41
④
Press "START/STOP Button".
Press
⑤
The pump starts. "CLF" blinks for about 10 seconds,
and then the instantaneous flow rate which the
current calibration value is applied to is displayed.
↓
⑥
Wait until the flow rate displayed on GSP-501FT
stabilizes, and then measure the flow rate with the
calibrator.
Press
or
[If the adjustable range is exceeded]
⑦
Set the flow rate value of the pump to the measured
value of the calibrator by pressing the "INCREASE
Button" and "DECREASE Button". (Pressing the
buttons changes the flow rate value of this pump.)
※
When the adjustable range is exceeded, "C.Err" is
displayed.