G-Shield Instruction Manual
Version 1
26
Step 1.
Note the current gas readings.
Step 3.
HCN detected, continue test.
13.3 HCN G-Pod Response Test
Make sure the HCN sensor reads zero before the response test is started.
Turn on the gas. The gas must flow for 3 minutes.
Step 4.
HCN detected gas level after 3 minutes.
WARNING! The response test has failed. You must calibrate the HCN G-Pod. Incorrect
calibration can result in injury or death to personnel.
The HCN G-Pod has passed its response test and can be used. The reading was in the
tolerance of ± 10% for the applied calibration gas.
Examples of G-Pod response outside of tolerance.
Figure 36. HCN response test too high.
Figure 37. HCN response test too low.
Step 2.
Connect HCN 10 PPM calibration gas.