S222 & S222.250 & S222.100
|
USER GUIDE
Troubleshooting | 75
9.1 OMNI® IS NOT TURNING ON
9.2 CONNECTIVITY TO THE SIMULATOR CANNOT BE
ESTABLISHED OR IS LOST
9. TROUBLESHOOTING
Probable Cause
Solution
Power supply may not be
connected to the simulator
Make sure the power supply cable is connected
Communication cable is not
connected to OMNI® or the
simulator
Connect the communication cable to the simulator and to OMNI®
Probable Cause
Solution
Corrupted OMNI® connection
Reboot OMNI® by unplugging the communication cable and reconnecting
after a few seconds
OMNI®'s data cable is not
connected
Ensure the Ethernet cable is plugged into OMNI® and the simulator
Verify the power cable is connected to the simulator
OMNI®2 tablet is too far away from
the simulator
Minimize the distance between the tablet and the simulator.
The distance should not be more than 12 meters away.
More than one device is paired
Open the Bluetooth connection setup.
Select the menu icon and tap “Forget” for the unwanted connection.
The simulator is not paired
Tap the Menu icon and select "Bluetooth Connection Setup and tap the
serial number. It will appear under "Paired devices."
Power supply is not connected to
the simulator
Ensure that the power supply is connected to the simulator