4.5 REPLACING THE IV ARM SKIN (S222.100)
1.
Gently lift the arm skin that meets at
the seam of Chloe's upper arm.
S222 & S222.250 & S222.100
|
USER GUIDE
Working with Chloe | 35
Draining the Venous System with the Fill Kit
If the venous system was filled using the fill kit syringe, follow the steps below to drain the system.
1.
Fill the syringe with a 70:30 mix
of water to isopropyl alcohol and
connect to the shoulder port.
2.
Make sure the drainage tube is open and
push the fluid through the arm until the
fluid coming out of the arm is clear
3.
Inject air through the arm to dry the
vasculature and prevent mold from forming.
NOTE: Drain and flush the venous system with
clean water after each day of simulation.