background image

GE H

EALTHCARE

 PROPRIETARY 

TO 

GE

D

IRECTION

 5344303-100, R

EVISION

 3 

VIVID P3 S

ERVICE

 M

ANUAL

Chapter 8 Replacement Procedures

8-69

Section 8-7  RX Board (FRU P/N: 5315027)

8-7-0-1

Tools

C

ommon Phillips screwdrivers

8-7-0-2

FRU BOM

This FRU Kit P/N:5315027 consists of following parts.

8-7-0-3

Needed Manpower

1 person, 30 minutes

8-7-0-4

Preparations

Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.

8-7-0-5

Removal Procedure

1) Remove the connector board assembly. Refer 

section 8-6-2 on page 63

2) Loosen the Five screws (1,2,3,4,5,) of TX board from the system. Refer 

Figure 8-59 on page 8-67 

3) Remove the Tx board from the system. Refer 

Figure 8-59 on page 8-67 

Figure 8-60   TX board Assembly

4) Remove the Tx conn board cable. Refer 

Figure 8-61 on page 8-70 

Part Number

Part Description

2404906

Receive Board

5308959

RX mounting Bracket

2139752

Fasteners- STR.SCR M3 X 8+BUILT IN SPR+PLNWAS

1

2

3

4

5

Summary of Contents for 5350375

Page 1: ...Technical Publication Direction 5344303 100 Revision 3 GE Healthcare Vivid P3 Service Manual Copyright 2009 2011 by General Electric Company Inc All Rights Reserved GE Healthcare ...

Page 2: ...This page was intentionally left blank ...

Page 3: ...NGUE QUE L ANGLAIS IL INCOMBE AU CLIENT DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL DE MAINTENANCE N A PAS ÉTÉ CONSULTÉ ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN L OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES MÉCANIQUES OU AUTRES DIESES KUNDENDIENST HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRAC...

Page 4: ...AVISO PODE PÔR EM PERIGO A SEGURANÇA DO TÉCNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS MECÂNICOS OU OUTROS IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN INGLESE SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE SI PROCEDA ALLA MANUTENZIONE DELL APPARECCHIATURA SOLO DOPO ...

Page 5: ...ΡΟΝ ΕΓΧΕΙΡΙ ΙΟ ΣΕΡΒΙΣ ΕΑΝ Ε ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙ ΟΠΟΙΗΣΗ ΑΥΤΗ ΕΝ ΕΧΕΤΑΙ ΝΑ ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΣΤΟ ΧΕΙΡΙΣΤΗ Ή ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ ΥΝΟΥΣ EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE NE PRÓBÁLJA ELKEZDENI HASZ...

Page 6: ...ANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE PÅ GRUND AF ELEKTRISK MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN OPERATØREN ELLER PATIENTEN DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST DAN IS DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE ONDERHOUDSHANDLEIDING WER...

Page 7: ... MEKANISKE ELLER ANDRE FARER NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU ANGIELSKIM JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA KLIENCIE NIE PRÓBOWAĆ SERWISOWAĆ NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM NIEZASTOSOWANIE SI...

Page 8: ...E TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA POSKYTOVATEĽA SLUŽIEB OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA ELEKTRICKÝM PRÚDOM PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO NEBEZPEČENSTVA DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA ÖVERSÄTTNINGSTJÄNSTER FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNI...

Page 9: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL vii ...

Page 10: ...lectrical contractors Other connections between pieces of electrical equipment calibrations and testing shall be performed by qualified GE Healthcare personnel In performing all electrical work on these products GE will use its own specially trained field engineers All of GE s electrical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE...

Page 11: ... or duplicated in any form in whole or in part without prior written permission of GE Healthcare GE Healthcare may revise this publication from time to time without written notice TRADEMARKS All products and their name brands are trademarks of their respective holders COPYRIGHTS All Material Copyright 2009 2010 by General Electric Inc All Rights Reserved ...

Page 12: ...3 29 3 Chapter 8 Replacement Procedures pages 8 1 to 8 83 3 Important Precautions pages i to x 3 Chapter 4 Functional Checks pages 4 1 to 4 43 3 Chapter 9 Replacement Parts pages 9 1 to 9 19 3 Table of Contents pages xi to xxviii 3 Chapter 5 Theory pages 5 1 to 5 47 3 Chapter 10 Periodic Maintenance pages 10 1 to 10 25 3 Chapter 1 Introduction pages 1 1 to 1 18 3 Chapter 6 Service Adjustments page...

Page 13: ... 1 3 Conventions Used in Book 1 3 Standard Hazard Icons 1 4 Product Icons 1 5 Safety Considerations 1 9 Introduction 1 9 Human Safety 1 9 Mechanical Safety 1 9 Electrical Safety 1 10 Labels Locations 1 11 Warning labels location Tables 1 16 Dangerous Procedure Warnings 1 17 Lockout Tagout Requirements For USA Only 1 17 Returning Shipping Probes and Repair Parts 1 17 EMC EMI and ESD 1 18 Electromag...

Page 14: ...te Power Outlets 2 4 Unit Power Plug 2 4 Power Stability Requirements 2 4 EMI Limitations 2 5 Scan Probe Environmental Requirements 2 6 Time and Manpower Requirements 2 6 Facility Needs 2 7 Recommended Ultrasound Room Layout 2 7 Purchaser Responsibilities 2 7 Required Features 2 8 Desirable Features 2 8 Recommended and Alternate Ultrasound Room Layout 2 9 Networking Pre installation Requirements 2...

Page 15: ... 3 9 Checking the Components 3 9 Completing the Installation 3 10 Power On Boot Up 3 10 Scanner Power On 3 10 Turn on the system 3 11 Power Off Shutdown 3 11 TMST Power Down 3 11 Scanner Shutdown 3 12 Transducer Connection 3 13 System Configuration 3 14 System Specifications 3 14 Physical Dimensions 3 14 Approved peripherals 3 15 Connecting Cables 3 16 Peripherals Accessories Connector Panel 3 16 ...

Page 16: ...GE HEALTHCARE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 4 Table of Contents Paperwork 3 29 Product Locator Installation 3 29 User Manual s 3 29 ...

Page 17: ... 6 System Features 4 7 Control Panel 4 7 Vivid P3 SoftMenu Key Tour 4 8 Monitor Display 4 9 B Mode Checks 4 10 Preparations 4 10 B Mode OP Panel Controls 4 11 B Mode Softmenu Key 4 12 M Mode Controls 4 13 Preparations 4 13 M Mode OP Panel Controls 4 14 M Mode Softmenu Key 4 15 Color Flow Mode Checks 4 16 Preparations 4 16 Color Flow Mode OP Panel Controls 4 17 Color Flow Mode Softemenu Key 4 18 Do...

Page 18: ...obe 4 31 Using Cine 4 31 Activating CINE 4 31 Quickly Move to Start End Frame 4 31 Start Frame End Frame 4 31 Adjusting the CINE Loop Playback Speed 4 32 Moving through a CINE Loop Frame By Frame 4 32 Image Management QG 4 32 Backup and Restore Database Preset Configurations and Images 4 33 Formatting Media 4 33 Backup System Presets and Configurations 4 34 Patient Healthcare Information PHI 4 34 ...

Page 19: ...ower Down sequence description 5 9 PWA Assy Diagrams 5 10 TMST 5 10 Description 5 12 RX64 5 13 Description 5 13 TX64 5 14 Description 5 14 CWD 5 15 Description 5 15 Connector Board 5 16 Rear Panel 1 Rear panel 2 5 17 Keyboard 5 19 Indicator LEDs 5 20 Power Diagrams 5 21 Overview 5 21 Common Service Platform 5 22 Introduction 5 22 Global Service User Interface GSUI 5 22 Internationalization 5 22 Se...

Page 20: ... 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 8 Table of Contents CHAPTER 6 Service Adjustments Overview 6 1 Purpose of this chapter 6 6 1 Monitor Adjustments 6 2 Adjustments Procedures 6 2 Monitor Calibration procedure 6 2 ...

Page 21: ... the P3 Key Function 7 6 Setting the P3 Key to Screen Capture 7 6 Capturing a Screen 7 6 Reset the P3 Key to Customer s Functionality 7 8 Wire LAN Network 7 9 Troubleshooting 7 10 Console Troubleshooting Trees 7 10 System Doesn t Boot 7 10 B Mode Low Sensitivity 7 12 B Mode Low Image Quality 7 13 Noise in B Mode 7 14 Color Flow Low Sensitivity 7 15 Spectrum Doppler Low Sensitivity 7 16 Noise in Co...

Page 22: ...7 LCD Arm Cover FRU P N 5310093 8 8 Tools 8 8 Needed Manpower 8 8 Preparations 8 8 Removal procedure 8 8 Mounting procedure 8 8 Functional Checkout Procedure 8 8 Handle P3 Console FRU P N 5315120 8 9 Tools 8 9 FRU BOM 8 9 Needed Manpower 8 9 Preparations 8 9 Removal procedure 8 9 Mounting procedure 8 10 Functional Checkout Procedure 8 10 Keyboard Assembly FRU P N 5340674 8 11 Tools 8 11 Needed Man...

Page 23: ...unctional Checkout Procedure 8 20 Keyboard Rear Cover Assy FRU P N 5315122 8 21 Tools 8 21 FRU BOM 8 21 Needed Manpower 8 21 Preparations 8 21 Removal procedure 8 21 Mounting procedure 8 22 Functional Checkout Procedure 8 22 Probe Holder Gel Bottle Holder 8 23 Needed Manpower 8 23 Preparations 8 23 Removal procedure 8 23 Mounting procedure 8 23 Functional Checkout Procedure 8 23 Keyboard Knob Set ...

Page 24: ... 30 Mounting procedure 8 31 Functional Checkout Procedure 8 32 Front Cover FRU No 5315116 8 33 Tools 8 33 FRU BOM 8 33 Needed Manpower 8 33 Preparations 8 33 Removal Procedure 8 33 Mounting procedure 8 34 Functional Checkout Procedure 8 35 Top Cover FRU P N 5315121 8 36 Tools 8 36 FRU BOM 8 36 Needed Manpower 8 36 Preparations 8 36 Removal Procedure 8 36 Mounting procedure 8 38 Functional Checkout...

Page 25: ...rocedure 8 46 Mounting procedure 8 47 Functional Checkout Procedure 8 48 MST BOARD ASSEMBLY 8 49 Tools 8 49 FRU BOM 8 49 Needed Manpower 8 49 Preparations 8 49 Removal Procedure 8 49 Mounting procedure 8 52 Functional Checkout Procedure 8 52 HARD DISK DRIVE ASSEMBLY PART NO 5315113 8 53 Tools 8 53 FRU BOM This FRU Kit P N 5315113 consists of following parts 8 53 Needed Manpower 8 53 Preparations 8...

Page 26: ...ls 8 63 Needed Manpower 8 63 Preparations 8 63 Removal Procedure 8 63 Mounting procedure 8 65 Functional Checkout Procedure 8 66 TX Board Flex Cable Con to Tx Assy 8 67 Tools 8 67 FRU BOM 8 67 Needed Manpower 8 67 Preparations 8 67 Removal Procedure 8 67 Mounting procedure 8 68 Functional Checkout Procedure 8 68 RX Board FRU P N 5315027 8 69 Tools 8 69 FRU BOM 8 69 Needed Manpower 8 69 Preparation...

Page 27: ...ations 8 80 Removal Procedure 8 80 Mounting procedure 8 80 Functional Checkout Procedure 8 81 Signal Cable Set 8 82 Tools 8 82 FRU BOM 8 82 Needed Manpower 8 82 Preparations 8 82 Removal Procedure 8 82 Functional Checkout Procedure 8 82 Power Cable Set FRU P N 5315036 8 83 Tools 8 83 FRU BOM 8 83 Needed Manpower 8 83 Preparations 8 83 Removal Procedure 8 83 Functional Checkout Procedure 8 83 CWD A...

Page 28: ...verview 9 1 Purpose of Chapter 9 9 1 List of Abbreviations 9 1 Operator Console Assy 9 2 LCD Assy 9 3 Keyboard Assy 9 4 Mechanical Assy 9 6 Pwa Assy 9 10 Flex Pcb Cables 9 12 Signal Cable set Power Cable set Fru USB Cables Fru Part num 9 13 Accessories and Kits 9 16 Probe 9 17 Vivid P3 Fru List With Part Numbers 9 18 ...

Page 29: ...minary Checks 10 7 Functional Checks 10 8 System Checks 10 8 Peripheral Option Checks 10 9 Input Power 10 9 AC Power Inspection 10 9 Cleaning 10 9 General Cleaning 10 9 Physical Inspection 10 10 Optional Diagnostic Checks 10 11 View the Logs 10 11 Probe Maintenance 10 11 Probe Related Checks 10 11 Basic Probe Care 10 11 Basic Probe Cleaning 10 11 Using a Phantom 10 12 Electrical Safety Tests 10 12...

Page 30: ... Source Leakage Lead to Lead 10 22 Dale 600 Patient Lead Tests 10 22 Isolated Patient Lead Sink Leakage Isolation Test 10 22 Data Sheet for ECG Leakage Current 10 23 Probe Leakage Current Test 10 24 Definition 10 24 Test Equipment 10 24 Generic Procedure for Leakage current 10 26 Meter Procedure Using Dale Meter to Measure Leakage Current 10 26 Data Sheet for Transducer Source Leakage Current 10 2...

Page 31: ...tional Checks Contains functional checks that are recommended as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Components and Functions Theory Contains block diagrams and functional explanations of the electronics 6 Chapter 6 Service Adjustments Contains instructions on how to make available adjustments to the Vivid P3 7 Chapter 7 Diagnostics Trouble...

Page 32: ...e Installation 1 1 5 Vivid P3 Models Covered by this Manual 1 1 6 Purpose of Operator Manual s The Operator Manual s should be fully read and understood before operating the Vivid P3 and also kept near the unit for quick reference Table 1 2 Vivid P3 Model Designations Part Number Description 5350375 Vivid P3 System INDIA 230V 5350378 Vivid P3 System CHINA 230V 5350379 Vivid P3 System CHILE KOREA 2...

Page 33: ... potential hazards are labeled in one of following ways NOTE Notes provide important information about an item or a procedure Information contained in a NOTE can often save you time or effort DANGER DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE IGNORED WARNING WARNING WARNING IS USED TO INDICATE THE PRESENCE OF A HAZARD ...

Page 34: ...pter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that should be followed Table 1 3 Standard Hazard Icons ELECTRICAL MECHANICAL RADIATION LASER HEAT Table 1 4 Standard Icons Indicating a Special Procedure to be Used AVOID STATIC ELECTRICITY TAG ...

Page 35: ...EC 878 02 03 indicates B Type equipment having a floating applied part Probe and PCG marked Type BF ATTENTION Consult accompanying documents is intended to alert the user to refer to the operator manual or other instructions when complete information cannot be provided on the label Various CAUTION Dangerous voltage the lightning flash with arrowhead in equilateral triangle is used to indicate elec...

Page 36: ... current limits An example of a high risk patient would be a special procedure where the patient has an accessible conductive path to the heart such as exposed cardiac pacing leads Rear of console Alternating Current symbol is in accordance with IEC 60878 01 14 Rear Panel Rating Plate Circuit breaker label of console and front panel if applicable Date of manufacture The date could be a year year a...

Page 37: ...centration values for electronic information products as set by the People s Republic of China Electronic Industry Standard SJ T11364 2006 include the hazardous substances of lead mercury hexavalent chromium cadmium polybrominated biphenyl PBB and polybrominated diphenyl ether PBDE 20 indicates the number of years during which the hazardous substance s will not leak or mutate so that the use of th...

Page 38: ...incline due to the weight of the unit On the console where easily seen during transport This product consists of devices that may contain mercury which must be recycled or disposed of in accordance with local state or country laws Within this sytem the backlight lamps in the monitor display contain mercury Consult accompanying document is intended to alert the user to refer to the operator manual ...

Page 39: ...ansporting place the system in its special storage case Ensure that the system is firmly secured while inside the vehicle Secure system with straps or as directed otherwise to prevent motion during transport Prevent vibration damage by driving cautiously Avoid unpaved roads excessive speeds and erratic stops or starts WARNING WARNING Ultrasound probes are highly sensitive medical instruments that ...

Page 40: ...equipment chassis must be connected to an electrical ground The system is equipped with a three conductor AC power cable This must be plugged into an approved electrical outlet with protective ground The power outlet used for this equipment should not be shared with other types of equipment Both the system power cable and the power connector meet international electrical standards ...

Page 41: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 1 Introduction 1 11 1 3 5 Labels Locations Figure 1 1 Label Location for Vivid P3 ...

Page 42: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 1 12 Section 1 3 Safety Considerations 1 3 5 Labels Locations cont d Figure 1 2 Label Location for Vivid P3 ...

Page 43: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 1 Introduction 1 13 1 3 5 Labels Locations cont d Figure 1 3 Label Location for Vivid P3 ...

Page 44: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 1 14 Section 1 3 Safety Considerations 1 3 5 Labels Locations cont d Figure 1 4 ...

Page 45: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 1 Introduction 1 15 1 3 5 Labels Locations cont d Figure 1 5 ...

Page 46: ...near this equipment D8 Caution Federal Law restricts this device to sale by or on the order of a physician D9 The CE Mark of Conformity indicates this equipment conforms with the Council Directive 93 42 EEC D10 CISPR CAUTION The Vivid P3 conforms to the CISPR11 Group 1 Class A of the international standard for Electromagnetic disturbance characteristics D11 WEEE Label This symbol indicates that th...

Page 47: ... purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or open this package NOTE The US Department of Transportation DOT has ruled that items that were saturated and or dripping with human blood that are now caked with dried blood or which were used or intended for use in patient care are regulated medical waste for transportation p...

Page 48: ...power of signal cables In addition to electromagnetic energy EMC also includes possible effects from electrical fields magnetic fields electrostatic discharge and disturbances in the electrical power supply 1 4 2 CE Compliance The Vivid P3 unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge radiated and conducted RF fields magnetic fi...

Page 49: ... Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa WI 53226 Service On site Service Parts Applications support 1 800 437 1171 1 800 558 2040 1 800 682 5327 or 1 262 524 5698 Canada 1 800 668 0732 Latin America Service Applications support 1 800 321 7937 1 262 524 5698 Europe GE Ultraschall Deutschland GmbH Co KG BeethovenstraBe 239 Postfach 11 05 60 D 42665 Solingen Germany Phone 33 0...

Page 50: ...N 3 VIVID P3 SERVICE MANUAL 1 20 Section 1 5 Customer Assistance 1 5 2 System Manufacturer Table 1 7 System Manufacturer Manufacturer Wipro GE Healthcare Pvt Ltd No 4 Kadugodi Industrial Area Bangalore Karnataka INDIA 560067 TEL 91 80 41801000 FAX 91 80 28452924 ...

Page 51: ...f this chapter 2 This chapter provides the information required to plan and prepare for the installation of a Vivid P3 Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit 2 1 2 Chapter Contents Table 2 1 Contents in Chapter 2 Section Description Page Number 2 1 Overview 2 1 2 2 General Console Requirements 2 2 2 3 Facility Needs 2 7 ...

Page 52: ...dim bright is recommended Keep in mind that lighting controls and diameters can be a source of EMI which could degrade image quality These controls should be selected to minimize possible interface Table 2 2 Environmental Requirements for Vivid P3 Scanners Operational Storage Transport Temperature 10 40 degree C 50 104 degree F 10 70 degree C 14 158 degree F 40 60 degree C 40 140 degree F Humidity...

Page 53: ... of one phase two lines not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Please note that image artifacts can occur if at any time within the facility the ground from the main facility s incoming power source to the Ultrasound unit is only a conduit 2 2 2 1 Vivid P3 Power Requirements Table 2 4 Electrical Specifications for Vivid P3...

Page 54: ...s Other adequate outlets for the external peripherals medical and test equipment needed to support this unit must also be present within 1 m 3 2 ft of the unit Electrical installation must meet all current local state and national electrical codes 2 2 2 5 Unit Power Plug If the unit arrives without a power plug or with the wrong plug you must contact your GE dealer or the installation engineer mus...

Page 55: ... most likely reason a unit will have noisy images Check grounding of the power cord and power outlet Replace all screws RF gaskets covers cores After you finish repairing or updating the system replace all covers and tighten all screws Any cable with an external connection requires a magnet wrap at each end Install the shield over the front of card cage Loose or missing covers or RF gaskets allow ...

Page 56: ...kes time Begin Pre installation checks as soon as possible If possible allow six weeks before delivery for enough time to make necessary changes NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF 20 TO 50 degrees C WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS THE PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE CAUTION Have two people available to deliv...

Page 57: ...d electrical contractors Other connections between pieces of electrical equipment products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All electrical work on these product must comply with the requirements of applicable electrical codes The purchaser of GE equipment must only utilize qualified personnel to perform e...

Page 58: ...ltrasound unit is only a conduit Dedicated single branch power outlet of adequate amperage meeting all local and national codes which is located less than 2 5 m 8 ft from the unit s proposed location Door opening is at least 76 cm 30 in wide Proposed location for unit is at least 0 3m 1 ft from the wall for cooling Power outlet and place for any external peripheral are within 2 m 6 5 ft of each ot...

Page 59: ...00 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 2 Pre Installation 2 9 2 3 3 1 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment Figure 2 6 RECOMMENDED ULTRASOUND ROOM LAYOUT ...

Page 60: ...manner frees up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 4 4 DICOM Option Pre installation Requirements To configure the Vivid P3 to work with other network connections the site s network administrator must provide some necessary information Information must...

Page 61: ...ts cont d Figure 2 7 Worksheet for DICOM Network Information Vivid P3 Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses Default DICOM APPLICATION INFORMATION NAME Store 2 Store 3 Store 1 MAKE REVISION IP ADDRESSES PORT AE TITLE Store 5 Store 6 Store 4 Storage Commit MPPS Worklist ...

Page 62: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 2 12 Section 2 3 Facility Needs This page was intentionally left blank ...

Page 63: ...on Time The VIVID P3 installation and functional checkout will take approximately Four hour VIVID P3 consoles with optional equipment may take slightly longer Table 3 1 Contents in Chapter 3 Section Description Page Number 3 1 Overview 3 1 3 2 Receiving and Unpacking the Equipment 3 3 3 3 Packing the Equipment 3 8 3 4 Preparing for Installation 3 9 3 5 Completing the Installation 3 10 3 6 System C...

Page 64: ...ay cause the system to be damaged Table 3 3 Time for Settlement C 60 55 50 45 40 35 30 25 20 15 10 5 0 5 10 15 20 25 30 35 40 F 140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 4 13 22 31 40 hrs 8 6 4 2 0 0 0 0 0 0 0 2 4 6 8 10 12 14 16 18 20 DANGER WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE I E METER S GROUND SWITCH IS OPEN DON T TOUCH THE UNIT CAUTION If the ...

Page 65: ...complete or Damaged Shipment Please read these procedures before packing unpacking the VIVID P3 We strongly advice you to store the VIVID P3 packing material in undamaged condition in case of future transportation 1 Cut the two Metal Bands CAUTION Do not lift the unit by the Keyboard This may cause equipment damage CAUTION The crate with the VIVID P3 weighs approximately 80 kg Be prepared for a su...

Page 66: ... SERVICE MANUAL 3 4 Section 3 2 Receiving and Unpacking the Equipment Section 3 2 Receiving and Unpacking the Equipment cont d 2 Lift the TOP Cover up and off 3 Remove the Monitor Cap up and off Figure 3 2 Remove the top cover Figure 3 3 Removing the Monitor Cap Monitor Cap ...

Page 67: ...ction 3 2 Receiving and Unpacking the Equipment cont d 4 Remove the three Plastic Joints from the Outer Sleeve Refer to the Labels on the Packing Box for more detailed instructions 5 Remove the Outer Sleeve 6 Remove the Inner Sleeve Figure 3 4 Removing Plastic Joints and Sleeves Palette Assy Plastic Joint Inner Sleeve ...

Page 68: ...t cont d 7 Remove the Plastic Wrapping around the VIVIDTM P3 8 Remove the Monitor Support and Monitor Packing 9 Remove the adhesive tapes attached at the four corners of the Top Cover 10 Put the Ramp Board on floor and prepare the slope to put the console down 11 Unlock the brakes on the front castors then carefully put the console off the Palette Figure 3 5 Removing System from the Pallete ...

Page 69: ...age Although VIVID P3 is a compact and mobile machine two people should move it over rough surfaces or up and down grades 3 2 2 Adjusting System Clock Set the system clock for the VIVID P3 to the local time For procedure of adjusting the system clock refer to 4 5 Software Configuration Checks in Chapter 4 FUNCTIONAL CHECKS 3 2 3 Product Locator Installation Card Fill out proper customer Informatio...

Page 70: ...ERVICE MANUAL 3 8 Section 3 3 Packing the Equipment NOTE The Product Locator Installation Card shown may not be same as the provided Product Locator card PRODUCT LOCATOR INSTALLATION CARD Section 3 3 Packing the Equipment Please pack VIVID P3 in the reverse order of unpacking ...

Page 71: ...agnetic Interference EMI Many of the covers shields and screws are provided primarily to protect the system from image artifacts caused by this interference For this reason it is imperative that all covers and hardware are installed and secured before the unit is put into operation 3 4 4 Checking the Components When a new system arrives check that none of the components are damaged or in short sup...

Page 72: ...ckly 3 5 1 1 Scanner Power On 1 Connect the Power Cable to the back of the system 2 Ensure the Cable Clip slips securely over the shoulders on the molded plug 3 Connect the Main Power Cable to a hospital grade power receptacle with the proper rated voltage Never use an adapter that would defeat the safety ground 4 Switch ON the Circuit Breaker at the rear of the system a Circuit Breaker b Power Ca...

Page 73: ...nce the TMST starts and the software code is distributed to initiate the scanner No status messages are displayed during this process 3 5 2 Power Off Shutdown NOTE After turning off a system wait at least ten seconds before turning it on again The system may not be able to boot if power is recycled too quickly 3 5 2 1 TMST Power Down To power down the system 1 Press the Power On Off switch at the ...

Page 74: ...n 4 The shutdown process takes a few seconds and is complete when the power status LED is turned Amber color 5 Disconnect the probes Clean or disinfect all probes as necessary Store them in their shipping cases to avoid damage 3 5 2 2 Scanner Shutdown 1 Switch OFF the Circuit Breaker at the back of the system 2 Disconnect the Mains Power Cable if necessary For example Relocating the scanner Figure...

Page 75: ...t the transducer connector on the receptacle guide pin until it touches the receptacle mating surface C Twist the transducer lock lever vertically to lock it in place Twist the lever to the horizontal position to disconnect the transducer NOTE Please ensure that the probe latch is in unlocked position before you connect the probe to the system NOTE It is not necessary to turn OFF power to connect ...

Page 76: ...uration Section 3 6 System Configuration 3 6 1 System Specifications 3 6 1 1 Physical Dimensions The physical dimensions of the VIVID P3 console are summarized in Figure 3 9 on page 3 14 Table 3 4 Physical Dimensions of VIVID P3 Height Width Depth Unit 1347 445 595 mm 53 0 17 5 23 4 inches Figure 3 9 Overall Dimensions ...

Page 77: ...Interface B W Printer SONY UP D897 USB B W Printer Mitsubishi P93D USB Digital Color Printer SONY UP D23MD USB HP Color Printer HP HP Officejet Pro K550 USB HP HP Laser Jet P2015dn USB HP HP Officejet Pro K5400 USB HP HP Officejet 6000 USB HP HP Deskjet 460 USB 3 pedal footswitch Steute MKF 2 MED GP26 USB USB Memory SanDisk SanDisk 1GB USB ECG ECG Module ECG USB1 USB USB HDD Iomega USB HDD 160GB U...

Page 78: ...panel are USB Ports VGA S Video Composite Out Ethernet port Power outlet A USB ports for printers ECG B W Color and USB Memory Stick Footswitch DVD RW USB HDD B Ethernet port C VGA Output D S Video Out E Composite video out F Power outlet WARNING WARNING Equipment damage possibility Be sure to use the following recommended connecting cables to connect recording devices and a network with VIVID P3 ...

Page 79: ...d on the Top Right side panel are one USB Port for control connections for printer USB Thumb Drive and service tools NOTE Each outer case ground line of peripheral accessory connectors are protectively grounded Figure 3 11 Left Side Panel NOTICE The USB devices should be connected to VIVID P3 first power on USB devices before turning VIVID P3 to work ...

Page 80: ...4 Pin mini Din Table 3 7 S Video Connector 4 Pin 2 Pin Assignment of USB Table 3 8 Pin assignment of USB1 Table 3 9 Pin assignment of USB2 Pin No Output Signal Description Picture 1 SVIDEO OUT IN YG Y Luma GND 2 SVIDEO OUT IN CG C Chroma GND 3 SVIDEO OUT IN Y Y Luma SIGNAL 4 SVIDEO OUT IN C C Chroma SIGNAL Pin No Signal Pin No Signal 1 5VDC 3 DATA 2 DATA 4 GND Pin No Signal Pin No Signal 1 5VDC 3 ...

Page 81: ...Connecting B W printer UP D897 Mitsubishi P93D or Color printer UP D23MD to the system Only B W Printer UP D897 can be mounted in Front Utility area tray as shown in the figure below Figure 3 12 Connect B W printer Tray to the system Fig A Fig B VGP Tray Bottom VGP Tray top Using two Captive fastener mount VGP Tray bottom to the Sytem Utility box ...

Page 82: ...ws as indicated below Figure 3 13 Mounting of VGP Tray to Printer A A A A Mounting the bracket to the printer by using M3X8 Screws Front Side Of the Printer VGP UP897 Mounting Bracket 5314101 Rear Side of the Printer A Indicates the location of the mounting holes on the VGP UP D897 Mounting Bracket B B Indicates the location of the mounting holes on the Printer Front side of the Printer Rear Side ...

Page 83: ...Tray fixed on the System and then connect USB Power cables in the rear side of the printer and push back fully and tighten the front captive fastener as indicated in the figure below D Mounting procedure for Color printer UP D23MD connect the USB Cable to any of the USB ports available on rear panel and connect the power cable to the rear panel Power port Figure 3 14 Mounting the Sony UPD 897 on V...

Page 84: ... on rear panel G Connecting the CRT to the system SVGA connector available on Rear panel H Connect the USB Memory stick to the system The USB Memory stick can be properly connected in top panel USB port or in any of the Rear panel USB Ports I Connect the USB HDD to the system The USB Harddisk can be properly connected to Rear Panel USB ports NOTE Please scan the devices USB Pen Drive HDD etc for v...

Page 85: ...rint and select the respective printer K Connect the external monitor to the system The monitor can be properly connected using the VGA port in the Rear panel Press Ctrl Alt V on the keyboard a dialog box appears Select Intel R Dual Display Clone check the Same display configuration driver on both display box and select OK NOTE Please refer to the operation manual for each peripheral information n...

Page 86: ...ist of Probes for Vivid P3 Probe Name Area of Using TYPE Catalog Number 4C GENERAL PURPOSE CONVEX H4904PC E8CS TRANSVAGINAL TRANSRECTAL MICRO CONVEX H44801HR 8C Abdomen OB GYN Urology MICRO CONVEX H40412LJ 11L Vascular Small Parts LINEAR H40412LY 8L SMALL PARTS PERIPHERAL VASCULAR LINEAR H4001DB 3S Cardiac Transcranial SECTOR H4701SZ 5cs Abdomen OB GYN Urology CONVEX H4001DC 6S Cardiac Pediatric S...

Page 87: ...or information on configuring items like Hospital Department Language Units of measure Date Time and Date Format For information on configuring Software Options refer to the VIVID P3 Basic User Manual Chapter 16 Customizing Your System For information on configuring DICOM Connectivity refer to the VIVID P3 Basic User Manual Chapter 16 Customizing Your System NOTE After Enabling the Software Option...

Page 88: ...orksheet Site System Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Name AE Title Dept LOGIQ SN Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone E Mail Address Site IP Settings IP Address Subnet Mask Default Gateway Remote Archive Setup Remote Archive IP Remote Archive Name ...

Page 89: ... GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 3 Installation 3 27 Section 3 8 Loading Base Image Software This information has been moved please refer to Section 8 13 Loading Base Image Software on page 8 86 ...

Page 90: ...ck out Section 3 9Software Version check out 3 9 1 Functional Check out 1 Power on VIVID P3 scanner and wait until system booting to main screen 2 Press Config key on control panel 3 Choose the About button on the right 4 Check whether Software version is the right version for use Figure 3 19 About Figure 3 20 Software version ...

Page 91: ...evant safety and user information is available during the operation and service of the complete system 3 10 1 Product Locator Installation NOTE The Product Locator Installation Card shown may not be same as the provided Product Locator card 3 10 2 User Manual s User Check that the correct User Manual s for the system and software revision is included with the installation Specific language version...

Page 92: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 3 30 Section 3 10 Paperwork ...

Page 93: ... Vivid P3 console diagnostics by using the built in service software and power supply adjustments Section 4 2 Required Equipment To perform these tests you ll need any of the sector linear or convex transducers normally you should check all the transducers used on the system Section Description Page Number 4 1 Overview 4 1 4 2 Required Equipment 4 1 4 3 General Procedure 4 2 4 4 Software Configura...

Page 94: ...mber color 4 3 1 1 Boot Up In the control Panel press the power On Off key the led color will turn in Green color the system will boot CAUTION SYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place The covers are required for safe operation good system performance and cooling purposes NOTICE Lockout Tagout Requirements For USA only Follow OSH...

Page 95: ...s pressed once the Back end Processor starts and the software code is distributed to initiate the scanner No status messages are displayed during this process 4 3 2 Power Off Shutdown NOTE After turning off a system wait at least ten seconds before turning it on again The system may not be able to boot if power is recycled too quickly 4 3 2 1 TMST Power Down To power down the system 1 Press the Po...

Page 96: ...plete when the power status LED is turned Amber 5 Disconnect the probes Clean or disinfect all probes as necessary Store them in their shipping cases to avoid damage 4 3 2 2 Scanner Shutdown Disconnect the Mains Power Cable is necessary For example Relocating the scanner CAUTION DO NOT unplug and or transport the unit until after the power off sequence has been completed Failure to do so may resul...

Page 97: ...4 3 3 1 Archiving Presets to an DVD R Disk or USB memory device 1 Insert an empty blank DVD R disk into the DVD RW 2 Access the Config Menu and select System The Backup sheet will be shown on the LCD display 3 Select the item to back up from the Backup section 4 Enter backup destination or browse through the disk to locate the destination 5 Select Backup now The backup status for each item is disp...

Page 98: ...nto the DVD RW 2 Access to the Config Menu and select System The Restore sheet will be shown on the LCD display see Figure 4 4 3 Select the item to restore from the Backup section 4 Enter restore destination or browse through the disk to locate the destination 5 Select Restore The restore status for each item is displayed on the Result column 4 3 4 Adjusting the Display Monitor Please refer to Sec...

Page 99: ...atient 2 App Selection Preset Key 3 Work sheet 4 Review Report End Exam 5 TGC 6 Sub menu keys 7 Userdefined key 2 Measure Trackball Zoom Set pause Depth Cursor Pointer Userdefined key1 8 2D M CF PW PDI Alt Gain 9 L R Store P1 10 Freeze 11 Steer 12 Auto Ao 13 Print keys Physio 14 Protocol Report 15 Alphanumeric keys 16 NTUPI Top menu keys Figure 4 5 Control Panel Tour ...

Page 100: ...enu Key Tour In general there are two types of NTPUI SoftMenu keys Paddle Switch and adjustable knobs 1 The Paddle Switch is used to access and adjust the Sub SoftMenu 2 Press the adjustable knobs to toggle option menu between line one and line two 3 Rotate the adjustable knobs to adjust the corresponding parameters Figure 4 6 SoftMenu Key Tour ...

Page 101: ...cal Zone 3 Acoustic Output Readout 15 TGC not shown on the image 4 GE Symbol Probe Orientation Marker Coincides with a probe orientation marking on the probe 16 Body Pattern 5 Image Preview 17 Depth Scale 6 Gray Color Bar 18 SoftMenu 7 Cine Gauge 19 Caps Lock On Off 8 Measurement Summary Window 20 Start menu icon 9 Image 21 Network icon 10 Measurement 22 Trackball Functionality Status Scroll M A M...

Page 102: ...pter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on already Figure 4 8 Controls available in B Mode Figure 4 9 B Mode Screen Picture Example SoftMenu Key Primary Menu Scan Area Zoom Secondary Menu Frequency Virtual Convex Gray Map Dynamic Range Focus Position Rotation Focus Number Frame Average Biopsy Kit Edge Enhance Focus Width B Softner Suppression Powe...

Page 103: ...ress the control to toggle between Focus Position and Focus Number Press Up Down Button to move or adjust the focal numbers 5 Activate Auto Optimize Optimize the image based upon a specified region of interest or anatomy Press the Center Button in the Gain Dial to toggle the ATO ACE On and Off 7 Adjust Time Gain Compensation TGC Amplifies the returning signals to correct for the attenuation caused...

Page 104: ...lti Frequency mode lets you downshift to the probe s next lower frequency or shift up to a higher frequency 6 Adjust Frame Average Temporal filter that averages frames together This has the effect of presenting a smoother softer image 7 Adjust Rotation Rotates the image by selecting the value from the pop up menu 9 Adjust Line Density Optimizes B Mode frame rate or spatial resolution for the best ...

Page 105: ... Available Probes on page 3 24 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on already Figure 4 10 Controls available in M Mode Figure 4 11 M Mode Screen Picture Example Primary Menu Secondary Menu M Mode M Cursor frequency Dynamic Range Sweep Speed Display Format Full Timeline Colorize Edge Enhance Anatomical M Gray Map Gain Rejection Colorize P...

Page 106: ...OP Panel Controls Step Task Expected Result s 1 Press M Mode key M Mode Starts 2 Adjust Gain Controls the amount of echo information displayed in an image Turn B Mode dial to the left right to increase decrease Gain Gain displays on the monitor in G dB 3 Display M Mode Cursor Displays the M Mode cursor on the B Mode image Press Cursor and Trackball to position M Mode Cursor ...

Page 107: ...sponding to the edges of structures Adjustments to M Mode s edge enhancement affects the M Mode only 4 Activate Gray Map Determines how the echo intensity levels received are presented as shades of gray 6 Activate Colorize Enables gray scale image colorization To deactivate reselect a Gray Map 7 Activate Full Timeline Displays only timeline screen Press the Full Timescreen to activate 8 Select Dis...

Page 108: ...ge 3 24 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on already Figure 4 12 Controls available in Color Flow Mode Figure 4 13 CFM Mode Screen Picture Example Primary Menu CF Mode Secondary Menu Frequency Angio Line Density Variance Baseline Packetsize Scale Map Invert Low Vel Reject Threshold Spatial Filter Flash Suppression Transparency Map Focu...

Page 109: ...7 4 3 8 2 Color Flow Mode OP Panel Controls Table 4 7 Color Flow Mode OP Panel Controls Step Task Expected Result s 1 Press CFM Mode key CFM Mode Starts 2 Adjust Gain Amplifies the overall strength of the echoes processed in the Color Flow window Turn the Gain dial CFM Mode key to the left right to increase decrease Gain ...

Page 110: ...Range Dynamic Range controls how echo intensities are converted to shades of gray thereby increasing the adjustable range of contrast 9 Activate ACE Eliminates the motion artifacts Press Ace to activate 10 Adjust Angle Steer Slants the Color Flow region of interest or the Doppler line to obtain a better Doppler angle 11 Move Baseline Adjusts the baseline to accommodate faster or slower blood flows...

Page 111: ...canner if it isn t turned on already Figure 4 14 Controls available in Doppler Mode Figure 4 15 Doppler Mode Screen Picture Example Primary Menu Secondary Menu PW Mode B Pause Frequency Baseline Quick Angle Sweep Speed Scale Angle Correct Invert Low Vel Reject Rejection Dynamic Range Display Format Full Timeline Trace Method Trace Sensitivity Trace direction Modify Auto Calcs PW CF Ratio Duplex Ti...

Page 112: ...plifies the overall strength of the echoes processed in the Color Flow window Turn the Gain dial PW Mode key to the left right to increase decrease Gain 3 Display M D Mode Cursor Displays the M D Mode cursor on the B Mode image Press Cursor and Trackball to position sample volume graphic Click SV gate to adjust sample volume gate size 4 B Pause Toggle between simultaneous and update presentation w...

Page 113: ...ed value The selected frequency is displayed in the status window 6 Trace Direction Allows to select different trace direction 7 Invert Vertically inverts the spectral trace without affecting the baseline position Press invert to invert the spectral trace The Plus and Minus signs on the velocity scale reverse when the spectrum is inverted 8 Auto Calculation Enables or disables auto calculation 9 M...

Page 114: ...nvert the spectral trace The Plus and Minus signs on the velocity scale reverse when the spectrum is inverted 8 Auto Calculation Enables or disables auto calculation 9 Modify Calcs Activates the window to modify the auto calculation items 10 Trace Method Allows to select different trace method 11 Activate Colorize Colorize the gray scale image to enhance the eyes discrimination capability Press th...

Page 115: ...esolution in frequency domain 20 Spectral Average Optimizes the smoothness of the spectrum Different levels can be selected 21 Power output Optimizes image quality and allows user to reduce beam intensity 10 increments between 0 100 Values greater than 0 1 are displayed 22 SV Length Sizes the sample volume gate 23 Low Vel Reject Wall Filter insulates the Doppler signal from excessive noise caused ...

Page 116: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 4 24 Section 4 3 General Procedure Figure 4 15 CW Doppler Mode Screen Picture Example ...

Page 117: ...plifies the overall strength of the echoes processed in the Color Flow window Turn the Gain dial PW Mode key to the left right to increase decrease Gain 3 Display M D Mode Cursor Displays the M D Mode cursor on the B Mode image Press Cursor and Trackball to position sample volume graphic Click SV gate to adjust sample volume gate size 4 B Pause Toggle between simultaneous and update presentation w...

Page 118: ...gray scale image to enhance the eyes discrimination capability Press the Cololize Trackball to cycle through available maps and press Set to select 10 Activate Gray Map Displays a map window adjacent to the image Move the trackball to select the map The image reflects the map as scrolled through the selections Press Set to select 11 Dynamic Range Controls how echo intensities are converted to shad...

Page 119: ...ness of the spectrum Different levels can be selected 19 Power output Optimizes image quality and allows user to reduce beam intensity 10 increments between 0 100 Values greater than 0 1 are displayed 20 SV Length Sizes the sample volume gate 21 Low Vel Reject Wall Filter insulates the Doppler signal from excessive noise caused from vessel movement Table 4 12 CW Doppler Mode Touch Panel Controls S...

Page 120: ...stem 2 Turn ON the scanner if it is not turned on already 4 3 11 2 Activating CW Doppler Ensure that the appropriate CW probe is connected Press CW key on the keyboard The Doppler spectrum appears along with the CW Top Sub menu The following CW parameters are displayed Frequency Gain Acoustic Output Scale Wall Filter and Dynamic Range 4 3 11 3 Exiting CW Doppler To exit CW Doppler mode press CW ke...

Page 121: ... 2 ECG Check Table 4 13 ECG Check Step Task Expected Result s 1 Connect the ECG harness to the connector on the front of the system The unit displays a straight curve along the bottom edge of the image sector on the screen 2 Connect the three leads to a ECG simulator or Fasten the three ECG Pads to your body and connect the three leads to respective ECG Pad When connecting the signal on the screen...

Page 122: ...ta that has been measured to this point but not data entered onto worksheets press CLEAR 4 3 13 2 Circumference Area Ellipse Measurement 1 Press MEASURE once an active caliper displays 2 To position the active caliper move the TRACKBALL 3 To fix the start point press SET The system fixes the first caliper and displays a second active caliper 4 To position the second caliper move the TRACKBALL 5 Ad...

Page 123: ...e keyboard toward the probe holder Ensure that the probe is placed gently in the probe holder 4 3 14 4 Disconnecting the probe Probes can be disconnected at any time However the probe should not be selected as the active probe 1 Unlock the probe latch downward 2 Pull the probe and connector straight out of the probe port 3 Carefully slide the probe and connector away from the probe port and around...

Page 124: ...djust the Loop Speed up down Two Button Softkey to increase decrease the CINE Loop playback speed 4 3 15 5 Moving through a CINE Loop Frame By Frame Adjust the Frame by Frame up down Two Button Softkey to move through CINE memory one frame at a time 4 3 16 Image Management QG For Image Management functionality refer to the Vivid P3 Quick Guide It talks about several topics Clipboard Printing Image...

Page 125: ...perly label and Insert the backup media 3 Select the media type from the drop down menu 4 Enter the label for the media as shown in It is best to use all capital letters with no spaces or punctuation marks Press Format 5 The Ultrasound system displays a pop up menu as shown in When the formatting has been completed press OK to continue 6 If desired verify that the format was successful by returnin...

Page 126: ...leshooting the PHI must be encrypted or protected from unauthorized access and the files must be compressed using PKZIP with a password The PHI must be sent to GE employees only through the GE intranet For VIVID P3 the GE intranet address is http libraries ge com foldersIndex do entity_id 12065698101 sid 101 sf 1 NOTE PHI Patient HealthCare Information data should not be sent to the GE Employees I...

Page 127: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 4 Functional Checks 4 35 Figure 4 18 Backup Restore Menu Check here to backup presets and configurations ...

Page 128: ...privileges the Operator Login window is displayed Log on with administrator privileges 5 In the Restore list select Patient Archive Report Archive and User Defined Configuration 6 In the Media field select the Backup Restore DVD RW or USB memory device 7 Select RESTORE The system performs the restore As it proceeds status information is displayed on the Backup Restore screen CAUTION The restore pr...

Page 129: ...m the hard drive by date Therefore the best way is to label media by date NOTE Images will be moved from the hard drive by date Therefore the best way to label media is by date When images are moved to the archive media they will be deleted from the system hard drive However the patient database backed up earlier maintains pointers to the location of the images on the archive media 5 Select Backup...

Page 130: ...5 Archiving Images cont d 7 Press PATIENT and set the Dataflow to store images directly to DVD ROM 8 From the image screen press EZBACKUP MOVE The Move Images pop up appears 9 Press Next on pop up message 10 A media check message appears Press OK to continue Figure 4 22 Image Archive Move Pop up Menu Figure 4 23 Media check message ...

Page 131: ...11 Press Finish after Backup Move complete All databases presets and images should now be saved to removable media 4 3 17 6 Full Backup 1 Connect the USB HDD to system 2 Select Backup Restore then Emergency Repair Disk 3 Select Reminder Dialog Interval Days Figure 4 24 EZBackup Move complete Figure 4 25 Emergency Repair Disk Reminder Dialog Interval Days ...

Page 132: ...General Procedure 4 3 17 6 Full Backup cont d 4 Return to scan mode Press Ctrl B the Emergency Disk Making window pop up appears 5 Press Next on pop up message 6 A storage information message appears Press Next to continue Figure 4 26 Emergency Disk Making Pop up Menu Figure 4 27 Storage Information message ...

Page 133: ...1 4 3 17 6 Full Backup cont d 7 Press Finish after Emergency Disk Making complete All databases presets and images should now be saved to Emergency Disk 8 Press Ctrl R Emergency Disk Recovery window pop up appear 9 Press Next on pop up message Figure 4 28 Emergency Disk Making Complete Figure 4 29 Emergency Disk Recovery ...

Page 134: ...ICE MANUAL 4 42 Section 4 3 General Procedure 4 3 17 6 Full Backup cont d 10 A recovery information message appears Press Next to continue 11 Press Finish after Emergency Disk Recovery complete Figure 4 30 Recovery Information message Figure 4 31 Emergency Disk Recovery Complete ...

Page 135: ...t 3 Check Language settings Desired Language is displayed 4 Check assignment of Printer Keys The default function for Print1 3 Keys is P1 store image P2 print P3 USB Quick Save Print1 3 Keys can also be assigned as desired by the customer 5 Check that all of the customer s options are set up correct All authorized functions are enabled Table 4 15 Peripheral Checks Step Task to do Expected Result s...

Page 136: ... products will be infected by internet spread viruses or trojan horses Network worms All GE U S products can be networked and are potentially vulnerable to infection GE U S has taken and will continue to take precautions to limit this vulnerability Removable Disks GE U S products support the use of removable disks But these GE products will actively use the following kinds of disks only GE softwar...

Page 137: ... chapter explains VIVID P3 s system concepts component arrangement and subsystem function It also describes the Power Distribution System PDS and probes Table 5 1 Contents in Chapter 5 Section Description Page Number 5 1 Overview 5 1 5 1 1 Block Diagram Vivid P3 5 2 5 2 PWA Assy Diagrams 5 10 5 3 Power Diagrams 5 21 5 4 Common Service Platform 5 22 5 5 RFS Request for Service 5 24 5 6 Machine RFS ...

Page 138: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 55344303 100 REVISION 6 VIVID P3 SERVICE MANUAL 5 2 Section 5 1 Overview 5 1 1 Block Diagram Vivid P3 Figure 5 1 Vivid P3 System Block Diagram ...

Page 139: ...nd White imaging 2D Color Flow M Mode Black and White imaging Doppler A number of combinations of the above VIVID P3 is a digital beam forming system that can handle up to 192 elements linear probes Signal flow from the Probe Connector Panel to the Front End to the Mid Processors and TMST and finally to the LCD and peripherals System configuration is stored on a hard disk and all necessary softwar...

Page 140: ... the signal is presented to the mid processor section PAM PreAmplifier The preamplifier amplifies 64 echo signals The reception signals are sent to ADC on RX64 RX64 VIVID P3 Front Processor Convert the analog echo signal to digital TX64 VIVID P3 Transmission Board This has 64 channel bipolar drivers Local Bus 32bits Probe Connector TX64 TMST DSP PCI CHA COM FPC 128 HV MUX Pulser x64 TPG2 x4 TX CTR...

Page 141: ...rt stores it in a memory performs scan conversion to pixel domain and drives the system LCD display 5 1 5 Top Console The Top Console includes a Standby On switch a keyboard different controls for manipulating the picture quality controls for use in Measure Analyze M A and loudspeakers for stereo sound output used during Doppler scanning HDD FPGA DSP PCI Bus To CPU Power Supply AGP Graphic SVGA LC...

Page 142: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 55344303 100 REVISION 6 VIVID P3 SERVICE MANUAL 5 6 Section 5 1 Overview Figure 5 4 Top Console Speakers On Off Switch Keyboard ...

Page 143: ...5 5 External I O module overview External I O The external I O is the interface between the scanner and all external items Examples Network USB interface medical grade printer and external medical grade SVGA displayer 5 1 7 Peripherals DVR Color printer Footswitch VGA monitor TV and a Black White Printer and Bluetooth printer can be connected to the external I O ...

Page 144: ...CE MANUAL 5 8 Section 5 1 Overview 5 1 8 Wiring Figure 5 6 Wiring Diagram Vivid P3 TX64 VIVID P3 Transmit board RX64 VIVID P3 Front end processing board MST VIVID P3 Master board PWR SW Power Switch FPC Flexible Print Circuit board PDB Power Distribution Board SSR Solid State Relay ...

Page 145: ...nput to SOM and the power supply unit which will power ON enabling the 12V and 5V to the entire system 5 1 8 2 Power Down sequence description 5 1 8 2 1 Overview There are two ways to switch off the system 1 Select shutdown from the Application 2 Keep the Keyboard ON OFF button pressed for more than 4 seconds Figure 5 7 On Off Sequence Keyboard power button If pressed ON for 4 secs MST SOM PS ON L...

Page 146: ...HCARE PROPRIETARY TO GE DIRECTION 55344303 100 REVISION 6 VIVID P3 SERVICE MANUAL 5 10 Section 5 2 PWA Assy Diagrams Section 5 2PWA Assy Diagrams 5 2 1 TMST Figure 5 8 TMST Block Diagram To FPGA To Power Block ...

Page 147: ...HEALTHCARE PROPRIETARY TO GE DIRECTION 55344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 5 Components and Functions Theory 5 11 5 2 1 TMST cont d TMST Block Diagram cont d TUSC FPGA Cont Power Block ...

Page 148: ...ower diagnostic circuit ETX Card CPU unit 5 2 1 1 Description This diagram describes the TMST board It controls the front end of VIVID P3 and also communicate with PC system through PCI interface The main function Generate clock signal and distributing each clock signal Generate DUSC bus cycle Power diagnostics HV LV Transmit image raw data B CFM DOP to PC after assembling packet including header ...

Page 149: ...tal converter OQX2 Beamform ASIC CHACOM B M CFM Dop signal processor ASIC FPIC FPGA Front Process interface control 5 2 2 1 Description This diagram describes the RX64 Front Process board It resides below TX board located at the bottom layer The main function Convertor 64 channels echo which is from preamp to digital signal Focus the received RF signals by digital beamforming technology Control CH...

Page 150: ...rams 5 2 3 TX64 Figure 5 10 TX64 Block Diagram TX64 VIVID P3 Transmit board TPG2 Transmit Pulse Generator TMST VIVID P3 Master board TXMX FPGA Transmit and MUX controlling FPGA 5 2 3 1 Description This diagram describes VIVID P3 TX64 board The main function Generate the transmit pulse Drive the Tx pulse with high voltage ...

Page 151: ...nents and Functions Theory 5 15 5 2 4 CWD Figure 5 11 CWD Block Diagram CWD Continue wave doppler board DEM Demodulator HPF High pass filter LPF Low pass filter 5 2 4 1 Description This diagram describes VIVID P3 CWD board it resides under RX64 board The main function Generate continue wave form image data ...

Page 152: ...Board one each for each probe connector There is one FPGA that acts like a buffer for the Probe related signals and is the interface between the probe and the main TX Board FPGA There is a I2C slave also implemented on this FPGA that responds and interfaces with the Host I2C controller on the TX board it is mainly for probe selection and control The Power supply to the TX board from the PDB board ...

Page 153: ...r like 1 ethernet port 1 VGA port 4 USB ports BNC Connector and one S Video connector The Ethernet signals and VGA signals are directly routed from the MST board via the FRC cable with buffering For the Video out signals BNC and S Video there is a video conversion logic built which converts the VGA input into the S video and Composite video formats The resolution of the Video output would be 640x4...

Page 154: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 55344303 100 REVISION 6 VIVID P3 SERVICE MANUAL 5 18 Section 5 2 PWA Assy Diagrams Figure 5 14 Rear Panel 1 2 Vivid P3 Standard VGA output Resolution 640x480 ...

Page 155: ...aces 1 USB controller that interfaces the Keyboard to the System SW The ultrasound Keys are directly mapped to this controller 2 Trackball and Alphanumeric Keyboard is connected to the system via the PS2 interface There is one Microcontroller that decodes the Alphanumeric key press information and relays this information to the system via the PS2 3 One USB port is extended from the MST board and i...

Page 156: ...n Control panel Power switch Amber When StandBy Turns Green when switched ON Main Power activity RX64 DS1 Green Normal Condition FPGA Config TX64 DS1 Green Normal Condition FPGA Config PDB Green Normal Condition 5v Green Normal Condition 12v Green Normal Condition 5v_ Stand by Connector Board D5 Green Normal Condition Probe 1 Connected D6 Green Normal Condition Probe 2 Connected D7 Green Normal Co...

Page 157: ...5 16 AC Power Distribution Block Diagram Vivid P3 The mains cord has plugs in one side end On Wall A male plug of Power Cord connects to the mains outlet on site From the Main Circuit Breaker the AC power is routed via an Inrush Current Limiter to a internal outlet connector for the Main power supply The power ON OFF is actually controlled by the MST PWA This signal is routed from the Keyboard ON ...

Page 158: ... service platform is designed for GE personnel and as such is in English only There is no multi lingual capability built into the Service Interface 5 4 2 2 Service Login Select Utility Service This button links the user or the Field Engineer FE to the service login screen 5 4 2 3 Access Security The service interface has different access and security user levels Each user is only granted access to...

Page 159: ...4 For a service call with no FRU replacement use the following debrief Script Refer the Service Manual Chapter 4 Basic Functional Checks in the service manual Equipment has passed all the required tests and is ready for use Table 5 3 Service Login User Levels User Level Access Authorization Password Operator Authorized access to specified diagnostics error logs and utilities Same acquisition diagn...

Page 160: ...nnected to network configured and checked out to be able to send the RFS 1 Right Click on Connect to GE icon Figure 5 19 Contact GE icon All the fields that are marked with asterisk are mandatory and should be filled in order to send the RFS request to On Line Center 2 Fill out the Last name First name and Phone number Extension Email and Other system ID fields are optional 3 Select Problem Type I...

Page 161: ...etailed problem description in the Problem Description field Figure 5 20 RFS Window 6 Then the Send button is enabled Figure 5 21 RFS Window 7 Verify that Send button is enabled 8 Click Send button and wait for confirmation window with reference number on the screen 9 Verify that confirmation message pops up in the screen 10 RFS request is completed ...

Page 162: ...r high or low voltage 5 6 1 Enabling MRFS NOTE In order to enable MRFS System must be connected to network and should be checked out to a product server Default MRFS user should also be set 1 If a system is not already checked out the system should be checked out to a product server 2 MRFS function is an automatic call for service from the system to the OLC when system detects high temperature or ...

Page 163: ...dd User All the fields that are marked with asterisk are mandatory and should be filled 6 When fields with asterisk are filled Add User button is enabled Figure 5 25 Add User 7 Pess Add User button 8 Check the checkbox next to the name to set to the default contact Then Press Set Default Machine Contact button Figure 5 26 Set Default Machine Contact ...

Page 164: ...Y TO GE DIRECTION 55344303 100 REVISION 6 VIVID P3 SERVICE MANUAL 5 28 Section 5 6 Machine RFS 9 Now the MRFS is enabled 10 When system detects high temperature or abnormal voltage it will send the RFS to the OLC automatically ...

Page 165: ... 1 Chapter 6 Service Adjustments Section 6 1 Overview 6 1 1 Purpose of this chapter 6 This section describes how to test and adjust the scanner These tests are optional You may use them to check the system for errors Table 6 1 Contents in chapter Section Description Page Number 6 1 Overview 6 1 6 2 Monitor Adjustments 6 2 ...

Page 166: ...s Indicator 4 Light Button 5 Adjustment Button 6 Toggle Button 7 Adjustment Button NOTE Default Factory setting for Contrast Bightness is 80 and Dim brightness 50 as shown in Fig A Dim brightness is the LCD lamp brightness control 6 2 2 Monitor Calibration procedure 1 Select Utility in A N Keboard Test patterns Select the Brightness calibration and perform the monitor calibration procedure as disp...

Page 167: ...he tools and software that help maintain image quality and system operation Very basic host system and board level diagnostics are run whenever power is applied Some Service Tools may be run at the application level However most software tests are required Table 7 1 Contents in Chapter 7 Section Description Page Number 7 1 Overview 7 1 7 2 Gathering Trouble Data 7 2 7 3 USB Quick Save 7 4 7 4 Scre...

Page 168: ...on to be sent back to the manufacturer for analysis There are different options to acquire this data that would give different results 7 2 2 Collect Vital System Information The following information is necessary in order to properly analyze data or images being reported as a malfunction or being returned to the manufacturer Product Name VIVID P3 From the Config System About screen Applications So...

Page 169: ...og Error Logs Crash Log Power Supply Temperature NOTE Power Supply and Temperature logs are not currently being updated by the VIVID P3 This Alt D function is available at all times When Alt D is pressed a menu box appears that allows for A place to enter a description of the problem A choice to store to a pre formatted DVD R RD Removable Disk or to the Export directory D export The subsequent fil...

Page 170: ...e However the default is for the Image Area only or the customer may have customized the P3 Key function 7 3 2 Check and Record the P3 Key Function Check the function of the Print 3 Key in the event that the customer may have made some customized settings 1 Click Config on the keyboard 2 Select Connectivity from the Utilities Menu 3 Select the Buttons tab on the Connectivity screen 4 In the Physic...

Page 171: ...3 Key to USB Quick Save If the P3 Key is not set to USB Quick Save 1 While on the Connectivity screen with the Buttons tab displayed go to the Destinations list 2 From the list select USB Quick Save Press to add the selection to the Printflow View section 3 Ensure that the Physical Print Buttons section for capture Area is set to Image Area and No Image Compression 4 The P3 Key should now be set u...

Page 172: ...u 3 Select the Buttons tab on the Connectivity screen 4 In the Physical Print Buttons field select Print3 The Connectivity Buttons Screen will be displayed like the one shown in Figure 7 2 on page 7 4 If P3 is not set to Whole Screen as shown in Figure 7 2 proceed to step 5 to record the customer s customized settings 5 In the Destinations section record the service that is displayed 6 In the Phys...

Page 173: ... cont d 3 Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom 4 Highlight the snapshot to be stored to RD Removable Disk or DVD R 5 Select Menu on the right side of the image screen then highlight and select SAVE AS Figure 7 3 Select Image to Capture Figure 7 4 Menu Save As Clipboard Thumbnails Menu Save As ...

Page 174: ...onality recorded in section 7 4 1 on page 7 6 Refer to Figure 7 2 1 Click Config on the keyboard 2 Select Connectivity from the Utilities Menu 3 Select the Buttons tab on the Connectivity screen 4 In the Physical Print Button field select Print3 5 In the Destinations list select the service s recorded in step 5 Section 7 4 1 6 In the Physical Print Buttons section select the parameters related to ...

Page 175: ... LAN Network 1 If user wants to setup static IP address uncheck Enable DHCP option input static address in IP Address box Subnet Mask and Default Gateway box In Network Speed box choose the proper speed available 2 Click Save in popped up dialog and choose Ok 3 After reboot the cross mark on the network icon at the left bottom of screen disappears ...

Page 176: ... d Is the power On Off LED in Amber color lit YES Check 220vac 10 at i p available to the transformer by conforming whether stand by led is lit NO Does power On Off LED is Green NO Check main cable between KBD and PDB board Are the LEDs on PDB module lit NO YES 1 Check cable connections from KBD to PDB 2 If problem persists replace PDB Module the Cable b w PDB KBD System doesn t boot Replace PDB b...

Page 177: ...splay on the LCD Go to Scan Procedure section NO Reboot the system and Reload latest system Application software with DVD RW YES Is BIOS screen displayed on the LCD NO Is BIOS screen displayed on external Monitor NO Check flat Ribbon cable B W DCK RPI YES 1 Check LCD brightness 2 Check LCD Power Signal Cable 3 Check LCD Replace if found defective else replace MST DCK YES YES Reboot system if still...

Page 178: ...S Is low sensitivity for all types of probe YES check HV voltages on the Service Diagnostics Check the HV power supply unit on TX64 refer to error code list NO Run Service diagnostics and replace defective board Replace the probe with low sensitivity NO Is Image sensitivity recovered after changing probe YES Refer to Ch 6 Service Adjustment to adjust LCD brightness NO END YES Reload application SW...

Page 179: ...iver ROD SCAN B Mode image quality is low Reconnect the probe check NO YES Is irregular pattern found by rod scan Are low sensitivity and lack of sensitivity found by rod scan Check Conn Board TX64 TMST RX64 refer to error code list YES NO Is the scanner set to the proper configuration YES Reset the system configuration as default i e reload the factory defaults NO Adjust the brightness of LCD dis...

Page 180: ...e any electrical devices that affect the Scanner Keep the Scanner away from those electrical devices NO YES If the probe connect properly END NO YES Is the PDB TX64 RX64 assembled properly NO Ensure board connection YES 1 Re connect the probe 2 Check the probe whether it is broken on the surface Is the shielding emi cover assembled properly NO Assemble it properly YES Are all cable located in righ...

Page 181: ...or Flow Low Sensitivity Color Flow sensitivity is low Go to B Mode Low Sensitivity section YES NO Is the sensitivity normal for B Mode Is the sensitivity low for all types of probe Check the probes NO YES Is the sensitivity low for all area of the ROI YES Is Conn board power supply on TX64 ok Check TMST board refer to error code list NO Replace the TX64 board NO Check RX64 board YES ...

Page 182: ...oppler Low Sensitivity Figure 7 12 Spectrum Doppler Low Sensitivity Spectrum Doppler sensitivity is low Go to B Mode Low Sensitivity section YES NO Is the sensitivity normal for B Mode Is the sensitivity low for all types of probe Check the probes NO YES Is Conn board power supply on TX64 ok Replace the TX64 board NO YES Check the RX64 board ...

Page 183: ... NO Is there any electrical devices that affect the Scanner Is the System Ground taken in right way Set the System Ground NO YES Does the noise vary according to probe cable routing NO END Replace the probe YES Is the TX64 RX64 Flex Cables assembled or in serted properly YES NO Re Insert the PWAS Cables Screw the screws Is the shielding emi cover assembled properly YES Are all cable located In ser...

Page 184: ...the probe connection YES NO Is probe connected in right way Whether the system can display the probe name Whether the probe connected well NO YES Is the system in Freeze status YES UnFreeze the system NO Go to the next page NO Reconnect the probe YES 1 Check Connector Board TX RX and TMST 2 Check TMST to RX Flex cable Whether the system can be unfreezes NO 1 Check KBD 2 Check TMST 3 Check the wire...

Page 185: ...leshooting 7 19 7 6 1 8 Scan Procedure cont d Figure 7 15 Unable to scan cont d Unable to scan con d Increase the Gain Is TX64 ok Restart the system YES NO NO YES Is Flex cables ok Replace Flex cables Restart the system YES Is there back ground noise NO Replace TX64 board Is TMST2TX ok YES YES NO Replace TMST2TX board ...

Page 186: ...3 SERVICE MANUAL 7 20 Section 7 6 Troubleshooting 7 6 1 9 Trackball Figure 7 16 Trackball The trackball is low sensitive Remove the ball from the Trackball and clean it YES NO Doesn t the trackball move at all END The trackball module is defective and replace it in Key board ...

Page 187: ...7 6 1 10 LCD Display Figure 7 17 LCD Display Error on the LCD display Check the cable from Docking board to LCD also check LCD NO YES Does the LCD display nothing Does the colors seem abnormal Check cable from Docking board to LCD YES NO Does the Brightness adjustment seem improper Set up the brightness YES Check the LCD Assy NO END ...

Page 188: ...g by Printer Figure 7 18 Unable Recording by Printer Unable to record by the Printer 1 Check power supply connection to the printer 2 Check the printer connection to the system YES NO Is the connection set in the right way Are there printer papers left Insert the printer papers NO YES Is the proper configuration for P key set Set the P key by the Utility menu NO Change printer END YES ...

Page 189: ... 7 23 7 6 2 2 Drive Issue DVD RW Figure 7 19 Unable Recording by DVD RW Unable to record by the DVD RW Check all signal and power supply connections YES NO Is the connection set in the right way Is there blank DVD R disk inside Insert the blank DVD R NO YES Is the blank has been formatted Format the DVD NO Change DVD RW END YES ...

Page 190: ...k for Video cable of LCD Change LCD setting in system Replace the LCD 7 6 4 Network Troubleshooting Trees 7 6 4 1 No Connection to the Network at All 1 Check if the LCD indicates the network connection is working in normal status Click Start button on the bottom left of the status bar select Network Connections refer to 7 69 a Check whether your Wireless Network Connection or Local Area Connection...

Page 191: ...to the Network at All cont d 2 Check that the network cable between the scanner and the wall network is connected and well seated in both ends 3 Try a network cable that is known to be OK 4 Connect a network cable between the Scanner and your PC Try to ping from the scanner to the IP address on the PC If ok the hardware connection inside the scanner is OK ...

Page 192: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 55344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting ...

Page 193: ...8 2 3 Standard tools list for VIVID P3 8 3 8 2 4 LCD Assy FRU P N 5315112 8 4 8 2 5 LCD Arm Cover FRU P N 5310093 8 8 8 2 6 Handle P3 Console FRU P N 5315120 8 9 8 2 7 Keyboard Assembly FRU P N 5340674 8 11 8 2 8 TGC PWA FRU Assy FRU P N 5315107 8 15 8 2 9 Alpha Numeric Keyboard with cables FRU P N 5340669 8 17 8 3 Keyboard Rear Cover Assy FRU P N 5315122 8 21 8 4 Mechanicals 8 26 8 5 USB Cable Se...

Page 194: ...for repair ensure that the patient information is erased from the USB Pendrive USB HDD or the USB Pendrive USB HDD is removed from the Vivid P3 before shipping In case if any patient information is still residing on the Vivid P3 GE will contact the customer and request for urgent collection of that patient information GE will keep this patient information in a secured environment for a maximum per...

Page 195: ...4 in Standard Socket set 19 pc Xcelite hex Blade 5 32 inch 3 screw Screwdriver Slotted 1 4 in X 6 in Steel rule Metric 6 inch 4 screw Screw FH M2 5x3 NL Phillips 1 5 screw Screw SFH M2 5x4 NL Phillips 1 6 screw Screw FH M2 5x5 NL Phillips 1 7 screw Screw FHE M2 5x5 NL Phillips 1 8 screw Screw SJ2836 87 M2 5X6 I Phillips 1 9 screw SCREW SJ2836 87 M2 5x12 II Phillips 1 10 screw D2 SCREW M3X3 NYLOK P...

Page 196: ...3 Needed Manpower 1 person 11 minutes travel 8 2 4 4 Preparations Shut Down the System Switch of the circuit breaker at the bottom rear side of the system 8 2 4 5 Removal procedure 1 Shut down the Vivid P3 system 2 LCD monitor rearcover of Vivid P3 system as shown below Figure 8 1 on page 8 4 Figure 8 1 LCD rear cover 3 Bend down the LCD monitor by 90 degree 4 Bend the LCD monitor as shown below F...

Page 197: ...river and remove the power cable cover Refer Figure 8 2 on page 8 5 Figure 8 2 Power cable cover 6 Remove the Power cable cover of LCD monitor as shown below Figure 8 3 on page 8 5 7 Unfasten the cable clamp for power cable and DVI cables and conncted to the LCD monitor Refer Figure 8 3 on page 8 5 Figure 8 3 Power cable DVI cable Power cable cover DVI Cable Power Cable Cable Clamp ...

Page 198: ...sembly of VIVID P3 8 2 4 LCD Assy FRU P N 5315112 cont d 8 Disconnect the power cable DVI cable from the LCD monitor system Refer Figure 8 4 on page 8 6 Figure 8 4 Remove Power cable DVI cable 9 Loosen the two screw of A B of LCD monitor Arm Refer Figure 8 5 on page 8 6 Figure 8 5 LCD monitor neck DVI cable power cable A B ...

Page 199: ...system Refer Figure 8 6 on page 8 7 Figure 8 6 Lift LCD monitor 8 2 4 6 Mounting Procedure 1 Install the new parts in the reverse order of removal 8 2 4 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 2 3 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 4 Adjusting ...

Page 200: ... rear side of the system 8 2 5 4 Removal procedure 1 Remove LCD Power Cable cover LCD Signal Cable Refer Refer section Section 8 3 on page 21 2 Remove LCD Arm Cover Assembly as shown below Figure 8 7 on page 8 8 Figure 8 7 LCD Arm cover monitor neck cover 8 2 5 5 Mounting procedure 1 Install the new parts in the reverse order of removal 8 2 5 6 Functional Checkout Procedure Functional Test Debrief...

Page 201: ... 6 2 FRU BOM This FRU Kit P N 5315120 consists of following parts 8 2 6 3 Needed Manpower 1 person 15 minutes 8 2 6 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 2 6 5 Removal procedure 1 Remove keyboard Top cover and Pull out the keyboard top assembly cover by the slot marked in the Figure 8 8 on page 8 10 2 Remove Gel holder an...

Page 202: ...head fasteners marked 1 2 3 in the Figure 8 8 on page 8 10 Figure 8 8 Handle P3 console 8 2 6 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 2 6 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 2 5 Equipment passes all requi...

Page 203: ...U P N 5340674 This is a description on how to remove and replace the keyboard Assembly 8 2 7 1 Tools Common Torx screwdrivers 8 2 7 2 Needed Manpower 1 person 10 minutes Note The Freeze key is part of Keyboard 8 2 7 3 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 2 7 4 Removal Procedure ...

Page 204: ... Keyboard Assembly FRU P N 5340674 cont d 1 Unscrew the three screws A B C of keyboard bottom cover of the system Refer Figure 8 9 on page 8 12 Figure 8 9 keyboard Bottom view 2 Lift or Pull the Keyboard Top view Assembly of Vivid P3 system Refer Figure 8 10 on page 8 12 Figure 8 10 Move the keyboard out in the given direction Remove three screws ...

Page 205: ...eyboard Assembly FRU P N 5340674 cont d 3 Remove USB cable from keyboard PCB Assembly of system Refer Figure 8 11 on page 8 13 Figure 8 11 Keyboard cable location 4 Remove keyboard interface cable from keyboard Assembly of the system Refer Figure 8 12 on page 8 13 Figure 8 12 Keyboard cable location USB cable keyboard interface cable ...

Page 206: ...ocedure 1 Install the new parts in the reverse order of removal 8 2 7 6 Replace Keyboard Assembly 1 Connect keyboard cable to the new Keyboard Assembly 2 Locate the Keyboard top hinge appropriately 3 Replace keyboard Assembly by fixing three screws a b c 4 Refix Gel holder and Probe holders 8 2 7 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Serv...

Page 207: ... consists of following parts 8 2 8 3 Needed Manpower 1 person 10 minutes 8 2 8 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 2 8 5 Removal procedure 1 Remove keyboard Assembly Refer section 8 2 7 on page 11 2 Remove TGC Caps 8 Nos 3 Remove TGC assembly by removing the four screws as marked in Figure 8 14 on page 8 16 PWA with TGC...

Page 208: ...8 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 2 8 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 2 7 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks TGC Interface cable TGC Assembly ...

Page 209: ... 3 Needed Manpower 1 person 30 minutes 8 2 9 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 2 9 5 Removal procedure 1 Remove Keyboard Assembly Refer section 8 2 7 on page 11 2 Remove the Alphanumeric Keyboard by unfastening the 10 screws marked in Red color in Figure 8 15 on page 8 18 holding it and the Alphanumeric keyboard cable...

Page 210: ... d Figure 8 15 Alpha Numeric keyboard PCB 8 2 9 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 2 9 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 2 8 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks A N Keyboard 10 Sc...

Page 211: ...5315029 consists of following parts 8 2 10 3 Needed Manpower 1 person 10 minutes 8 2 10 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 2 10 5 Removal procedure 1 Remove Keyboard Assembly Refer section 8 2 7 on page 11 2 Unfasten the 2 screws A B holding the trackball bracket circled in Yellow color in Figure 8 16 on page 8 20 PART...

Page 212: ...all Interface cable Refer Figure 8 16 on page 8 20 Figure 8 16 Optical Trackball Assembly 8 2 10 6 Mounting procedure 1 New parts in the reverse order of removal 8 2 10 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 2 9 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown ...

Page 213: ... BOM This FRU Kit P N 5315122 consists of following parts 8 3 1 3 Needed Manpower 1 person 15 minutes 8 3 1 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 3 1 5 Removal procedure 1 Remove Keyboard assembly Refer Section 8 2 6 Figure 8 17 Keyboard 2 Remove the 2 screws A B as shown in the figure above 3 Remove LCD Arm cover as ment...

Page 214: ... on page 8 22 Figure 8 18 keyboard rear cover 8 3 1 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 3 1 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 3 1 Equipment passes all required tests and is ready for use Keyboard re...

Page 215: ...in Circuit Breaker at the bottom rear side of the system 8 3 2 3 Removal procedure 1 Pull the probe holder out by slightly pushing it upwards with your hands Refer Figure 8 19 on page 8 23 Figure 8 19 Probe Holder 8 3 2 4 Mounting procedure 1 Install the new parts in the reverse order of removal 8 3 2 5 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no...

Page 216: ...3 FRU BOM For Encoder set This FRU Kit P N 5262796 consists of following parts NOTE B Mode Gain Knob Encoder is not a FRU 8 3 3 4 Needed Manpower 1 person 15 minutes 8 3 3 5 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 3 3 6 Removal procedure 1 Pull the knobs out with hands by slightly pushing it up from the bottom edge Refer Figu...

Page 217: ... Mounting procedure 1 Install the new parts in the reverse order of removal 8 3 3 8 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 3 3 Equipment passes all required tests and is ready for use ...

Page 218: ...g parts 8 4 1 3 Needed Manpower 1 person 15 minutes 8 4 1 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 1 5 Removal Procedure 1 Remove the screw cap of four places 1 2 3 4 on RH side cover of the system Refer Figure 8 21 on page 8 26 2 Remove the RH side cover from the system Refer Figure 8 21 on page 8 26 Figure 8 21 RH side c...

Page 219: ... Figure 8 22 on page 8 27 Figure 8 22 RH side cover 8 4 1 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 4 1 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 1 Equipment passes all required tests and is ready for use ...

Page 220: ... person 15 minutes 8 4 2 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 2 5 Removal Procedure 1 Remove the screw cap of four places 1 2 3 4 2 Unscrew 4 screws 1 2 3 4 3 Remove the Left cover in the direction as shown in the figure below Figure 8 23 Left Cover 8 4 2 6 Mounting procedure 1 Install the new parts in the reverse orde...

Page 221: ...pter 8 Replacement Procedures 8 29 8 4 2 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 2 Equipment passes all required tests and is ready for use ...

Page 222: ... Needed Manpower 1 persons 15 minutes 8 4 3 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 3 5 Removal Procedure 1 Remove RH side cover from the system Refer section 8 4 1 on page 26 2 Remove LH side cover from the system Refer section 8 4 2 on page 28 3 Remove the two screws A B RH side from the system Refer Figure 8 26 on page...

Page 223: ...1 5 Loosen the two screws A B LH side from the system Refer Figure 8 25 on page 8 31 Figure 8 25 Removing screws of rear cover 6 Loosen Rear cover RH side from the system Refer Figure 8 26 on page 8 31 Figure 8 26 Removing screws of rear cover 8 4 3 6 Mounting procedure 1 Install the new parts in the reverse order of removal ...

Page 224: ... 8 32 Section 8 4 Mechanicals 8 4 3 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 3 Equipment passes all required tests and is ready for use ...

Page 225: ...lowing parts 8 4 4 3 Needed Manpower 1 persons 15 minutes 8 4 4 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 4 5 Removal Procedure 1 Remove the RH side cover from the system Refer section 8 4 1 on page 26 2 Remove the LH side cover from the system Refer section 8 4 2 on page 28 Figure 8 27 Front cover PART NO DESCRIPTION 53100...

Page 226: ...he two screws A B LH side of the system Refer Figure 8 28 on page 8 34 Figure 8 28 Removing side screws of front cover 4 Remove the two screws A B RH side of the system Refer Figure 8 29 on page 8 34 Figure 8 29 Removing side screws of front cover 8 4 4 6 Mounting procedure 1 Install the new parts in the reverse order of removal ...

Page 227: ...pter 8 Replacement Procedures 8 35 8 4 4 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 4 Equipment passes all required tests and is ready for use ...

Page 228: ...nd switch off the main Circuit Breaker at the bottom rear side of the system 8 4 5 5 Removal Procedure 1 Remove the LH side cover from the system Refer section 8 4 2 on page 28 2 Remove the RH side cover from the system Refer section 8 4 1 on page 26 3 Remove the Front cover from the system Refer section 8 4 4 on page 33 4 Remove the Rear cover from the system Refer section 8 4 7 on page 42 5 Loos...

Page 229: ...e LH side EMI cover and remove two screws from LH side bottom of the top assembly of the system Refer Figure 8 31 on page 8 37 Figure 8 31 Removing screw top assembly 7 Remove RH side EMI cover and remove two screws from RH side bottom of the top assembly of the system Refer Figure 8 32 on page 8 37 Figure 8 32 Removing screw top assembly ...

Page 230: ... whole assembly by pulling the top assembly front side from the system Refer Figure 8 33 on page 8 38 Figure 8 33 Top assembly 9 The top assembly should be to removed from two locating pins A B Refer Figure 8 34 on page 8 38 Figure 8 34 Top assembly 8 4 5 6 Mounting procedure 1 Install the new parts in the reverse order of removal A B ...

Page 231: ...pter 8 Replacement Procedures 8 39 8 4 5 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 5 Equipment passes all required tests and is ready for use ...

Page 232: ...sists of following parts Table 8 1 Front Table 8 2 Rear 8 4 6 3 Needed Manpower 1 person 15 minutes each 8 4 6 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 6 5 Removal Procedure 1 Remove the Transformer assy Refer section 8 8 1 on page 75 2 Lift the system slightly up on the side of the castor wheel that needs replacement 3 Pl...

Page 233: ...move it Refer Figure 8 35 on page 8 41 Figure 8 35 Castor wheel 8 4 6 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 4 6 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspectionand roll system around and ensure smooth wheel rotation and sweivwl Service Manual Direction 5344303 100 Section 8 4 6 Equipment passes all required tests and is re...

Page 234: ...M This FRU Kit P N 5308951 consists of following parts 8 4 7 3 Needed Manpower 1 person 15 minutes 8 4 7 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 7 5 Removal Procedure 1 Remove Right cover Refer section 8 4 3 on page 30 for details on how to remove Right cover 2 Remove rear cover Refer section 8 4 3 on page 30 3 Use the sc...

Page 235: ...Rear cover Figure 8 36 Rear side EMI cover 8 4 7 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 4 7 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 7 Equipment passes all required tests and is ready for use Rear side EMI ...

Page 236: ... 8 4 8 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 8 5 Removal Procedure 1 Remove Right cover Refer section 8 4 3 on page 30 for details on how to remove Right cover 2 Use the scewdriver and loosen the eight screws 1 2 3 4 5 6 7 8 3 Slide and remove RH side EMI cover Refer Figure 8 37 on page 8 44 for details on how to remove...

Page 237: ... Mounting procedure 1 Install the new parts in the reverse order of removal 8 4 8 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 8 Equipment passes all required tests and is ready for use ...

Page 238: ... 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 9 5 Removal Procedure 1 Remove Left cover Refer section 8 4 1 on page 26 for details on how to remove Left cover 3 Eight screws to be unfastened of LH side EMI cover of Vivid P3 system Refer Figure 8 38 on page 8 46 2 Remove LH side EMI cover Refer Figure 8 38 on page 8 46 for deta...

Page 239: ...loosen eight screws 1 2 3 4 5 6 7 8 from LH side EMI cover of VividP3 system Refer Figure 8 39 on page 8 47 Figure 8 39 Loosen screw of LH side EMI cover 5 Remove LH side EMI cover of Vivid P3 system Refer Figure 8 40 on page 8 47 Figure 8 40 Remove LH side EMI cover 8 4 9 6 Mounting procedure 1 Install the new parts in the reverse order of removal ...

Page 240: ... 8 48 Section 8 4 Mechanicals 8 4 9 7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure no scratches observed and fitment is done properly Service Manual Direction 5344303 100 Section 8 4 9 Equipment passes all required tests and is ready for use ...

Page 241: ...er 1 person 15 minutes 8 4 10 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 10 5 Removal Procedure 1 Removal of all cable assembly from MST to TX as shown below Figure 8 41 on page 8 49 2 Use a screwdriver to remove the two screw holding the flexi cable to the MST board Refer Figure 8 41 on page 8 49 Figure 8 41 Removing Flexi ...

Page 242: ...CE MANUAL 8 50 Section 8 4 Mechanicals 3 Remove USB cable from MST board Refer Figure 8 42 on page 8 50 Figure 8 42 Removing USB cable of MST board 4 Remove the FRC cable from MST board Refer Figure 8 43 on page 8 50 Figure 8 43 Removing FRC cable of MST board USB cable FRC cable ...

Page 243: ...apter 8 Replacement Procedures 8 51 5 Remove the Audio cable from MST board Refer Figure 8 44 on page 8 51 Figure 8 44 Removing Audio cable of MST board 6 Remove the Power cable from MST board Refer Figure 8 45 on page 8 51 Figure 8 45 Removing Power cable of MST board Audio cable Power cable ...

Page 244: ... Figure B Figure 8 46 MST Assy 8 4 10 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 4 10 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 4 10 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks 4 3 7 M Mode Controls 4 3 ...

Page 245: ...inutes 8 4 11 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 4 11 5 Removal Procedure 1 Remove Right side cover and Right side EMI cover Refer Section 8 4 8 on page 8 44 2 Loosen the one screw A and pull out the hard disk drive assembly as shown in Figure 8 47 on page 8 53 Figure 8 47 Removing Hard Disk Drive Power cable PART NO D...

Page 246: ...ge 8 54 Figure 8 48 Removing SATA cable 8 4 11 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 4 11 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 4 11 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown Sata Signal Cable Power cable ...

Page 247: ...side of the system 8 5 0 5 Removal Procedure 1 Remove the covers and parts as applicable Refer to the applicable section of this document 2 Unplug the USB cable from both the ends and remove the cable from the machine 8 5 0 6 Functional Checkout Procedure Part Number Part Description 5262527 MST To Keyboard 4FTAA 5267540 RPI to Printer 3FT AB2 5262590 Docking Board to Top panel 4FT AA 5269176 RPI ...

Page 248: ...4 consists of following parts 8 5 1 4 Needed Manpower 1 person 15 minutes 8 5 1 5 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 5 1 6 Removal Procedure 1 Loosen the two screw A and pull out the cables which are connected to PDB as shown in rectangle box B in the Figure 8 47 on page 8 53 PART NO DESCRIPTION 5273204 Low voltage Power...

Page 249: ...eplacement Procedures 8 57 2 Remove LV and PDB assembly Refer Figure 8 47 on page 8 53 Figure 8 49 Removing Hard Disk Drive Power cable 3 Remove SATA cable from Hard disk drive assembly Refer Figure 8 48 on page 8 54 Figure 8 50 Removing SATA cable B Power cables A Two Screws B Power cable PDB Bracket ...

Page 250: ... 5315039 8 5 1 7 Mounting procedure 1 Install the new parts in the reverse order of removal 8 5 1 8 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 5 1 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown ...

Page 251: ... 2 FRU BOM This FRU Kit P N 5313708 consists of following parts Table 8 3 110 V Rear Panel Table 8 4 220 V Rear Panel 8 6 1 3 Needed Manpower 1 persons 15 minutes 8 6 1 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 6 1 5 Removal Procedure 1 Remove the rear cover Refer section 8 4 3 on page 30 Part Number Description 5313708 Rear ...

Page 252: ...ear Panel Assy 110V 220V FRU P N 5313708 2 Loosen the five screws 1 2 3 4 of rear panel assembly of the system Refer Figure 8 51 on page 8 60 Figure 8 51 Rear pannel Assembly 3 Push upwards and pull out the Rear panel assembly by the slot marked A in the Figure 8 52 on page 8 61 Rear panel Assy Software DVD ...

Page 253: ...ment Procedures 8 61 4 Remove Rear panel assembly from the system Slide upwards as shown below Figure 8 52 on page 8 61 Figure 8 52 Removing Rear panel assembly 5 Disconnect the FRC cable Power cable and USB Cables from rear pannel assembly Refer Figure 8 53 on page 8 61 Figure 8 53 Removing FRC USB cables ...

Page 254: ...RU P N 5313708 8 6 1 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 6 1 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 6 1 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown ...

Page 255: ...ower and wear the wrist strap before you remove circuit boards do not un plug the power card to keep ground continuity Do not bend or flex the boards when mounting dismounting each boards surface mount IC boards are very susceptible to damage from flex torque 1 Remove Right cover Refer section 8 4 1 on page 26 for details on how to remove right cover 2 Remove Left cover Refer section 8 4 2 on page...

Page 256: ... 8 64 Section 8 6 Rear Panel Assy 110V 220V FRU P N 5313708 Figure 8 55 Removing Conn board assembly 5 Remove the conn board TX cable and PDB conn board cable or power cable Refer Figure 8 56 on page 8 64 Figure 8 56 Removing Power cable Conn board Tx cable Power cable Conn board Tx cable ...

Page 257: ... on page 8 65 Figure 8 57 Loosen screws of flexi cable 7 Remove flexi cable from conn board to Tx board Refer Figure 8 58 on page 8 65 Figure 8 58 Removing Flexi cable conn board 8 Use the screwdriver and remove four screws 1 2 3 4 of connector board from the system Refer Figure 8 54 on page 8 63 9 Slide the connector board out 8 6 2 5 Mounting procedure 1 Install the new parts in the reverse orde...

Page 258: ...5313708 8 6 2 6 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 6 2 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks 4 3 7 M Mode Controls 4 3 8 Color Flow Mode Checks 4 3 9 Doppler Mode Checks ...

Page 259: ...following parts 8 6 3 3 Needed Manpower 1 person 30 minutes 8 6 3 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 6 3 5 Removal Procedure 1 Remove the connector board assembly Refer section 8 6 2 on page 63 2 Loosen the Five screws 1 2 3 4 5 of TX board from the system Refer Figure 8 59 on page 8 67 3 Remove the Tx board from the s...

Page 260: ...r board and other side on TX PWA 7 To remove the TX remove the 7 fasteners holding TX to RX as shown in Figure 8 62 on page 8 70 8 6 3 6 Mounting procedure 1 Install the new parts in the reverse order of removal 8 6 3 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 6 3 Equipment passes all required tes...

Page 261: ...switch off the main Circuit Breaker at the bottom rear side of the system 8 7 0 5 Removal Procedure 1 Remove the connector board assembly Refer section 8 6 2 on page 63 2 Loosen the Five screws 1 2 3 4 5 of TX board from the system Refer Figure 8 59 on page 8 67 3 Remove the Tx board from the system Refer Figure 8 59 on page 8 67 Figure 8 60 TX board Assembly 4 Remove the Tx conn board cable Refer...

Page 262: ...le from TX board as shown below in Figure 8 61 on page 8 70 Figure 8 61 Removing TX conn board cable Flexi cable 6 Remove the Screws fixed in Tx Pwa seperate the RX PWA from TX PWA as shown in Figure 8 62 on page 8 70 Figure 8 62 Tx RX combined 8 7 0 6 Mounting procedure 1 Install the new parts in the reverse order of removal Tx conn board cable Flexi cable ...

Page 263: ...7 0 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 7 0 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks 4 3 7 M Mode Controls 4 3 8 Color Flow Mode Checks 4 3 9 Doppler Mode Checks ...

Page 264: ...parts 8 7 1 3 Needed Manpower 1 person 30 minutes 8 7 1 4 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 7 1 5 Removal Procedure 1 Remove the connector board assembly Refer section 8 6 2 on page 63 2 Loosen the Five screws 1 2 3 4 5 of TX board from the system Refer Figure 8 59 on page 8 67 3 Remove the Tx board from the system Refe...

Page 265: ...ure 8 61 on page 8 70 5 Remove the two flexi cable from TX board as shown below in Figure 8 64 on page 8 73 Figure 8 64 Removing TX conn board cable Flexi cable 6 To remove the Fex cable remove one side from connector board other side on TX PWA 7 Use screwdriver to remove the screws of MST to TX flexi cable and remove the core Refer Figure 8 65 on page 8 73 Figure 8 65 Removing Flexi cable of MST ...

Page 266: ...ew parts in the reverse order of removal 8 7 1 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 7 1 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks 4 3 7 M Mode Controls 4 3 8 Color Flow Mode Checks 4 3 9 Doppler Mode Checks ...

Page 267: ...system NOTE For 110V System Configuration use 110V wired Plug with 3 Amps Fuse NOTE For 230V System Configuration use 230V Wired plug With 2 Amps Fuse NOTE To replace Cable Guard Clamp Donot remove the transformer remove power cord and press the cable clamp from both the sides the clamp will come out side NOTE Before Replacing the transformer remove the Labels stuck on the old transformer and stuc...

Page 268: ...8 Power Block 4 Pull out the transformer tray assembly from the system Refer Figure 8 67 on page 8 76 Figure 8 67 Removing Tansformer Assembly 5 Pull out the transformer tray so the connector marked J4 is accessible and disconnect cable Refer Figure 8 68 on page 8 76 Figure 8 68 Removing Transformer cable ...

Page 269: ...ransformer assy in the reverse order of removal 8 8 1 6 Functional Checkout Procedure See Section Functional and or Leakage Current Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 8 1 Leakage Current measured at Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 10 7 5 Chassis Leakage Current Test ...

Page 270: ...ions Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system 8 9 1 5 Removal Procedure 1 Loosen two screws and remove the DVD tray assembly Refer Figure 8 69 on page 8 78 2 Disconnect the cables connected to the DVD drive To disconnect the cables remove the Keyboard 3 Remove the DVD drive from the DVD tary by removing the four screws holding it Refer Figur...

Page 271: ... MANUAL Chapter 8 Replacement Procedures 8 79 8 9 1 7 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 10 1 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown ...

Page 272: ...rcuit Breaker at the bottom rear side of the system 8 10 0 4 Removal Procedure 1 Pull the metal clamp upwards 2 Remove the power cable as shown in the figure below Figure 8 70 Power Cable 8 10 0 5 Mounting procedure 1 Install the new parts in the reverse order of removal Part Number Description 5337332 Power Cord India 5248395 Power Cord China 5248396 Power cable Europe 5248393 Power cable Japan 5...

Page 273: ...cedures 8 81 8 10 0 6 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 11 0 Leakage Current measured at Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 10 7 5 Chassis Leakage Current Test ...

Page 274: ...he system 8 11 0 5 Removal Procedure 1 Remove the covers and parts as applicable Refer to the applicable section of this document 2 Unplug the signal cable from both the ends 8 11 0 6 Functional Checkout Procedure Part Numbers Description 5265925 LCD signal Cable 5262402 Docking Brd to Speaker 5273129 SATA signal Cable for HDD 5266424 Keyboard ground Cable 5272618 Fan Cable Assembly 5269755 TGC In...

Page 275: ...vers and parts as applicable Refer to the applicable section of this document 2 Unplug the power cable from both the ends 8 12 0 6 Functional Checkout Procedure Part Number Description 5270783 Cable Assembly for Transformer to PS SSR and Peripheral Connector 5262720 Cable Assembly for Panel mount connector to power supply solid state relay 5262046 Peripheral Connector to Rear panel sockets 5269059...

Page 276: ...f the main Circuit Breaker at the bottom rear side of the system NOTE If the CWD Board is Defective Not Present in the image mode Mode not available on this probe message will apear Removal Procedure of CWD Refer to Figure 8 71 Figure 8 71 1 Unscrew two screws 1 2 2 Disconnect the CWD assy 8 12 1 4 Removal procedure Refer to 8 12 1 on page 8 84 1 Unscrew two screws 1 2 to remove the CWD cover see ...

Page 277: ... 1 5 Functional Checkout Procedure See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 13 15 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks 4 3 7 M Mode Controls 4 3 8 Color Flow Mode Checks 4 3 9 Doppler Mode Checks ...

Page 278: ...y immediately after power on and enter BIOS Password for LP3 as Lakshya NOTE Please ensure AC adapter is connected during system upgrade 1 Before loading the Software backup the database For reference see 4 3 3 1 on page 4 5 and after loading the Software restore the backup database 2 Insert the disk labeled System Application Software into the DVD drive 3 Properly turn off the scanner by momentar...

Page 279: ...ARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 8 Replacement Procedures 8 87 6 Press any key to continue when below message display as shown in Figure 8 73 on page 8 87 Figure 8 73 Update message ...

Page 280: ...ystem Select choice a to load the complete disk Refer to Figure 8 74 on page 8 88 8 Press Yes or No to continue 9 System DVD will be loaded as shown in Figure 8 76 on page 8 88 Figure 8 74 Selection for loading the system WARNING WARNING While the software install procedure is designed to preserve data you should select choice b to format disk C only Figure 8 75 Confirmation on loading the system ...

Page 281: ...ocedures 8 89 Section 8 13 Loading Base Image Software cont d 10 After finish updating system refer to Figure 8 77 on page 8 89 11 Press any key to reboot system When system first time boot up after upgrading complete you will see the following message Figure 8 77 System upgrade complete Figure 8 78 Windows start ...

Page 282: ...e rotating security Password 2 Select Utilities 3 Select Invoke SW Reload This brings up the dialog as shown in the figure below Figure 8 79 Invoke SW Reload 4 Call the Online Center OLC Tell OLC that you need a Challenge Code and on request you must read the Challenge string from the screen OLC will return a password that you need to type in as the Response 5 Enter the Response password you got f...

Page 283: ...s until the GRUB menu is displayed on the screen Figure 8 80 GRUB menu 3 On the Grub menu select Recovery Console The GE Healthcare Recovery Console is displayed Enter the Pass word 4 Call the Online Center OLC Tell OLC that you need a Challenge Code and on request you must read the Challenge string on the screen OLC will return a password that you need to type in as the Response 5 Enter the Respo...

Page 284: ...ervice Manual Direction 5344303 100 Section 8 14 0 Equipment passes all required tests and is ready for use 4 3 2 Power Off Shutdown 4 3 6 B Mode Checks 4 3 7 M Mode Controls 4 3 8 Color Flow Mode Checks 4 3 9 Doppler Mode Checks See Section Functional Test Debrief Script 4 3 1 Power On Boot Up Service Manual Direction 5344303 100 Section 8 14 1 Equipment passes all required tests and is ready for...

Page 285: ...placement part available in part hub KBD Keyboard LCD Liquid Crystal Display BnV Brightness and Volume RX64 Front Processor Board TMST Master Board TX64 Transmit Board PDB Power Distribution Board Table 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Overview 9 1 9 2 List of Abbreviations 9 1 9 3 Operator Console Assy 9 2 9 4 LCD Assy 9 3 9 5 Keyboard Assy 9 4 9 6 Mechanical Assy 9 5...

Page 286: ...Operator Console Assy Figure 9 81 OPERATOR CONSOLE ASSY 1 Peripheral Device Signals I O Port Power In 9 Gel Holder 2 Three Probe Port 10 VGA Connector 3 ECG 11 Ethernet Connector 4 DVD RW Drive 12 S Video Connector 5 Monitor 13 BNC Connector 6 USB Port 14 USB Ports 7 B W Printer Option 15 Circuit Breaker 8 Probe Holder 16 Power Connector ...

Page 287: ...REVISION 3 VIVID P3 SERVICE MANUAL Chapter 9 Renewal Parts 9 3 Section 9 4 LCD Assy Figure 9 82 LCD Assy Table 9 3 LCD Assy Item Part Name Part Name Description Quantity 100 LCD Monitor 5315112 LCDMON II 1 101 LCD Arm Cover 5310093 LCD Arm Cover 1 100 101 ...

Page 288: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 9 4 Section 9 5 Keyboard Assy Section 9 5Keyboard Assy Figure 9 83 Keyboard Assy 200 202 201 203 204 ...

Page 289: ...s 9 5 Section 9 5 Keyboard Assy cont d Table 9 2 Keyboard Assy Item Part Name Part Number Description Quantity 200 Keyboard Assy 5340674 Keyboard Assy 1 201 A N Keyboard Assy 5340669 A N Key Assy 1 202 TGC Assy 5315107 TGC 1 203 Track Ball Assy 5315029 Track Ball Assy 1 204 Knob Set 5315502 Gain Knob TGC Knob 1 ...

Page 290: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 9 6 Section 9 6 Mechanical Assy Section 9 6Mechanical Assy Figure 9 84 Mech Assy ...

Page 291: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 9 Renewal Parts 9 7 Section 9 6 Mechanical Assy cont d Figure 9 85 Mech Assy ...

Page 292: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 9 8 Section 9 6 Mechanical Assy ...

Page 293: ...r A Left cover B 5350996 5350992 Right cover A Left cover B 1 304 Top Cover Assy 5315121 Top Cover 1 305 Probe Gelbottle holder 5340677 5340678 Probe Gelbottle holder 1 306 Rear cover 5315117 Rear cover 1 307 Front cover 5315116 Front cover 1 308 EMI Cover Ser 5315111 EMI Cover Set includes 1 LH EMI cover 2 RH EMI cover 3 Rear EMI cover 1 309 RPA 230V RPA 100V 5313708 Rear Panel ASSY 100V 220V 1 3...

Page 294: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 9 10 Section 9 7 Pwa Assy Section 9 7Pwa Assy Figure 9 86 PWA Assy 401 402 403 ...

Page 295: ...er Description Quantity 401 CWD 5124847 CWD 1 402 Sata Hard disk Assy 5315113 SATA HArd Disk Assy 160GB 1 403 3PP Connector Board Assy 5314411 2 3PP Connector Board Assy 1 404 MST 5315025 MST 1 405 SOM 5308994 System on Module SOM 1 406 RX PWA 5315027 RX PWa ASM with Bracket 1 407 TX PWA 5389402 TX PWA 1 408 LV PDB power supply 5315028 LV Power Distribution Board assy 1 404 406 407 408 405 ...

Page 296: ...L 9 12 Section 9 8 Flex Pcb Cables Section 9 8Flex Pcb Cables Wiring Diagram Refer Chapter 5 page 5 8 Figure 9 88 Cable Assy Item Part Name Part Num Description Quantity 501 Flex Cable Con to Tx 5315108 Flex Cable Con to Tx 1 502 Flex Cable MST to TX 5315109 Flex Cable MST to TX 1 502 501 ...

Page 297: ...344303 100 REVISION 3 VIVID P3 SERVICE MANUAL Chapter 9 Renewal Parts 9 13 Section 9 9Signal Cable set Power Cable set Fru USB Cables Fru Part num Wiring Diagram Refer Chapter 5 Sec 5 8 Figure 9 89 Cable Assy 503 507 505 504 506 Power Cable Set ...

Page 298: ...VID P3 SERVICE MANUAL 9 14 Section 9 9 Signal Cable set Power Cable set Fru USB Cables Fru Part num Section 9 9 Signal Cable set Power Cable set Fru USB Cables Fru Part num cont d Figure 9 90 Cable Assy cont d 509 511 513 514 512 510 508 Signal Cable Set USB Cable Set ...

Page 299: ...ower Cable FRU Set 1 Cable Assembly for Transformer to PS SSR and Peripheral Con nector 2 PDB to Conn Brd and MST Brd 3 PDB to KBD and Fan 4 LCD Power cable 5 Dual SATA Power Cable for PDB to HDD and DVD 1 508 509 510 514 Signal Cable set 5315042 Parts under Signal Cable FRU Set 1 SATA Signal Cable to DVD 2 MST To Keyboard 4FTAA 3 LCD signal Cable 4 SATA signal Cable for HDD 1 511 512 513 USB Cabl...

Page 300: ...308 2 Vivid P3 Application Software with Base ImageDVD old version 1 606 Vivid P3 Application Software with Base ImageDVD R1 0 3 5344308 3 Vivid P3 Application Software with Base ImageDVD new version 1 607 USB Hard Disk 160GB Option 5313598 USB Hard Disk 160GB Option 1 608 DeskJet Printer Color Option 5309104 Option part Hp 5400 1 609 Laser Jet B W Printer option xxxxxxx Option Part Hp2015dn 1 610...

Page 301: ...5123455 Probe Center Frequency 3 2MHz 1 702 8C 2348094 Probe Center Frequency 6 5MHz 1 703 5CS 5143934 Probe Center Frequency 5 6MHz 1 704 8L 5140738 Probe Center Frequency 6 2MHz 1 705 3S 2323337 Probe Center Frequency 2 0MHz 1 706 11L 5171885 Probe Center Frequency 7 5MHz 1 707 E8CS 47236865 Probe Center Frequency 6 5MHz 1 708 T739 2259246 Probe Center Frequency 6 4MHz 1 709 6S 47236867 Probe Ce...

Page 302: ...mbers 5314414 Fan Assembly 5315026 CWD Assy 5315104 Power distribution Board Assembly 5314404 Keyboard Asm complete 5315118 Front Castor Wheel Assy 5315119 Back Castor Wheels Assy 5219747 Cable Guard Clamp 5196448 Transformer Assembly 5351324 Knob Set 5262765 ECG Module Option 5312274 SATA DVD Writer 5308996 1GB_USB_Thumb_Drive 5262796 Encoder set for KBD 5316731 NTPUI Encoder PWA Asm ...

Page 303: ...s performance Read this chapter completely and familiarize yourself with the procedures before performing a task Table 10 1 Contents in Chapter 10 Section Description Page Number 10 1 Overview 10 1 10 2 Why do Maintenance 10 2 10 3 Maintenance Task Schedule 10 2 10 4 Tools Required 10 4 10 5 System Maintenance 10 7 10 6 Using a Phantom 10 12 10 7 Electrical Safety Tests 10 12 10 8 When There s Too...

Page 304: ...ction and the effects of corrective action must be documented and maintained on the site Your GE service representative can help you with establishing performing and maintaining records for a quality assurance program Please contact us for coverage information and or price for service Section 10 3 Maintenance Task Schedule 10 3 1 How often should care maintenance tasks be performed The Care Mainte...

Page 305: ...r before each use Clean Probe Holders Inspect AC Mains Cable Mobile Unit Check Weekly Inspect Cables and Connectors Clean Console Clean Air filter More frequently depending on your environment Clean LCD Inspect Wheels Casters brakes and Swivel locks Mobile unit check Daily Console Leakage Current Checks See Note Twice Annually Peripheral Leakage Current Checks See Note Twice Annually Surface Probe...

Page 306: ...ent 9 4151 Screwdriver Blade Stubby 9 41322 Pickup Tool Claw type 9 41421 Screwdriver Blade Pocket clip 9 6757 6 pc Needle File Set 9 41594 Screwdriver Blade 1 8 in x 4 in 9 9487 Utility Knife 9 41581 Screwdriver Blade 3 16 in x 4 in 9 45341 Pliers Vice Grip 10 inch 9 39451 20 Steel Tape locking Spring load 9 3001 Xacto Pen Knife 9 GH807 Ratchet Offset Slotted 9 HT62002 Solder Aid Fork and Hook 68...

Page 307: ...tion 9 45381 Pliers Arc Joint 9 1 2 inch 9 44067 Socket 1 1 16 in for 1 2 in drive 9 45092 Pliers Linesman 8 1 2 inch 9 42679 Socket 10MM Hex for 1 2 in drive 2273333 9 42882 Punch Pin 3 32 inch 9 44262 Extension 10 inch for 1 2 in drive 2273405 9 42884 Punch Pin 5 32 inch 9 4258 3 8 inch to 1 2 inch Adapter 9 42886 Punch Pin 1 4 inch 9 34374 3 8 inch Metric Socket Set 12 PT 9 42973 Cold Chisel 1 ...

Page 308: ...ti static mat wrist strap and cables for 200 to 240 V system 3M 2204 Large adjustable wrist strap 3M 2214 Small adjustable wrist strap 3M 3051 conductive ground cord Anti Static Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Safety Analyzer 46 285652G1 DALE 600 KIT or equivalent for electrical tests QIQ Phantom E8370RB RMI Grayscale Target Model 403GS B W Printer Cleaning Sheet See printer u...

Page 309: ... Step Item Description 1 Ask Listen Ask the customer if they have any problems or questions about the equipment 2 Paperwork Fill in the top of the Ultrasound Inspection Certificate see page 10 30 Note all probes and system options 3 Power up With AC input Turn the system power on and verify that all fans and peripherals turn on Watch the displays during power up to verify that no warning or error ...

Page 310: ...ation M Mode Verify basic M Mode operation Check the basic system controls that affect this mode of operation Applicable Software Options Verify the basic operation of all optional modes such as Multi Image 3D Harmonics Cine etc Check the basic system controls that affect each options operation TXmit Recv Elements Use the Visual Channel Utility on the loop connect to verify that all system Txmit r...

Page 311: ... footswitch is functioning as programed Clean as necessary 5 ECG Verify basic operation with customer 6 DVD Verify that the DVD is functioning properly Clean heads and covers if necessary Table 10 9 AC Inspection Step Item Description 1 Unplug Cord Disconnect the mains cable from the wall and system 2 Inspect Inspect it and its connectors for damage of any kinds 3 Verify Verify that the LINE wires...

Page 312: ...Control Panel Inspect keyboard and control panel Note any damaged or missing items 4 Cables Connectors Check all internal cable harnesses and connectors for wear and secure connector seating Pay special attention to footswitch assembly and probe strain or bend reliefs 5 Shielding Covers Check to ensure that all EMI shielding internal covers air flow panels and screws are in place Missing covers an...

Page 313: ...equipment damage Failure to properly handle or maintain a probe may also void its warranty Any evidence of wear indicates the probe cannot be used Do a visual check of the probe pins and system sockets before plugging in a probe 10 5 7 3 Basic Probe Cleaning Refer to the User s Manual for details on probe cleaning NOTE To help protect yourself from blood borne diseases wear approved disposable glo...

Page 314: ...ectrical failure Do not use probes or equipment having excessive leakage current To minimize the risk that a probe may shock someone the customer should Not use a probe that is cracked or damaged in any way Check probe leakage current once a year on surface probes twice a year on endocavitary probes whenever probe damage is suspected WARNING WARNING THE USER MUST ENSURE THAT THE SAFETY INSPECTIONS...

Page 315: ...surement limits per IEC 60601 1 Medical Equipment Safety Standards Table IV Most meters like Dale 600 601 measure a composite of the AC and DC leakage current Table 10 13 Chassis Leakage Current Limits Accessible Metal Surfaces Normal Condition Open Ground Open Neutral 0 1 mA 0 5 mA 0 5 mA Table 10 14 Type BF Applied Part Leakage Current Limits Non Conductive Floating Surface and Cavity Probes All...

Page 316: ...spital immediately and the receptacle should not be used The Dale 600 has self contained lamps designed for testing the outlet wiring arrangement Plug the Dale 600 into each outlet to be tested comparing the lamp status NOTE No outlet tester can detect the condition where the Neutral grounded supply conductor and the Grounding protective earth conductor are reversed If later tests indicate high le...

Page 317: ...ps to test the ground wire resistance 1 Turn the VIVID P3 unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of the VIVID P3 unit Set the meter s FUNCTION switch to the RESISTANCE position 4 Set the meter s POLARITY switch to the OFF ...

Page 318: ...of the attachment plug to the meter s Chassis Cable clamp Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 601 1 1 procedure as above Refer to the Dale 600 Instruction Manual for meter self tests and operation Record measured resistance of the grounding continuity The ground wire resistance should be less than 0 2 Use any safety analyzer Figure 10 4...

Page 319: ... with the unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current When using the Microguard or a similar test instrument its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter This places the meter in the grounding conductor and the current flowing from the case to grou...

Page 320: ...akage current 1 Turn the VIVID P3 unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of the VIVID P3 4 Set the tester s FUNCTION switch to CHASSIS position 5 Follow the test conditions described for respective test points shown in Tab...

Page 321: ...de with the ground open and closed with power line polarity normal and reversed and with the ultrasound console Off and On For each combination the operating controls such as the lead switch should be operated to find the worst case condition Generic Procedure Table 10 17 Typical Data Sheet for enclosure Source Leakage Current Unit Power Tester Polarity Switch Tester Neutral or Ground Switch Test ...

Page 322: ... meter and the meter into the tested AC wall outlet 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of the VIVID P3 unit 4 Connect the patient leads to the corresponding snaps located at the upper front of the Dale 600 600E Lead nomenclature for this test is not important Figure 10 6 Test Circuit for Measuring Non ...

Page 323: ...easure the patient lead to ground leakage current 6 Select and test each ECG lead positions except ALL of the LEAD selector testing each to the power condition combinations Figure 10 7 ECG Leakage Current Test Table 10 18 Testing Power Conditions ECG Power Meter s Polarity Switch Meter s Neutral Switch ON NORM CLOSED ON NORM OPEN ON REVERSE CLOSED ON REVERSE OPEN OFF NORM CLOSED OFF NORM OPEN OFF ...

Page 324: ... patient lead isolation current measured on the data sheet 10 7 7 Isolated Patient Lead Source Leakage Lead to Lead Reference the procedure in the IEC 60601 1 When using the Dale 600 switch the meter s function selector to the LEAD LEAD position Select and test each of the five ECG lead positions except ALL on the LEAD selector testing each to the power condition combinations found in the table Re...

Page 325: ...Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current Test 115V 10uA 10uA 230V 500uA 10uA Table 10 20 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit Patient Lead Isolation Current Test 115V 20uA 230V 5mA Table 10 21 Typical Data Sheet for ECG Leakage Current ECG Power Tester Polarity S...

Page 326: ...10 7 9 2 Test Equipment Table 10 22 Test Equipment kits Chapter 10 VIVID P3 Service Manual Kit Description Contents 46 285652G1 Dale 601 ULTRASOUND SAFETY ANALYZER FIELD KIT for 120V unit ELECTRICAL SAFETY ANALYZER DALE MODEL 601 for 120V unit Vendor MODEL 600 100 8FT CHASSIS GROUND CABLE Vendor MODEL 600 102 6FT SALINE BATH GROUND CABLE Vendor MODEL 600 202 ULTRASOUND PROBE ADAPTER Vendor MODEL 6...

Page 327: ...for tight spaces Also referred to as Saline Probe or Saline Bath Ground Cable Measures earth resistance and enclosure leakage current Also used for grounding saline baths for isolation testing of probes Used on DALE601 601E This probe may be substituted for the 600 100 Chassis Cable and used as a probe instead of a clamp Vendor MODEL 600 200 46 285647P6 ISO EXTERNAL LEAKAGE CABLE Black coiled cord...

Page 328: ...ughly Follow these steps to test each probe for leakage current 1 Turn OFF the VIVID P3 unit 2 Plug the unit s mains power cord into the test meter and plug the test meter into the tested AC wall outlet 3 Plug the Chassis Ground Probe saline probe into the test meter s CHASSIS connector 4 Set the test meter s FUNCTION switch to CHASSIS Dale 600 or ENCLOSURE LEAKAGE Dale 601 5 Connect the probe to ...

Page 329: ...EXTERNAL position 12 Add the saline probe and the imaging area of the probe into the saline bath 13 Have unit power ON for the first part turn it OFF for the second half 14 Depress the ISO TEST rocker switch and record the highest current reading 15 Follow the test conditions described in Table 10 24 for every transducer 16 Keep a record of the results with other hand copies of PM data Figure 10 9...

Page 330: ...quipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the POLARITY switch and or the NEUTRAL switch Otherwise the unit may be damaged Table 10 24 Typical Data Sheet For Transducer Source Leakage Current Transducer Tested Unit Power Tester Power Polarity Switch Tester GROUND or NUETR...

Page 331: ...e another probe to confirm if the fail is caused by console NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test lead placement The maximum allowable leakage current for body surface contact probe differs from inter cavity probe Be sur...

Page 332: ...Customer Name System ID Dispatch Number Date Performed Warranty Contract HBS System Type Model Number Serial Number Manufacture Date Probe 1 Frequency Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number Serial Number Probe 4 Frequency Scan Format Model Number Serial Number Probe 5 Frequency Scan Format Model Num...

Page 333: ...___________________________________________________________ Electrical Test Performed Max Value Allowed Value Measured OK Comments Outlet correct ground wiring config Type BF Applied Part Leakage Current Limits Probe enclosure Source Leakage Current Chassis Leakage Current Limits Peripheral 1 Leakage Current Peripheral 2 Leakage Current PROBES Probe Number from previous page Max Value Allowed Max ...

Page 334: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5344303 100 REVISION 3 VIVID P3 SERVICE MANUAL 10 32 Section 10 8 When There s Too Much Leakage Current ...

Page 335: ... Electrostatic Discharge Warning 18 EMI 18 ESD 18 Exam study display location 9 F Focal zone display location 9 Functional Checks Basic Measurements 30 Control Panel 7 Image Management 32 Monitor Display 9 Peripherals 43 Probes Connector Usage 31 G Gathering Trouble Data 2 General Cleaning 9 Gray color bar display location 9 H Hazard Icons 4 Hospital name display location 9 Human Safety 9 I Image ...

Page 336: ...tifier display location 9 Probe orientation marker display location 9 Probes Connector Usage Functional Checks 31 R Required Features 8 Restore Patient Database 36 Preset Configurations 36 S Safety Considerations 9 Screen Captures 6 System Maintenance 7 System Manufacturer 20 T TGC display location 9 Touch Panel Functional Checks 7 Trouble Image with Logs 3 Troubleshooting Gathering Trouble Data 2...

Reviews: