GASMAX/TX Operation & Maintenance Manual, Revision 1.0
Page 43
DATE
adjusts date in month, day and two-digit year.
WARM UP DELAY
sets the amount of time from power-on until the 4-20mA output signal begins to track
the sensor output. Default setting is 60 seconds. Maximum value is 255 seconds.
CAL PURGE DELAY
sets the amount of time from the completion of span calibration until the 4-20mA
output signal restarts tracking the sensor output. The default setting is 60 seconds and the maximum
value is 255 seconds.
Figure 11-11: Security Menu
CONTACT NAME
is a 12 character ASCII field available for displaying a phone # or name of personal who
knows the Pass Code. Lost Pass Codes may be recovered by entering the locked security menu and
holding the UP key for 5 seconds. The 4-digit code appears near the bottom of the screen.
SECURE LEVEL
sets LOW, HIGH or OFF modes. A LOW level allows CAL MODE sensor calibrations but
requires the 4-digit Pass Code prior to altering menus. HIGH level locks the entire menu database and the
CAL Mode until the correct Pass Code is entered. LOW and HIGH security levels always allow static
viewing of configuration menus.
Figure 11-12: LCD Contrast Menu
LCD CONTRAST
Select the LCD contrast command and use the UP or DOWN to adjust the contrast of the
LCD display.
SYSTEM SECURITY
SYSTEM SECURITY
SYSTEM SECURITY
SYSTEM SECURITY
Contact Name
Secure Level LOW
Pass Code
xxxx
UNLOCKED
DEVICE SETUP
DEVICE SETUP
DEVICE SETUP
DEVICE SETUP
RF Link Setup
Clock/Delays
Security
LCD Contrast
Turn Power OFF
Reset Range Icon
CONTRAST
CONTRAST
CONTRAST
CONTRAST
UP/ DOWN to change
Next to Exit
DEVICE SETUP
DEVICE SETUP
DEVICE SETUP
DEVICE SETUP
RF Link Setup
Clock/Delays
Security
LCD Contrast
Turn Power OFF
Reset Range Icon