46
●
When previous sampling results are being displayed.
Error
Message
Status
Lamp
Name
Cause and meaning
Countermeasure
Flashing
red light
Battey
Error
During the last
sampling, the pump
stopped due to a drop in
the battery voltage.
Replace with new AA alkaline
batteries or recharged AA
nickel-metal hydride batteries.
(Refer to "5.1 Prepare the
Power Supply")
Flashing
red light
Clogging
The pump stopped due
to clogging, etc.
Check if there is no clogging
in the sampling path, filter,
etc.
(Refer to "8.1 Replacing the
Dust Filter and O-ring")
Off
Sampling
time/
Integration
Error
●
When timer mode
If the measured
integrated value falls out
of ±5% of the value
calculated from
multiplying the set
sampling time by the
flow rate, the error
message is displayed
after the pump stops
●
When Volume Mode
If the actual sampling
time is out of ±5% of
the value calculated
from the set integrated
value and flow rate (In
intermittent operation, if
calculated sampling
time + 1 minute is
exceeded), the error
message is displayed
after the pump stops.
※Not applicable if
sampling is stopped
with "Start/Stop Button"
The flow rate may not reach
the set value due to a large
resistance of the sampling
media.
Check if the flow resistance of
sampling media at the set flow
rate is within the
specifications of the pump.
(Refer to “10. Specifications”)