Maintenance
5-9
October 2007
“Balance” Remains on the
Status Line
Check that the sensor and sensor cable are connected. If necessary,
connect them, and the unit will complete the PACER cycle after a
short time (5 to 15 minutes).
The sensor bridge may be out of balance (refer to page 5-1).
No Analog Output
If there is no analog output, but the digital display indicates correctly,
check the analog output scaling.
No Serial Output
Check the serial port programming for proper settings.