Guidance and manufacturer’s declaration – electromagnetic emissions
The AED PRO is intended for use in the electromagnetic environment specified below. The customer or the user of the
AED PRO should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The AED PRO uses RF energy only for its
internal function. Therefore its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
The AED PRO is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Guidance and manufacturer’s declaration – electromagnetic immunity
The AED is intended for use in the electromagnetic environment specified below. The customer or the user of the AED
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
Not applicable
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
61000-4-11
<5 % U
T
(>95 % dip in U
T
) for 0.5
cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Not applicable
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
80 A/m
Power frequency magnetic fields
should be at levels no higher than
those characteristic of a typical
location in typical heavy industrial
and power plants and the control
rooms of H.V. sub-stations.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.