3/52
EN
OPERATING CONDITIONS
STORAGE, TRANSPORT
CONDITIONS
-20/+60 °C
-30/+70 °C
10/100%
20/70%
600/1200 mbar
600/1200 mbar
Before first use the product must be kept in its original packaging to
ensure that the product is not contaminated. For transport and storage
GCE recommends use of the original packaging (including internal sealing
bag and caps).
Statutory laws, rules and regulations for medical gases, accident prevention
and environmental protection must be observed.
4. PERSONNEL INSTRUCTIONS AND TRAINING
According to Medical Devices Directive 93/42/EEC the product provider
must ensure that all personnel handling the product are provided with the
operating instructions & performance data and are fully trained to carry out
that operation. Trainees need to be supervised by an experienced person.
Do not use the device without being trained. Training can be only done
by a person with an appropriate education, experience and knowledge
that has been also trained by the manufacturer.
For further information about training programs, please contact GCE.
5. PRODUCT DESCRIPTION
Terminal units consist of two basic parts:
•
housing
•
quick connection valve
Individual parts must always be designed for the identical type of gas. Each
of these basic parts is marked with stamping displaying the gas for which
it is intended.
An integral part of each terminal unit is a check valve enabling the
through-pass of the gas when a counterpart is connected and which
automatically prevents the gas from passing through when the counterpart
is disconnected.