time, the time will be displayed in red.) 

CO

2

: 99999999 days 

O

2

: 99999999 days 

CO: 99999999 days 

CH

4

: 99999999 days 

Pump: 99999999 min 

Reset to factory 

Yes / No (Recover the factory setting to cancel customize setting.) 

Sensor calibration (

+

)  Please refer to the calibration section. 

5. Power off:

Switch the Power Key (

A

) off.

Note: When the device is charged with USB, it cannot be turned off. 

6. Calibration:

(1) Single-point calibration

After entering the setup menu, then press Mode key (

I

) to continue to the next page. Press Down key (

H

) to

select "Sensor calibration (

+

)" and press Enter key (

J

) to confirm it. Press Up (

E

) and Down (

H

) keys

at the same time to proceed the calibration. Then, press Enter key (

J

) to enter the sensor settings.

Target sensor 

CO

2

/O

2

/CO/CH

4

/AMB/RH 

Raw reading 

Unadjusted reading 

Offset 

The amount of offset 

Perform reading 

Adjusted reading 

(2) Two-point calibration (calibrate by professional)

After entering the setup menu, then press Mode key (

I

) to continue to the next page. Firstly, press and hold

Up key (

E

), then press Down key (

H

). ”1” will appear next to "Sensor calibration (

+

)". Press Down

key (

H

) to select "Sensor calibration (

+

1" and press Enter key (

J

) to confirm it. Press Up (

E

) and

Down (

H

) keys at the same time to proceed the calibration. Then, press Enter key (

J

) to enter the sensor

settings.

Target sensor 

CO

2

/O

2

/CO/CH

4

/AMB/RH 

Raw reading 

Unadjusted reading 

Level 1 input 

Low concentration value of standard gas 

Level 2 input 

High concentration value of standard gas 

Level 1 catch 

Connect the low concentration gas to the device, and 

press Pump key (

K

) to start the pump. (Set the data 

log time to less than 1 min). After the reading is 

stable, press Enter key (

J

) to confirm it. 

Level 2 catch 

Connect the high concentration gas to the device, 

and press Pump key (

K

) to start the pump. (Set the 

data log time to less than 1 min). After the reading is 

stable, press Enter key (

J

) to confirm it. 

Cali Set 

When "Start" turns red, press Enter key (

J

) to 

confirm the calibration. 

Gain 

The amount of Gain. 

Offset 

The amount of offset 

Perform reading 

Adjusted reading 

(Display Features and Modes) 