124
PIPS II Owners Manual and Users Guide
7.
Tighten the set screws to the timing pulley.
8.
Plug the motor into the motor drive cable.
9.
Install the cover.
10.
Turn on power to the PIPS II.
Figure 5-14 Motor Drive Removal.
5.13.
Replacing the Stage Encoder
The stage encoder is located under the specimen chamber. It does not need to
be replaced unless it fails, which typically does not happen during the lifetime
of the instrument.
The stage encoder can be replaced without removing the stage assembly from
the chamber, but it may be easier to remove the stage assembly first and then
change the encoder on a bench.
1.
Remove the cover.
2.
Shut down power to the PIPS II.
Unplug the power cable from the back
of the system.
3.
Remove the cover.