background image

 

Extension hose No.351A-5

No.351A-10 

Instruction manual

 

FOR SAFE OPERATION

 

Read this manual, the instruction manual of your Gastec Gas sampling pump, and the instruction manual of your 
Gastec Gas detector tube carefully.   

CAUTION

 

If not observed, injuries to the operator or damage to the product may result. 

 

When breaking the tube ends, keep away from eyes. 

 

Do not touch the broken glass tubes, pieces and reagent with bare hand(s). 

NOTE

 

For maintaining performance and reliability of the test result. 

 

Use only Gastec Gas Sampling Pump (GV-100 or GV-110). 

 

Refer to the instruction manual of your Gastec Gas Detector tube for use temperature 
and humidity range. 

 

Avoid sunlight, acids, and alkaline materials. 

 

Specifications 

Product name 

5 m Extension hose 

10 m Extension hose 

Product code 

No.351A-5 

No.351A-10 

Length 

5m±2% 

10m±2% 

Weight 

200g (approx.) 

340g (approx.) 

Outer / Inner diameter 

φ5 mm / φ1.5 mm 

Applicable with 

GASTEC Gas sampling pump model GV-100 or GV-110. 

Optional parts 

Rubber shroud No.358 

 

 

 

MEASUREMENT PROCEDURE

(Read the instruction manual of your Gastec Gas sampling pump and the 

instruction manual of your Gastec Gas detector tube carefully.) 
1. 

Replace the rubber inlet flange clamping nut of the sampling pump by the metal clamping nut of the extension 
hose. The metal clamping nut should be firmly tightened to eliminate leakage caused by poor seal between the 
flange of the rubber inlet and the pump body. 

2. 

Insert the nipple of the extension hose securely into the pump inlet. 

3. 

Tighten the metal retaining nut firmly and fix the tubing on the pup head. 

4. 

After breaking off tube tips, place the tube into the rubber shroud of the extension hose. The tube must be 
attached in the holder so that sample is drawn into the pump in the direction of the arrow (►)printed on the 
tube. 

5. 

When use of twin tube, connect rubber tube shroud No. 358 which is sold seperately.   

 

6. 

As necessary, multiply the correction factors of temperature, humidity, pump strokes, and atomospheric 
pressure. Refer to the instruction manual of your Gastec Gas Detector tube for temperature, humidity, and 
pump strokes correction.

 

 

WARRANTY

 

If you have any questions regarding gas detection and quality of the products, please feel free to contact your Gastec 
representatives. 

 

Manufacturer: Gastec Corporation 

8-8-6 Fukayanaka, Ayase-City, Kanagawa 252-1195, Japan 

http://www.gastec.co.jp 
Tel81-467-79-3010 Fac81-467-79-3979 

Printed

 

in

 

Japan

 

Summary of Contents for 351A-5

Page 1: ... Gastec Gas detector tube carefully 1 Replace the rubber inlet flange clamping nut of the sampling pump by the metal clamping nut of the extension hose The metal clamping nut should be firmly tightened to eliminate leakage caused by poor seal between the flange of the rubber inlet and the pump body 2 Insert the nipple of the extension hose securely into the pump inlet 3 Tighten the metal retaining...

Page 2: ...順 検知管及び検知管式気体測定器の取扱説明書を参照してください 1 気体採取器の入口ナットを取り外し 代わりに採取管接手 A を取り付けてください この時採取管接手 A を 充分に締め付けないと漏れの原因になりますのでしっかり締め付けてください 2 採取管ニップルを採取器インレットゴムにしっかり差し込んでください 3 採取管接手 B を採取管接手 A にしっかり締め付けてください 4 検知管の両端をチップブレーカーで折り取った後 延長採取管のガードゴム 先端部 に検知管の矢印 が 採取器に向くようにセットしてください 5 二連管検知管 2 本接続して使用する検知管 を使用する場合は 別売りのガードゴム No 358 を接続して ください 6 検知管に補正が必要な場合には 読み値に対して温度 湿度 吸引 気圧の順にそれぞれの係数を掛けて ください 温度 湿度 吸引の補正係数の有無は各検知管の...

Reviews: