![background image](http://html.mh-extra.com/html/ge/001c-051d-003a/001c-051d-003a_service-manual_128510131.webp)
Troubleshooting:
Review errors
2000966-542D
Dash 3000/4000/5000
6-15
NOTE
The patient monitor may be referred to as a display or scope in the
error code descriptions.
Severity
Severity is a measure of how the event/error affected the system. There
are three levels of severity.
Continue—the event/error was logged, the task may have or may not
have been finished, but the system was able to go on. Most log entries
will have a severity of Continue
.
Fatal—the event/error was such that the task is not able to go on.
Recovery was not possible. This always is followed by a WARM
START.
Forced Restart—the system was restarted by a known condition
(internet address change, video test, etc).
Date and time
The date and time the event or problem occurred.
Error number
A sequential number that is used to identify each event/problem.
70F
Date was changed from this patient monitor. This helps determine
how the system-wide date was altered.
710
Incompatible software was detected. If the main processor software
finds that the software operating on the communication software
incompatible, it turns off the communication (network) controller and
enters this data into the error log. When the patient monitor won’t
“talk” to the network, looking for this entry in the error log is one part of
the troubleshooting process.
Definition of error codes
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......