NOELLE S554.100 |
User Manual
59 | Other
Noelle’s delivery baby does not release at delivery end
Dystocia is “ON”
When “Dystocia On” button is checked, the baby is not released until check-
ing “Dystocia Off”.
Motor is “disoriented”
Reset the delivery motor by going to the Gear> Setup> Options menu, “NO-
ELLE Features” tab, and then click on “Labor Motor” under “Reset…”. If de-
livery mechanism is at the very beginning in a way that it is compressing the
rubber boots, the grinding noise won’t go away for a minute or so. Should that
be the case either let it go (it won’t break) or simply move motor forward a
few turns, and then reset.
Initial fetus position was not
specified on the software
Lock the baby into the birthing mechanism and manually turn the baby to
either ROA, LOA, LOP, and ROP. Then synchronize the baby’s position in the
womb with the labor position icon located on the labor tab.
Motor is making a loud grind-
ing noise
Click the Gear> Setup> Calibration> Factory Settings. Reset the Force Sensor
and Reset the labor motor.
Pre-built scenarios do cannot be loaded
Wrong profile.
Select “Quick Start Scenarios” when starting the software.
To change profiles from within the UNI. Go to “File/Profile” menu and then
select “Modeled Scenarios”