6.  When this extension hose is used, sampling time may be different from the instruction manual of detector 

tube. Also, correction factor may have to be applied to tube reading. (See below chart for sampling time 

and correction factor.) As necessary, multiply the correction factors of temperature, humidity, pump 

strokes, extension hose, and atomospheric pressure.(When measuring a gas listed on “APPLICATION 

FOR OTHER SUBSTANCES”in the instruction manual, apply the correction factor/scale after these 

corrections.) Refer to the instruction manual of your Gastec Gas Detector tube for temperature, humidity, 

and pump strokes correction. 

 

 

 

Sampling time and correction factors for No.351A-30 

(Sampling times are approximate value.) 

    Currently only the detector tubes listed below are applicable to No. 351A-30. If you need to use other 

tube, please contact your Gastec representatives. 

 

Tube No. 

Gas or Vapour to be Measured 

Sampling time 

for 1 pump 

stroke

100mL

 

Sampling time for a 

half pump stroke

50mL

 

Correction factor 

1L 

Carbon monoxide 

1.5min. 

45sec. 

1.13 

1LK 

Carbon monoxide 

3min. 

1.5min. 

NONE 

1LL 

Carbon monoxide 

3min. 

 

 

1.14 

2HH 

Carbon dioxide 

2min. 

1min. 

1.12 

2H 

Carbon dioxide 

1min. 

30sec. 

1.12 

2L 

Carbon dioxide 

3min. 

1.5min. 

NONE 

2LC 

Carbon dioxide 

3min. 

1.5min. 

NONE 

2LL 

Carbon dioxide 

2min. 

 

 

NONE 

3M 

Ammonia 

1.5min. 

45sec. 

NONE 

3L 

Ammonia 

1min. 

30sec. 

1.11 

3La 

Ammonia 

45sec. 

30sec. 

NONE 

4HM 

Hydrogen sulphide 

1.5min. 

45sec. 

NONE 

4HH 

Hydrogen sulphide 

1min. 

30sec. 

NONE 

4M 

Hydrogen sulphide 

1.5min. 

45sec. 

NONE 

4H 

Hydrogen sulphide 

1min. 

30sec. 

NONE 

4L 

Hydrogen sulphide 

1.5min. 

45sec. 

NONE 

4LL 

Hydrogen sulphide 

1min. 

30sec. 

NONE 

4LK 

Hydrogen sulphide 

1min. 

30sec. 

NONE 

4LB 

Hydrogen sulphide 

1min. 

30sec. 

NONE 

4LT 

Hydrogen sulphide 

1.5min. 

45sec. 

NONE 

5La 

Sulphur dioxide 

1min. 

 

 

NONE 

5Lb 

Sulphur dioxide 

1min. 

 

 

1.12 

5LC 

Sulphur dioxide 

3min. 

   

NONE 

6L 

Water vapour 

30sec. 

30sec. 

NONE 

7H 

Phosphine 

1min. 

30sec. 

NONE 

7J 

Phosphine 

1min. 

30sec. 

1.29 

Phosphine 

1.5min. 

 

 

NONE 

7LA 

Phosphine 

1.5min. 

 

 

NONE 

7L 

Phosphine 

1.5min. 

 

 

NONE 

9L 

Nitrogen dioxide 

1min. 

   

1.60 

11L 

Nitrogen oxides 

2min. 

 

 

1.35 

15L 

Nitric acid 

45sec. 

30sec. 

1.33 

19La 

Arsine 

2min. 

 

 

NONE 

30 

Hydrogen 

3min. 

   

NONE 

35 

Sulphuric acid 

1.5min. 

 

 

NONE 

40 

Mercury vapour 

2min. 

1min. 

1.20 

60 

Phenol 

2min. 

1min. 

1.12 

70 

Mercaptans 

2min. 

 

 

NONE 

Summary of Contents for 351A-30

Page 1: ...m φ1 5 mm Weight 900g approx Applicable with GASTEC Gas sampling pump model GV 100 or GV 110 Optional parts Rubber shroud No 358 MEASUREMENT PROCEDURE Read the instruction manual of your Gastec Gas sampling pump and the instruction manual of your Gastec Gas detector tube carefully 1 Replace the rubber inlet flange clamping nut of the sampling pump by the metal clamping nut of the extension hose Th...

Page 2: ...n monoxide 1 5min 45sec 1 13 1LK Carbon monoxide 3min 1 5min NONE 1LL Carbon monoxide 3min 1 14 2HH Carbon dioxide 2min 1min 1 12 2H Carbon dioxide 1min 30sec 1 12 2L Carbon dioxide 3min 1 5min NONE 2LC Carbon dioxide 3min 1 5min NONE 2LL Carbon dioxide 2min NONE 3M Ammonia 1 5min 45sec NONE 3L Ammonia 1min 30sec 1 11 3La Ammonia 45sec 30sec NONE 4HM Hydrogen sulphide 1 5min 45sec NONE 4HH Hydroge...

Page 3: ... 118 Cyclohexanol 2min 1 35 120 Aromatic hydrocarbons 1 5min 45sec 1 35 121 Benzene 1 5min 1 22 121S Benzene 2min 1 53 121SL Benzene 2min 1 30 121L Benzene 3min 1 19 122 Toluene 2min 1min 1 35 122L Toluene 1 5min 1 22 123 Xylene 2min 1min 1 33 123L Xylene 1 5min 1 47 124 Styrene 1min 30sec 1 48 124L Styrene 1min 30sec 1 46 131 Vinyl chloride 2min 1min 1 17 131L Vinyl chloride 1 5min 1 25 131La Vin...

Page 4: ...3 Butadiene 1 5min NONE 174L 1 3 Butadiene 3min NONE 191L Acrylonitrile 2min NONE 230 Methyl iodide 1 5min 45sec 1 17 230H Methyl iodide 2min 1min 1 28 231 Sulphuryl fluoride 2min NONE 232 1 2 Dichloroethane 3min 1 13 233 Chloropicrin 2min 1 12 WARRANTY If you have any questions regarding gas detection and quality of the products please feel free to contact your Gastec representatives Manufacturer...