GasLab Pro
Multi Gas
Sampling
Data
Logger
Operating Instructions
CM-1000
Product Overview
Thank you for selecting
the
CM-1000
multi gas sampling
data
logger.
The CM-1000 is designed to
simultaneously measure multiple gas concentrations
through sampling methods
. With long-term data storage,
the user can assess t
he previous records
. When
the
CM-1000 measures the gas concentration reaching the
alarm setting or higher, the data display and alarm functions are activated.
The
CM-1000 is equipped with
an RS485 interface for connecting to a computer for remote
monitoring. The
Multi Gas
Sampling
Data
Logger
can be used in
a wide-range of industries and applications.
Features:
V
□
O
ptions for CO
2
measurement range:
1%,
5%,
10%, 20%
30%,
and
100%
V
□
Measure CO
2
, CO, O
2
, CH
4
, RH, AMB, DP, ALTI
V
□
Large LCD for easy reading of gas concentration and temperature and humidity.
V
□
Audible alarm
V
□
Data logging with SD card
V
□
Through the RS485 interface connection, multiple CM-1000 units can be connected to the back-end
computer system as a security monitoring application.
V
□
Built
-
in LCD
back light
for easy reading in the dark.
V
□
This device is supplied with Li- ion 18650 3.7V rechargeable batteries, can be used for a long time.
A. Power Key
B. Temperature & Humidity Sensor
C. LCD
D. Rubber Boot
E. Up Key
F. Left Key
G. Right Key
H. Down Key
I. Mode Key
J. Enter Key
K. Pump Key
L. USB Socket
M. SD Card Slot
N. RS485 Jack
O. Analog / Voltage Output Port
P. Luer Cap
Q. Screw Position
R. Battery Cover
Key description:
MODE (
○
I
)
Enter the setting menu.
Up (
○
E
) / Down (
○
H
) / Left (
○
F
) / Right (
○
G
)
Change parameter values or select options.
Enter (
○
J
)
Execute a command
PUMP (
○
K
)
Start the pump or data logger.
LCD display symbol description:
Symbol
Meaning
Description
CO
2
concentration, parts per
million (ppm)
The current CO
2
concentration
CO concentration, parts per
million (ppm)
The current CO concentration
O
2
concentration, % (percent)
The current O
2
concentration
CH
4
concentration, parts per
million (ppm)
The current CH
4
concentration
Relative humidity
Relative humidity
Ambient temperature
Ambient temperature
Dew point
The current dew point
Atmospheric pressure
Compensate the pressure changes with
appropriate altitude of location when
measure.
Time weighted average (8 hours) CO
2
, 8-hr time weighted average
Short
‐
Term Exposure Limit
(15 min. weighted average)
CO
2
, 15 min. weighted average
Maximum value
Maximum value of gas measurement
Minimum value
Minimum value of gas measurement
High/low alarm
Customize the range of color Bar
Battery capacity
Battery capacity without segment display.
Real-time display
12/24-hour time display
Warning
Need to calibrate the sensor or exceed the
pump life time.
Pump
The pump icon rotates to indicate that the
pump has been started.
Micro SD card
Micro SD card working status (When the
icon flashes, it means that the Micro SD
card is broken or full; when the icon remains
stationary, it indicates normal operation.)
Operation Instructions
1. Power on:
Switch the Power Key (
○
A
) on.
2. Measurement:
After power on, the device starts to measure and update the data every second.
Press Enter key (
○
J
) to select zone. The selection order is: Zone 1
Zone 2
Zone 3. The selected zone will
display a red frame. Press Left (
○
F
) / Right (
○
G
) key to select the parameter or sensor to be adjusted. The
selected parameter or sensor displays the red background color, press Right key (
○
G
) to proceed the setting
and then press Up (
○
E
) / Down (
○
H
) key to edit the parameter value.
Zone 1 sensor display: CO
2
, O
2
, CO, CH
4
Zone 2 sensor display: O
2
, CO, CH
4
, RH
Zone 3 sensor display: CO, CH
4
, AMB, DP, ALTI
Zone 1 parameter display and setting
Item
Select 1
Select 2
Select 3
Select 4
Nothing
MAX
MIN
STEL
TWA
AL L / AL H
Press Up (
○
E
) / Down (
○
H
) key to
edit the parameter value.
–
Note: AL L and AL H are expressed as a percentage. Example: 0.04 = 400ppm
Zone 3 parameter unit selection:
Sensor
Unit 1
Unit 2
Unit 3
Unit 4
DP
DP=°C
DP=°F
–
–
AMB
°C
°F
–
–
ALTI
m
mf
mmHg
hpa
3. Display all sensor readings
Press and hold Enter key (
○
J
) to display all sensor readings. Press Mode key (
○
I
) to return to the main
screen.
4. Setting
Press and hold Mode key (
○
I
) to enter the setup menu, then press Mode key (
○
I
) to continue to the next
page. Press Up (
○
E
) / Down (
○
H
) key to select options or change parameter values. And then press Enter key
(
○
J
) to confirm it. Press and hold Mode key (
○
I
) to return to the main screen.
The first page:
Date
01/01/2017 ~ 2099
Time
12/24-hour time
Continue alarm
Yes (The buzzer sounds continuously.) / No (The buzzer sounds once.)
Screen auto rotate
Yes / No
Backlight level
1~32 (The larger the number, the brighter the screen)
Background color
□
□□□
(There are a total of 4 background colors.)
Backlight time
10~999 sec (Backlight display time)
Auto power off
1~999 min
Graph line 2
CO
2
/ O
2
/CO/ CH
4
(The second trend line shows the gas measurements.)
About
CO2Meter – www.co2meter.com
The second page:
Analog for Zone 1
0-10V/4~20mA (Zone 1 voltage or current output)
Analog for 4-20mA
CO
2
/O
2
/CO/CH
4
(Another analog output option)
Data log time
Data logging interval time
Flow rate level
1~5 (The larger the number, the greater the pump output)
Exhaust time
10~240 sec
Ventilate time
0~120 sec
Gas compensation
Yes / No (CO
2
and O
2
)
Sensor times info
Sensor recalibrated time and pump life time (The time is displayed in the
countdown mode. If you need to calibrate the sensor or exceed the pump
life time, the time will be displayed in red.)
CO
2
: 99999999 days
O
2
: 99999999 days
CO: 99999999 days
CH
4
: 99999999 days
Pump: 99999999 min
Reset to factory
Yes / No (Recover the factory setting to cancel customize setting.)
Sensor calibration (
▲
+
▼
) Please refer to the calibration section.
5. Power off:
Switch the Power Key (
○
A
) off.
Note: When the device is charged with USB, it cannot be turned off.
6. Calibration:
(1) Single-point calibration
After entering the setup menu, then press Mode key (
○
I
) to continue to the next page. Press Down key (
○
H
) to
select "Sensor calibration (
▲
+
▼
)" and press Enter key (
○
J
) to confirm it. Press Up (
○
E
) and Down (
○
H
) keys
at the same time to proceed the calibration. Then, press Enter key (
○
J
) to enter the sensor settings.
Target sensor
CO
2
/O
2
/CO/CH
4
/AMB/RH
Raw reading
Unadjusted reading
Offset
The amount of offset
Perform reading
Adjusted reading
(2) Two-point calibration (calibrate by professional)
After entering the setup menu, then press Mode key (
○
I
) to continue to the next page. Firstly, press and hold
Up key (
○
E
), then press Down key (
○
H
). ”1” will appear next to "Sensor calibration (
▲
+
▼
)". Press Down
key (
○
H
) to select "Sensor calibration (
▲
+
▼
) 1" and press Enter key (
○
J
) to confirm it. Press Up (
○
E
) and
Down (
○
H
) keys at the same time to proceed the calibration. Then, press Enter key (
○
J
) to enter the sensor
settings.
Target sensor
CO
2
/O
2
/CO/CH
4
/AMB/RH
Raw reading
Unadjusted reading
Level 1 input
Low concentration value of standard gas
Level 2 input
High concentration value of standard gas
Level 1 catch
Connect the low concentration gas to the device, and
press Pump key (
○
K
) to start the pump. (Set the data
log time to less than 1 min). After the reading is
stable, press Enter key (
○
J
) to confirm it.
Level 2 catch
Connect the high concentration gas to the device,
and press Pump key (
○
K
) to start the pump. (Set the
data log time to less than 1 min). After the reading is
stable, press Enter key (
○
J
) to confirm it.
Cali Set
When "Start" turns red, press Enter key (
○
J
) to
confirm the calibration.
Gain
The amount of Gain.
Offset
The amount of offset
Perform reading
Adjusted reading
(Display Features and Modes)