___________________________________________________Chapter 6 – Testing and Repair, Page 6-11
Replacing UV Lamps – Cont.
9.
Slide off each UV Lamp Sockets from the Lamps at each end of the Module (Refer to Figure 6-11-1).
10.
Lubricate the Lamps around the Grommets on each end of the UV Lamp with a small amount of
Formula G-510 or similar detergent.
11.
CAREFULLY slide the UV Lamp out one end.
12.
Inspect the Grommets around the Lamps (Refer to Figure 6-11-1). Replace any Grommet that show
cracks, or any other damage.
13.
Check all UV Lamp Sockets and wires for damage. Replace Lamp Sockets if the socket or wires show
damage. Refer to Page 6-13 for replacement instructions.
14.
Lubricate each end of the new UV Lamps before installing with a small amount of Formula G-510 or
similar detergent.
15.
From one end, slide in all the new UV Lamps.
16.
Reconnect all the UV Lamp Sockets to the appropriate numbered Lamp as shown in Figure 6-10-2.
17.
Check the Module End Cap Gaskets for cracks or deterioration (Refer to Figure 6-10-1). Replace if
needed.
18.
Re-install the Module End Caps with gasket and torque the bolts/nuts to 7-9 in – lbs.
19.
Check the UV Module Plate Gasket for cracks or deterioration (Refer to Figure 6-12-2). Replace if
needed. Refer to Page 6-41
for instructions on replacing.
20.
Slide the UV Module back into the Ventilator.
21.
The next step is to run a test to ensure all Lamps are operating. Secure the Module by installing one
of the holding nuts at each end.
Caution: Testing the Lamps requires the use or UV protective glasses as a small amount of UV light
may be visible.
22.
Reconnect the Lamp / Ballast Connector plug.
23.
Turn on the power to the Command Center.
24.
Turn on the circuit that supplies power to the UV Lamps.
UV Lamp Socket
Socket Number
Wrapped on
Socket Wire
Lamp Grommets
Figure 6-11-1
UV Lamp Sockets
End of UV Lamp –
Socket Removed