October 2007
4-10
Programming the OptiSonde
Automatic Cleaning and
Balance Function
OptiSonde analyzers can run the
Auto
or
PACER
self-cleaning and
rebalancing cycles at a preset time after the last cycle was run. The
automatic balance cycle will always run upon power up of the
analyzers.To program the
Automatic Cleaning & Balance Function
on the OptiSonde:
1.
From the main screen:
a.
Press the
ESC
key to enter the Main Menu.
b.
Scroll to
Settings
and press
ENT
.
c.
Enter the
Balance
option.
2.
Set the Balance
Type
:
a.
Enter
Type
.
b.
Enter either
AUTO
or
PACER
. Table 4-7 below explains the
difference between the options.
Note:
The
PACER
function will provide more thorough cleaning than
the
AUTO
function because it first develops a thick dew/frost
layer into which soluble contaminants dissolve. When heated,
some of the contaminants are flash evaporated and the
remaining residue accumulates in clusters, resulting in the
cleaning of much of the mirror’s surface. The
PACER
cycle
generally takes longer to complete.
3.
Select whether or not you will use
Auto Balance
.
a.
From the
Balance
menu, enter
Status
.
b.
Enter
Off
or
On
Table 4-7: Automatic Cleaning and Balance Functions
Item
Function
AUTO
Heats the mirror and balances the optics.
PACER
First cools the mirror to develop a thick dew/frost layer,
then heats and balances the optics.
Summary of Contents for 1111H
Page 1: ...GE Sensing OptiSonde General Eastern Chilled Mirror Hygrometer User s Manual ...
Page 9: ...Chapter 1 ...
Page 15: ...Chapter 2 ...
Page 33: ...Chapter 3 ...
Page 43: ...Chapter 4 ...
Page 57: ...Chapter 5 ...
Page 68: ...Appendix A ...
Page 73: ...Appendix B ...
Page 74: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 78: ...Appendix C ...
Page 79: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 82: ...Appendix D ...
Page 87: ...Appendix E ...
Page 88: ...Theory of Operation and Glossary Theory of Operation E 1 The PACER Cycle E 4 Glossary E 5 ...