GPD-100 Instruction Manual
PAGE 16 of 27
Rev 1.1(2019.05.23)
SPAN SPAN Calibration Mode
YES/no Span calibration setting
50.0
Span calibration value
50.3
Measured value of
standard gas
WAIt
Under span calibration
50.0
Measured value after
completing calibration
ALAm
Alarm
setting
LACH Alarm occurrence setting on/OFF
AL-1
Alarm1 value setting
90% setting of 0~Full Scale
1H/1L
Alarm1 operational
direction
H:Upward Alarm
L: Downward Alarm
AL-2
Alarm2 value setting
90% setting of 0~Full Scale
2H/2L
Alarm2 operational
direction
H:Upward Alarm
L: Downward Alarm
LOg
Saving log
CtrL
Log saving mode
On/OFF
AUto
Set log auto saving mode On/OFF
mSdC
Set type of external
memory
[Table 7. Menu Table]
6.3.
Configuration Mode
While the device displays ”Stby”(Stand by), press and hold “RUN/MODE” key
more than 2 seconds. Then you will enter menu selection mode.
Password entering screen shows up. Input password using “
”Key or
“
”Key. If you press “RUN/MODE” key after entering password, then you
enter a configuration menu (“ConF”) (Initial password: [--]).
Select “COnF”(configuration) using “
” key or “
” key and then press
“RUN/MODE” key. You will enter environment configuration mode.
“Add” message will show up. In this state, you can set an address that is
used to identify the device when communicating data.
If you press ”RUN/MODE” key, then you enter an address setting function.
Address is a mode in which you can set the device’s address so as to
identify each detector by PC uniquely. Address number(initially [01]) will
increase or decrease by pressing “
” or “
” key.
When desired address is displayed, press “RUN/MODE” key. Then the
address is set to be the number displayed in the screen. After that, the