ULTRATHERM 908i
15
gbo Medizintechnik AG 2004
Version 1.6
3
Installation
3.1
Mains connection
The standard therapy unit is to be connected to a mains voltage of 230 V +/- 15 %, 50/60 Hz.
However, a special version is also available for the connection to 115V +/- 15 %, 50/60 Hz. (The
actual value is shown on the rating plate at the rear of the unit).
Before connecting to the supply mains check the conformity of the voltage values of the available
mains and therapy unit. The therapy unit is to be only connected to a properly installed earthend
socket outlet by means of the shielded mains connecting cable included in the basic accessories.
For the 230-V mains we recommend a slow-acting 10-A fuse.
3.2
Switching on
Switch on the power with switch
(1)
on the left of the operating panel. Now, the therapy unit is in
the stand-by mode of operation. The indicators „Dosage"
(15)
and „Treatment time"
(14)
are lit.
Further settings are to be made on the keyboard (cf. figure).
3.3
Mode of operation
Select the mode of operation with the buttons
(3)
,
(4)
and
(5)
in the
„MODE"
field. Pressing
button
(3)
turns the therapy unit to the continuous mode of operation. Different pulsed modes of
operation can be selected by pressing button
(4)
or
(5)
. They differ in pulse width and pulse
repetition frequency (cf. paragraph 6). The chosen mode is shown by the indicator light in the
corresponding button marked by the symbol. The mode of operation can only be selected when
the dosage display
(15)
shows
„0"
.
3.4
Choice of the applicator
The choice of the appropriate applicator is made according to area and volume of the body parts
to be treated and according to preferring the coil or capacitor-field field application.
By pressing button
(6), (7), (8)
or
(9), (10), (11)
in the
"APPLICATOR"
button pad the dosage
(output power) is limited according to the selected applicator so that an unintended overheating
of the tissues is avoided.