www.gastron.com
16
_
17
[Figure 10. Span Calibration related Parts]
■ Ex.) Output Calculation Method for NH3
Range : 0-150ppm
Calibration gas : 100ppm NH
3
/N
2
balance
Output signal : 4-20mA
Test point signal ( TP5, TP6 )
■ Ex.) Output Calculation Method for CO
Range : 0-150ppm
Standard Gas : 100ppm
Output signal : 4-20mA
Test point signal (TP5, TP4)
(200 - 40) ×
100ppm NH
3
(Standard Gas)
+ 40 = 147 = 14.7mA
150ppm NH
3
(Measurement Range)
(200 - 40) ×
100ppm CO (Standard Gas)
+ 40 = 147 = 14.7mA
150ppm CO(Measuring Range)
6. Calibration and Maintenance