Rechargeable Batteries 

Battery message: 

‘Battery OK’:   

Measurements are possible 

‘Battery Low’:   

The battery needs to be recharged, 

measurements are still possible 

‘Battery Exhausted’: 

Measurements are not 

possible 

Battery installation:   

This device is supplied with Li-ion 18650 3.7V rechargeable battery *3 pcs. 

Please confirm whether bat polarity are Li ion18650 positive bump specification and install batteries in 

the correct polarity. If the battery polarity is reversed, the capacity will be insufficient and the battery life will be 

shortened. 

Battery charging: 

During battery charging, the temperature of the device will rise by 5°C~10°C. At this time, the measurements 

of temperature and humidity will be affected by temperature rise.  Cause  an impact on the accuracy of 

temperature when charging, please use a fan to blow toward the Temperature & Humidity Sensor (

B

) directly 

in order to  get a compensated balance  of  temperature and humidity between temperature  sensor  and 

surrounding area. (5V/1A USB adapter charger) 

Safety Instructions 

Warning: Your safety is very important to us. To ensure use of the device correctly and safely, we would like to 

draw your attention to read the warning and entire User Manual before using the device. These are important 

safety information and should always be observed . 

1. Please handle the device lightly, do not subject the device to impact or shock.

2. Do not immerse the device in water. Water can cause electric shock, fire or malfunction which may result in

damage.

3. Do not keep the device under the hot and moisture environment. Keep the device away from the heat source

or near water.

4. Please use a standard USB power supply (such as PC's USB port, universal AC adapter with USB port).

Improper power supply can cause serious damage to the device, or result in injury or death to the user.

Product Care 

To ensure you receive the maximum benefit from using this device, please observe the follow guidelines.

 

1.

The maximum capacity of the SD card is 16G.

2.

During battery charging, the temperature of the device rises. The temperature and humidity sensors are

affected. At this time, measurements are only for reference. After batteries are fully charged and the device

cools down, measurements are reliable.

3. Repair – Do not attempt to repair the device or modify the circuitry by yourself. Please contact with the local

dealer or a qualified repairman if the device needs servicing. 

4.

Caution – The CO sensor must be replaced every 3 years.

5. Cleaning – Disconnect the power before clean. Use a damp cloth, do not use the liquid cleaning agent, such

as benzene, thinner or aerosols.   

6.

Maintenance

 – 

Recommend that the user conducts a comprehensive test and calibration every year to

ensure the normal operation of the device. 

Specifications 

Measurement Range & 

Accuracy 

CO

2

: 1%, 5%, 

10%, 20%, 

30%,

 

100% 

CO: 0~1,000ppm 
O

2

: 0~25%

< 2% FS / 0.1 mbar 

CH

4

: 5%

 

0-1%: ≤±0.1% 

RH: 0~100%, ±

3

% @25°C (20~80%RH)

, others ±5%

 

Temperature: 

0

~50°C,±1

 

Barometer: 50~110

kPa

±

0.4

kPa

 

Display Resolution 

CO

2

: 10ppm 

CO: 1ppm 
O

2

: 0.01% 

CH

4

:

 

0.01% 

RH: 0.01% 
Temperature: 0.01°C 
Barometer: 0.1mmHg 

Alarm Volume 

80db±5%@10cm 

Operating Conditions 

0~50°C(32~122°F), 0~95%RH, non-condensing 

Storage Temp. 

-20~60°C(-4~140°F)

Comm. Interface 

RS485 ModBus BR19200

N

8

Power Supply 

Rechargeable Battery : Li-ion 18650 3.7V *3 
AC Adapter : 5V±5% >1A, 100~240 VAC, 50/60 Hz 

Storage Capacity 

depend on SD card capacity (max.16G SD card) 

Weight 

480g (without batteries) 

Dimension 

225.8x99.6x55 mm (8.9x3.9x2.2 inch) 

Note: After power-on, it would take 20 minutes for the device to stably measure the temperature and 

relative humidity. 

EMC/RFI: Readings may be affected if the unit is operated within radio frequency electromagnetic field 

strength of approximately 3 volts per meter, but the performance of the instrument will not be 

permanently affected. 

Package contents 

CM-1000 

Multi Gas Sampling Logger

Manual

Batteries

USB 1.5M cable