G-Shield Instruction Manual
Version 1
30
13.7 CO G-Pod Zero Calibration
NOTE: Use zero-grade air unless you know the ambient air does not contain CO.
Access the Settings screen. See section
Step 1.
Push the for sensor 2.
Step 2.
Current sensor reading shown.
Step 3.
Connect zero-grade air.
Step 4.
Turn the gas on. Push
Step 5.
Zero calibration started.
Step 6.
Zero calibration completed. Turn gas off.
If an error occurs please consult