EMERGENCY BYPASS
NOTE:
There are 2 bypasses per system, one on the formulation unit and one on the water processing unit. Both units must be put
into bypass to prevent flooding and damage to equipment. Always put formulation unit into bypass first to prevent flooding.
Putting the system into bypass
1. Turn OFF all the switches on the water storage unit.
2. Turn the yellow handle located on the front panel of the water storage unit towards “ON”. (1 in “Formulation System Diagram”
on page 4)
3. Turn the yellow handle located on the water processing unit towards “ON”. Location will be in the unit guide that came with
the water processing unit.
Note: Tank will not refill while in bypass. System pressure will read zero psi.
Removing the system from bypass
1. Turn the yellow handle located on the water processing unit towards “OFF”. Location will be in the unit guide that came with
the water processing unit.
2. Turn the yellow handle located on the front panel of the water storage unit towards “OFF”. (1 in “Formulation System Diagram”
on page 4)
3. Turn ON all the switches on the water storage unit.
Bypass ON
Bypass OFF
PAGE 12