NOELLE S554.100 |
User Manual
Other | 58
Streaming audio does not work, tab is not displayed
The backup battery on the
simulator is depleted
Plug charger into the simulator. Verify LED light on charger indicates
“charging” status.
•
NOELLE should always be plugged in while being operated. The birth
mechanism is power intensive and will drain the backup battery quickly.
•
Reconnect Newborn to the charger.
UNI has set the power mode to STAND-BY automatically
Wired communication module
is not connected
Connect the communication module to any USB port.
“Wired communication module not found” message is displayed when UNI is started
Wired communication is
module not identified by the
computer
Close the software and try disconnecting the communication module for at
least five seconds, then plug it back in and restart the software
Drivers not properly loaded.
Perform a full shutdown procedure on the tablet.
Is the communication indicator
panel consistantly yellow?
See solution above in section making reference to “Communication with the
simulator cannot be established”
CPR is not detected
Compression sensor is not
calibrated
See Calibration section inside the User Guide.
Ventilations are not detected
See Calibration section inside the User Guide.
Set the Respiratory Rate to 0.
Noelle’s delivery baby is disengaged prematurely
Wrong force sensor reading.
When baby is pulled with more than 35 lbs. of force, to avoid damage, the
baby is released. If user is not pulling the baby it means that “Force Sensor” is
offset. See the “Calibration Wizard” section inside User’s Manual and reset the
force sensor to its default value (“Reset to Default Force” button).
Summary of Contents for NOELLE S554.100
Page 2: ......