INSTALLATION
Your JUN-AIR compressor is easy to operate . Observe the
instructions and you will get many years service from your
compressor .
•
Visually inspect unit for shipping damage, contact your supplier
immediately if you think the unit may have been damaged .
•
Check that the performance of the compressor matches the
actual air consumption, please refer to Technical Specifications .
•
Check that the rating plate of the compressor corresponds with
the electrical voltage offered and check that fusing is adequate .
Placing
Place the compressor in a dust-free, dry and cool, yet frost-free room .
Sufficient cooling from the surroundings is important .
•
Ambient temperature: 0°C to 40°C, 32°F to 104°F
•
Relative humidity: Max 90%
Mount the two distance bolts (1) on the back of the cabinet to ensure
sufficient ventilation .
Installation
•
Mount the drain bottle (2) visibly outside the cabinet and mount
the hose at the back of the cabinet (3) .
•
Connect the cable at the back of the cabinet (5) .
•
Plug the compressor into a standard outlet switch .
•
Connect equipment at the back of the cabinet (4) .
•
The front wheels are delivered with brakes . Brake the wheels
before starting the compressor (6) .
Thermo Switch
The fans are controlled by a Stego thermo switch adjusted to 30°C
from the factory – do not alter this setting . Check that the setting is
correct and adjust if necessary .
The fans will start when the temperature inside the cabinet exceeds
30°C and will run continuously until the temperature drops below
30°C .