5. Man down alarm  

Falling by breathing dangerous gases can cause serious injury and even fatality to workers. If the 

Man down alarm function in SAN-11 is set on, SAN-11 can detect falls and send a man down alert 
which will activate the audible and visual alarms and alert other people in the area.  

  

The  man-down  detection  uses  a  three-

axis  accelerometer  to  automatically  monitor  the  user’s 

movements  in  order  to  identify  a  sudden  fall/impact  and  a  lack  of  movement  for  a  period  of  6 
seconds.  

Once alert, people can turn off the current man-down alarm by pressing any one of the two buttons.  

  

6. Reset Button  

Users can reset the unit by pushing a reset button through a hole on back of shell.  

  

  

MAINTENANCE   

Calibration   

The SAN-11 comes pre-calibrated from the factory. However, the O2 sensor should be calibrated 

at least once a year, or as described in your company’s safety procedures. You can perform the 
calibration yourself, or you can return it to GasLab for factory calibration at a nominal fee.  The 
temperature and humidity sensors do not require calibration and should remain accurate for the 
life of your unit.   
  

Cleaning and Storage   

Apply  sparingly  with  a  soft  cloth  and  allow  drying completely  before  using.  Do  not  use  soap  or 

Alcohol cleaning. Do not use aromatic hydrocarbons or chlorinated solvents for cleaning.  

SPECIFICATIONS  

Method - Dual Beam NDIR (Non-dispersive-Infrared)  

Summary of Contents for CO2Meter SAN-11

Page 1: ...User Manual R1 0 CO2Meter Inc Personal Safety CO2 Monitor with Data Logging SAN 11...

Page 2: ...Contents INTRODUCTION 3 FEATURES 3 MONITOR 3 LCD DISPLAY 4 OPERATION 4 MAINTENANCE 7 SPECIFICATIONS 7 TROUBLESHOOTING 8 SUPPORT 10 WARRANTY 10 LIABILITY 10 RETURNS 10 CONTACT US 11 Copyright 2019 Gas...

Page 3: ...isplay Rugged design with protective rubber enclosure Heavy duty metal clip Front facing sensor unit Rechargeable Battery 4 2v 1500mAh Micro USB cable and Wall USB charger Rechargeable Li Ion battery...

Page 4: ...CD DISPLAY 1 Low alarm indicator 2 High alarm indicator 3 Battery indicator 4 CO2 concentration 5 CO2 concentration units ppm or when 9 999 ppm 6 Calibration icon 7 Man down alarm 8 Air Temperature 9...

Page 5: ...d in the following table Menu Items Functional Description AIR AIR means Ambient Air Calibration User presses the power button to implement Ambient Air Calibration 300 600ppm N2 N2 means Nitrogen Cali...

Page 6: ...presses the power button to exit the menu status 3 Alarm Threshold There are high and low alarm thresholds in SAN 11 Both high and low alarm has five thresholds 5000 ppm 5000 ppm TWA 1 0 1 5 and 3 0...

Page 7: ...g any one of the two buttons 6 Reset Button Users can reset the unit by pushing a reset button through a hole on back of shell MAINTENANCE Calibration The SAN 11 comes pre calibrated from the factory...

Page 8: ...lity 20ppm 400ppm Temperature Dependence Typ 0 3 of reading per C or 4ppm per C whichever is greater referenced to 25 C Pressure Dependence 0 13 of reading per mmHg Response Time About 2 min for 90 of...

Page 9: ...y that the ftd2xx dll file is in the same folder as the executable program Try using a different USB cable Windows reports unknown device error Go to Gaslab com select the App Notes tab and follow the...

Page 10: ...ve we will repair or replace it at no additional cost to you This warranty does not cover any products that have been subjected to misuse neglect accident modifications or repairs by you or by a third...

Page 11: ...e etc or if it is beyond the warranty date we will inform you of the cost to fix or replace the product For additional warranty information visit our website gaslab com pages faq CONTACT US We are her...