Wall-mount type of Carbon Dioxide Detector 

Model: GTF220

 

GASTECH is a professional manufacturer in various gas detectors, carbon monoxide detector, carbon dioxide detector, 
oxygen detector, odor detector, VOC detector, hydrogen sulfide detector and carbon monoxide detector. Welcome tour 
our website 

www.gastech.com.tw 

6

 

In the display, user has to finish calibration. 
The last words of last row might show UP, press ADJ key will increase the 
current output. Press UPDN key, the word will change to DN; press ADJ key 
will decrease the current output. 
Press NEXT key to finish calibration of 20 mA and enter next function. 
 

In the display, user can read out the used life time of detector. Normally, the sensor of 
detector has 1,000 days life time. It is necessary to renew the sensor beyond that. 

Note 1: due to QC procedure, the used life time might be more than 1 day, but it should be 
less than 30 days. 
Note 2: Life time meter count up during power on only. If the detector is idle for a long 
time, or used in an environment with high concentration of carbon dioxide, frequentative 
power on and off, the lifetime will be shortened even the meter less than 1,000 days. 

Press NEXT key; the display will back to the concentration of carbon dioxide. 

3. Alarm message 

The concentration of gas is higher than the buzzer high limit, the detector not only has audible alarm but also 

provide clear alphanumeric alarm message for users. 

If the concentration of carbon dioxide is higher than the buzzer high limit, the detector will 
show this display. If the concentration of carbon dioxide is over 8,000 ppm will show 
>8,000 ppm. 
Press NEXT key, the buzzer will mute and the display will back to the concentration of 
carbon dioxide. If the concentration of carbon dioxide is still higher than buzzer high limit, 
this message and audible alarm will come again after 30 seconds that will disappear until 
the concentration of carbon dioxide is lower than the buzzer high limit. 
 
If the concentration of carbon dioxide is lower than the buzzer low limit, the detector will 
show this display. 
Press NEXT key, the buzzer will mute and the display will back to the concentration of 
carbon dioxide. If the concentration of carbon dioxide is still lower than buzzer low limit, 
this message and audible alarm will come again, that will disappear until the concentration 
of carbon dioxide is higher than the buzzer low limit. 
 

4. Warning message 

The GTF220 carbon dioxide detector may auto-off under low or high specified voltage. 

If the gas detector is 

auto-off, it cannot work for detection. 

At some of case, the detector cannot work but will provide warning message 

which reminds user to check the detector or its system. 

 
Due to input voltage lower than specified, the detector cannot work. 
 
 
 
Due to ambience temperature higher than specified, the detector cannot work. 
 

 

The copyrights reserved by GASTECH. 

GASTECH reserves the right to improve any product described in this operation manual without prior notice. 

High Alarm 

CO2  1368 ppm

 

Life Time 

29 DAYS 

Temp. High 

Calibration 

20 mA    UP 

Power Low

 

Low Alarm 

CO2  600 ppm

 

Summary of Contents for GTF220

Page 1: ...Security Notifications in detail firstly 2 The gas detector might install on wall or steel plate please refer following drawing to make marking then drill the holes Do not use the gas detector as a drilling template to avoid fault 3 Due to the space is very small that is impossible to use wrench Therefore to use female expansion bolts on the wall or tap female screws on the steel plate and use ext...

Page 2: ...astech com tw 2 4 Analog output A recommend fuse is 0 2 A in the above 5 Relay output A recommend fuse is 0 2 A the diode could be 1N4002 or 1N4004 in the above 6 In order to meet waterproof the wiring material and parts should be met the requirement also 7 The wiring size is not less 0 5 mm 2 or 20 AWG In the case of long distance wiring to apply bigger size or 30 VDC power is better 8 The detect...

Page 3: ...ons Specification maybe changed without notice model GTF220 A GTF220 R type wall mount sensor solid electrolyte target carbon dioxide sampling method diffusion measuring range 350 8 000ppm resolution accuracy sampling time 1 ppm 5 F S 350 5 000ppm continuity alarm level high alarm resetable analog output 4 20 mA max load resistance 500 ohm N A relay output rating N A single pole normal open relay ...

Page 4: ...s The following operating procedure describes the settings In the operation press NEXT key or idle 3 minutes the display will back to the concentration of carbon dioxide In the display press NEXT key to enter next function In the display press ADJ key the buzzer and flash lamp will action Press ADJ key again they will stop The relay will make action in TEST ALARM also if detector model has relay o...

Page 5: ...and ADJ key at the same time the detector will execute calibration function The calibration will take 10 minutes in the period time here the detectors will access the lowest detecting value to make a basis of 350 ppm of carbon dioxide In order to get good basis and to avoid the breath affect consequence the operators have to leave and come back again after 10 minutes It is necessary to provide a f...

Page 6: ... carbon dioxide is higher than the buzzer high limit the detector will show this display If the concentration of carbon dioxide is over 8 000 ppm will show 8 000 ppm Press NEXT key the buzzer will mute and the display will back to the concentration of carbon dioxide If the concentration of carbon dioxide is still higher than buzzer high limit this message and audible alarm will come again after 30...

Reviews: